Disinfection

消毒
  • 文章类型: Journal Article
    Objective: To evaluate the disinfection effect of high-energy pulse ultraviolet disinfection equipment in medical institution settings. Methods: The disinfection effect was evaluated through field tests and laboratory tests. Among them, 135 high-frequency contact points were selected from nine departments in the field test. Samples were collected before and after disinfection, and the disinfection effects of 75% alcohol wipes wiping disinfection, high-energy pulse ultraviolet disinfection robot disinfection and high-energy pulse ultraviolet handheld disinfection instrument were compared. In the laboratory test, 30 infected areas of the simulated test table were exposed to vertical ultraviolet irradiation and the bacterial-killing rate before and after disinfection was calculated. Results: In the field test, the bacteria-killing rates of 75% alcohol wipes, high-energy pulse ultraviolet disinfection robot and high-energy pulse ultraviolet handheld disinfection instrument were 94.99%, 91.53% and 95.94%, respectively, and the difference was statistically significant. The disinfection effect of the high-energy pulse ultraviolet handheld disinfection instrument was better than that of the high-energy pulse ultraviolet disinfection robot (P values <0.05). In the laboratory test, the killing log value of Staphylococcus aureus and Escherichia coli on the carrier were both greater than 3.00. In the simulated field test, the killing log value of Staphylococcus aureus on the surface samples were 4.99. Conclusion: Both the high-energy pulse ultraviolet handheld disinfection instrument and the high-energy pulse ultraviolet disinfection robot have good disinfection effects, which are similar to the disinfection effects of conventional 75% alcohol wipes.
    目的: 评价医疗机构场景下高能脉冲紫外线消毒设备消毒效果。 方法: 分别通过现场试验和实验室试验进行消毒效果评价。其中,现场试验选择9个科室135个高频接触点位,消毒前后采样,比较75%酒精湿巾擦拭消毒、高能脉冲紫外消毒机器人消毒和高能脉冲紫外手持消毒仪消毒效果;实验室试验将模拟试验桌面的30个染菌区块置于紫外线垂直照射下,计算消毒前后细菌杀灭率。 结果: 现场试验中75%酒精湿巾擦拭、高能脉冲紫外消毒机器人和高能脉冲紫外手持消毒仪细菌杀灭率分别为94.99%、91.53%和95.94%,差异有统计学意义,其中高能脉冲紫外手持消毒仪的消毒效果好于高能脉冲紫外消毒机器人(P<0.05)。实验室试验中高能脉冲紫外手持消毒仪对染菌载体上金黄色葡萄球菌和大肠杆菌的杀灭对数值均>3.00,模拟现场试验中高能脉冲紫外手持消毒仪对物体表面样本上金黄色葡萄球菌的杀灭对数值为4.99。 结论: 高能脉冲紫外手持消毒仪和高能脉冲紫外消毒机器人消毒效果均较好,与常规75%酒精湿巾擦拭消毒效果相近。.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: English Abstract
    超声诊断和治疗易于执行,花费的时间很少。由于其非侵入性,它在临床实践中被广泛使用,实时,和动态特性。在超声诊断和治疗过程中,探针可能会与皮肤接触,粘膜,甚至身体的无菌部分。然而,使用后难以对探头进行有效的实时消毒,并且经常重复使用,导致探针携带多种致病菌的可能性。目前,国内外探针的加工方法主要包括探针清洗、探头消毒,和物理隔离(使用探头盖或护套)。然而,每种方法都有其局限性,不能完全防止超声诊断和治疗引起的探头污染和感染。例如,当避孕套被用作探针鞘时,安全套破损率相对较高。保鲜膜或冷冻袋的切割和固定涉及复杂的程序并且难以执行。一次性塑料手套容易脱落并造成污染,因此不符合无菌原则。此外,一次性塑料手套的成像效果差。因此,迫切需要探索新材料,以制造不仅可以紧紧包裹超声探头的探头盖,还有助于实现有效保护和快速重用。基于物理障碍的概念,在这项研究中,我们开发了一种热密封系统,用于快速重复使用超声探头。该系统使用热封装置使保护膜收缩,使其紧贴超声探头表面,允许快速重复使用探针,同时降低医院感染的风险。目的设计超声探头快速复用热封系统,并验证其在超声探头快速复用中的应用效果。
    1)通过整合医学和工程方法,设计和测试了用于快速重复使用超声探头的热封系统。该系统包括保护膜(多层共挤出聚烯烃热收缩膜)和热密封装置,其中包括加热丝组件,鼓风机,一个光电开关,温度传感器,控制和驱动电路板,等。根据热收缩原理,快速加热配有热收缩膜的超声探头,膜将紧密包裹在放置在热封机顶部的超声探头周围。超声探头在热收缩过程完成后准备使用。在探头表面安装温度传感器,测试系统的隔热性能。该系统的操作程序如下:将覆盖有保护膜的超声探头放置在保护通风口上方的一定空间中,由光电开关检测;加热装置在设定温度值下,用恒定的热风流量对热收缩膜进行加热。然后,探针被旋转,使得热收缩膜将快速地缠绕在超声探针周围。热收缩完成后,探头可以直接使用。2)采用便利抽样方法,麻醉和围手术期医学部的90名患者,以西安交通大学第一附属医院为研究对象。所有患者均在超声引导下进行动脉穿刺。受试者分为3组,每组30名患者。使用临床上常用的三种方法对三组探针进行处理,并在使用前在穿刺部位周围进行水溶性荧光标记。在实验组中,探头用热封系统处理。执行用于快速重复使用超声探头的热密封系统的标准操作程序以覆盖超声探头并形成物理屏障以防止探头污染。有两个对照组。对照组1使用含双链季铵盐的消毒湿巾反复擦拭探头表面10-15次,然后探针一旦干涸就可以使用了。在对照组2中,使用一次性保护套覆盖探针的前端,并用螺纹将护套的手柄端绑住。使用前后探针表面的水溶性荧光标记(反映探针表面的菌落残基)和重复使用时间(即,从第一次使用结束到第二次使用开始的时间)在实验组和两个对照组之间进行。
    1)超声探头内部的温度低于40℃,用于快速重复使用的热封系统不影响超声探头的性能。2)热封系统组中的重复使用时间,表示为(中位数[P25,P75]),是(8.00[7.00,10.00])s,显著低于消毒擦拭组(95.50[8.00,214.00])s和保护套组(25.00[8.00,51.00])s,差异有统计学意义(P<0.05)。使用后,在热封系统组或保护护套组中的探针上均未发现荧光残留物。热封系统组的荧光残留量明显低于消毒湿巾组,差异有统计学意义(χ2=45.882,P<0.05)。
    本研究中设计和开发的热收缩膜可以根据设备的尺寸进行切割和修剪。当薄膜被加热时,它收缩并紧紧地包裹在设备上,形成坚固的保护层。随着超声波探头快速重复使用的热封系统,实现了热收缩膜与加热装置的半自动连接,减少了耗时和复杂的手工操作。此外,缩短了平均重用时间,系统易于使用,这有助于提高超声探头的重复使用和操作效率。热封系统减少了探针表面上的菌落残留,并在探针上形成有效的物理屏障。在研究中没有探针被损坏。超声探头快速重复使用的热封系统可作为超声探头加工的一种新方法。
    UNASSIGNED: Ultrasound diagnosis and treatment is easy to perform and takes little time. It is widely used in clinical practice thanks to its non-invasive, real-time, and dynamic characteristics. In the process of ultrasound diagnosis and treatment, the probe may come into contact with the skin, the mucous membranes, and even the sterile parts of the body. However, it is difficult to achieve effective real-time disinfection of the probes after use and the probes are often reused, leading to the possibility of the probes carrying multiple pathogenic bacteria. At present, the processing methods for probes at home and abroad mainly include probe cleaning, probe disinfection, and physical isolation (using probe covers or sheaths). Yet, each approach has its limitations and cannot completely prevent probe contamination and infections caused by ultrasound diagnosis and treatment. For example, when condoms are used as the probe sheath, the rate of condom breakage is relatively high. The cutting and fixing of cling film or freezer bags involves complicated procedures and is difficult to perform. Disposable plastic gloves are prone to falling off and causing contamination and are hence not in compliance with the principles of sterility. Furthermore, the imaging effect of disposable plastic gloves is poor. Therefore, there is an urgent need to explore new materials to make probe covers that can not only wrap tightly around the ultrasound probe, but also help achieve effective protection and rapid reuse. Based on the concept of physical barriers, we developed in this study a heat sealing system for the rapid reuse of ultrasound probes. The system uses a heat sealing device to shrink the protective film so that it wraps tightly against the surface of the ultrasound probe, allowing for the rapid reuse of the probe while reducing the risk of nosocomial infections. The purpose of this study is to design a heat sealing system for the rapid reuse of ultrasound probes and to verify its application effect on the rapid reuse of ultrasound probes.
    UNASSIGNED: 1) The heat sealing system for the rapid reuse of ultrasound probes was designed and tested by integrating medical and engineering methods. The system included a protective film (a multilayer co-extruded polyolefin thermal shrinkable film) and a heat sealing device, which included heating wire components, a blower, a photoelectric switch, temperature sensors, a control and drive circuit board, etc. According to the principle of thermal shrinkage, the ultrasound probe equipped with thermal shrinkable film was rapidly heated and the film would wrap closely around the ultrasound probe placed on the top of the heat sealing machine. The ultrasound probe was ready for use after the thermal shrinkage process finished. Temperature sensors were installed on the surface of the probe to test the thermal insulation performance of the system. The operation procedures of the system are as follows: placing the ultrasound probe covered with the protective film in a certain space above the protective air vent, which is detected by the photoelectric switch; the heating device heats the thermal shrinkable film with a constant flow of hot air at a set temperature value. Then, the probe is rotated so that the thermal shrinkable film will quickly wrap around the ultrasound probe. After the heat shrinking is completed, the probe can be used directly. 2) Using the convenience sampling method, 90 patients from the Department of Anesthesiology and Perioperative Medicine, the First Affiliated Hospital of Xi\'an Jiaotong University were included as the research subjects. All patients were going to undergo arterial puncture under ultrasound guidance. The subjects were divided into 3 groups, with 30 patients in each group. Three measures commonly applied in clinical practice were used to process the probes in the three groups and water-soluble fluorescent labeling was applied around the puncture site before use. In the experimental group, the probes were processed with the heat sealing system. The standard operating procedures of the heat sealing system for rapid reuse of ultrasonic probes were performed to cover the ultrasonic probe and form a physical barrier to prevent probe contamination. There were two control groups. In control group 1, disinfection wipes containing double-chain quaternary ammonium salt were used to repeatedly wipe the surface of the probe for 10-15 times, and then the probe was ready for use once it dried up. In the control group 2, a disposable protective sheath was used to cover the front end of the probe and the handle end of the sheath was tied up with threads. Comparison of the water-soluble fluorescent labeling on the surface of the probe (which reflected the colony residues on the surface of the probe) before and after use and the reuse time (i.e., the lapse of time from the end of the first use to the beginning of the second use) were made between the experimental group and the two control groups.
    UNASSIGNED: 1) The temperature inside the ultrasound probe was below 40 ℃ and the heat sealing system for rapid reuse did not affect the performance of the ultrasound probe. 2) The reuse time in the heat sealing system group, as represented by (median [P25, P75]), was (8.00 [7.00, 10.00]) s, which was significantly lower than those of the disinfection wipe group at (95.50 [8.00, 214.00]) s and the protective sleeve group at (25.00 [8.00, 51.00]) s, with the differences being statistically significant (P<0.05). No fluorescence residue was found on the probe in either the heat sealing system group or the protective sheath group after use. The fluorescence residue in the heat sealing system group was significantly lower than that in the disinfection wipes group, showing statistically significant differences (χ 2=45.882, P<0.05).
    UNASSIGNED: The thermal shrinkable film designed and developed in this study can be cut and trimmed according to the size of the equipment. When the film is heated, it shrinks and wraps tightly around the equipment, forming a sturdy protective layer. With the heat sealing system for rapid reuse of ultrasonic probes, we have realized the semi-automatic connection between the thermal shrinkable film and the heating device, reducing the amount of time-consuming and complicated manual operation. Furthermore, the average reuse time is shortened and the system is easy to use, which contributes to improvements in the reuse and operation efficiency of ultrasound probes. The heat sealing system reduces colony residues on the surface of the probe and forms an effective physical barrier on the probe. No probes were damaged in the study. The heat sealing system for rapid reuse of ultrasonic probes can be used as a new method to process the ultrasonic probes.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    目的是分析根尖壁架对根管消毒的影响。对44颗单根牙齿进行显微CT扫描,并接种粪肠球菌。在S组中,在工作长度(WL)处用ProTaperNext(PTN)直到X3进行成形。在L组中,使用K-Files#40创建了根尖壁架,并完成了直到PTNX3的整形。交替使用5%NaOCl和10%EDTA冲洗液。使用共聚焦激光扫描显微镜(CLSM)和活力染色来分析死(红色)和活(绿色)细菌的比例以及牙本质小管内的渗透能力。用具有Bonferroni校正的Mann-Whitney检验分析数据(p<0.05)。在L组中,红色荧光的量显著降低,根尖和中段穿透能力下降(p<0.05)。根尖壁架的存在可能会对根尖和中部三分之一的消毒产生负面影响。
    The aim was to analyse the influence of an apical ledge on root canal disinfection. Forty-four single-rooted teeth were micro-CT scanned and inoculated with Enterococcus faecalis. In Group S shaping was performed with ProTaper Next (PTN) up to X3 at working length (WL). In Group L an apical ledge was created with K-Files #40 and shaping completed up to PTN X3. NaOCl 5% and EDTA 10% irrigant solutions were alternated. Confocal laser scanning microscope (CLSM) and viability staining were used to analyse the proportions of dead (red) and live (green) bacteria and penetration ability inside dentinal tubules. Data were analysed with the Mann-Whitney test with Bonferroni correction (p < 0.05). In Group L the amount of red fluorescence resulted significantly lower, and penetration ability was decreased in the apical and middle portion (p < 0.05). The presence of an apical ledge may negatively influence the disinfection both in the apical and middle third.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    背景:安装的UV-C灯不需要单独的空间或劳动力进行运输,但是关于其表面消毒效果的数据很少。我们旨在评估天花板安装的UV-C灯的性能。
    方法:这项研究是在韩国一家三级保健医院的生物防护单元的一个装有UV-C灯的空房间中进行的。在病房的天花板上安装了UV-C灯,前厅,和厕所。每种病原体(金黄色葡萄球菌,大肠杆菌,克鲁斯念珠菌,蜡样芽孢杆菌,和Peregrinum分枝杆菌)接种在血琼脂平板上,并从UV-C灯放置在20个选定的地方,并照射15分钟。作为对照组,将细菌溶液稀释10,000倍并且不施加UV。
    结果:与对照相比,用UV照射观察到5.95±0.91log的平均值±SD减少。金黄色葡萄球菌的对数减少最大[中位数,7.05(IQR,6.49-7.26)]且至少适用于M.peregrinum[中位数,4.88(IQR,4.58-5.24)].对数减少的程度与距UV-C灯的距离的平方成反比(R2=-0.12,p<0.001)。
    结论:在这项研究中,安装在天花板上的UV-C在4-m的距离内证明了至少4-log减少的测试生物的有效消毒。安装UV-C照明是改善表面消毒的重要选择。
    BACKGROUND: The mounted UV-C lamp requires no separate space or labor for transport, but data on its effectiveness for surface disinfection are scarce. We aimed to evaluate the performance of ceiling-mounted UV-C lamps.
    METHODS: This study was conducted in an empty room with UV-C lamps in the biocontainment unit of a tertiary care hospital in South Korea. UV-C lamps were installed on the ceiling of the patient room, anteroom, and toilet. Each pathogen (Staphylococcus aureus, Escherichia coli, Candida krusei, Bacillus cereus, and Mycobacterium peregrinum) was inoculated on blood agar plates and placed in 20 selected places from the UV-C lamp, and irradiation was applied for 15 min. As a control group, the bacterial solution was diluted 10,000 times and UV was not applied.
    RESULTS: A mean ± SD of 5.95 ± 0.91 log reduction was observed with UV irradiation compared with the control. The log reduction was greatest for S. aureus [median, 7.05 (IQR, 6.49-7.26)] and least for M. peregrinum [median, 4.88 (IQR, 4.58-5.24)]. The degree of log reduction was inversely proportional to the square of the distance from the UV-C lamp (R2 = -0.12, p < 0.001).
    CONCLUSIONS: In this study, ceiling-mounted UV-C demonstrated effective disinfection of at least 4-log reduction of the test organisms within a 4-m distance. Mounted UV-C lighting is a considerable option for improving surface disinfection.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    目的:消毒剂的消毒效果因具体情况而异。本研究旨在探讨商业过氧化氢的消毒效率,二氧化氯,和氯消毒剂在真实的现场表面,并为精确消毒提供数据。
    方法:采用模拟现场消毒和现场消毒方法定量评价过氧化氢的消毒效率,二氧化氯,和二氯异氰尿酸钠.log10生物指标的减少,大肠杆菌(ATCC8099)和金黄色葡萄球菌(ATCC6538),已计算。接下来,确定了食品生产和加工车间和生物安全实验室表面上天然细菌的减少。
    结果:评估的3种商业消毒剂对大肠杆菌和金黄色葡萄球菌有效,减少超过3.00log10菌落形成单位/毫升测试15分钟与3.5%过氧化氢的暴露时间,100毫克/升二氧化氯,和250mg/L二氯异氰尿酸钠。使用10.5%的过氧化氢,暴露时间为30分钟,食品生产和加工车间的自然负荷降低了90%以上。在生物安全2级实验室中,通过在60分钟的暴露时间下500mg/L的二氧化氯和在60分钟的暴露时间下450mg/L的二氯异氰尿酸钠实现相同的消毒水平。
    结论:本研究为食品工业和生物安全实验室的表面精确消毒提供了参考。
    OBJECTIVE: The disinfection efficiency of disinfectants differs in specific conditions. This study aimed to investigate the disinfection efficiency of commercial hydrogen peroxide, chlorine dioxide, and chlorine disinfectant on real field surfaces and provide data for precise disinfection.
    METHODS: Simulated field disinfection and field disinfection methods were conducted to quantitatively evaluate the disinfection efficiency of hydrogen peroxide, chlorine dioxide, and sodium dichloroisocyanurate. The log10 reduction of biological indicators, Escherichia coli (ATCC 8099) and Staphylococcus aureus (ATCC 6538), was calculated. Next, the reduction in natural bacteria on the surfaces of a food production and processing workshop and a biosafety laboratory was determined.
    RESULTS: The 3 commercial disinfectants evaluated were effective against E coli and S aureus, with a reduction of more than 3.00 log10 colony-forming units/mL tested for an exposure time of 15 minutes with 3.5% hydrogen peroxide, 100 mg/L chlorine dioxide, and 250 mg/L sodium dichloroisocyanurate. The natural load in the food production and processing workshop decreased by more than 90% using 10.5% hydrogen peroxide with an exposure time of 30 minutes. The same disinfection level in the biosafety level 2 laboratory was achieved by 500 mg/L chlorine dioxide at an exposure time of 60 minutes and 450 mg/L sodium dichloroisocyanurate at 60 minutes.
    CONCLUSIONS: This study provides a reference for precise disinfection of surfaces in the food industry and biosafety laboratories.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    BACKGROUND: Central line-associated bloodstream infections (CLABSIs) are among the most epidemiologically relevant health care-associated infections. The aseptic non-touch technique (ANTT) is a standardized practice used to prevent CLABSIs. In a pediatric hospital, the overall CLABSI rate was 1.92/1000 catheter days (CD). However, in one unit, the rate was 5.7/1000 CD.
    METHODS: Nurses were trained in ANTT. For the implementation, plan-do-study-act (PDSA) cycles were completed. Adherence monitoring of the ANTT and epidemiological surveillance were performed.
    RESULTS: ANTT adherence of 95% was achieved after 6 PDSA cycles. Hand hygiene and general cleaning reached 100% adherence. Port disinfection and material collection had the lowest adherence rates, with 76.2% and 84.7%, respectively. The CLABSI rate decreased from 5.7 to 1.26/1000 CD.
    CONCLUSIONS: The implementation of ANTT helped reduce the CLABSI rate. Training and continuous monitoring are key to maintaining ANTT adherence.
    UNASSIGNED: Las infecciones relacionadas con catéteres venosos centrales son unas de las infecciones asociadas a la atención de salud con mayor relevancia epidemiológica. La técnica aséptica «no tocar» es una práctica estandarizada que se utiliza para prevenir estas infecciones. En un hospital pediátrico, la tasa de infecciones relacionadas con catéteres venosos centrales fue de 1.92/1000 días de catéter. Sin embargo, en una de las unidades la tasa fue de 5.7/1000 días de catéter.
    UNASSIGNED: Se capacitaron enfermeras en la técnica aséptica «no tocar». Para la implementación se cumplieron ciclos de planificar-hacer-estudiar-actuar (PHEA). Se realizaron seguimiento de la adherencia a la técnica y vigilancia epidemiológica.
    RESULTS: Se logró una adherencia a la técnica aséptica «no tocar» del 95% después de seis ciclos. La higiene de manos y la limpieza general alcanzaron un 100% de cumplimiento. La desinfección de los puertos y la recolección de material alcanzaron la menor adherencia, con un 76.2% y un 84.7%, respectivamente. La tasa de infecciones relacionadas con catéteres venosos centrales disminuyó de 5.7 a 1.26 por 1000 días de catéter.
    CONCLUSIONS: La implementación de la técnica aséptica «no tocar» ayudó en la reducción de infecciones relacionadas con catéteres venosos centrales. La capacitación y el seguimiento continuo son clave para mantener el cumplimiento de la técnica.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    目的:评估义齿清洁剂对白色念珠菌粘附的功效及其对表面的影响,光学,和常规树脂的机械性能,碾磨,和3D打印义齿基托。
    方法:共制备了240个树脂样品,120用于测试白色念珠菌的附着力,光学稳定性(ΔE00),粗糙度(Ra),亲水性(°),表面自由能(Owens-Wendt)和120个用于测试白色念珠菌粘附的样品,表面显微硬度(努普),三点试验中的弯曲强度和弹性模量,其中将它们分为3组假牙树脂(n=40)和5组假牙清洁剂(n=8)。数据通过双向方差分析和Tukey检验进行多重比较(α=0.05)。
    结果:使用碱性溶液和稀酸成分的义齿清洁剂在减少白色念珠菌方面表现出最大的效果(P<0.001),然而1%NaOCl显著影响树脂的性能(P<0.05)。3D打印义齿显示,所有清洁剂的表面显微硬度均显着降低(P<0.05)。
    结论:李斯特林在减少白色念珠菌方面表现出优异的疗效,对义齿性能的影响最小,而1%NaOCl对性能有显著的负面影响。对于所有义齿清洁剂,3D打印树脂的机械性能明显低于其他树脂。
    结论:正在出售义齿基托材料以适应CAD/CAM系统,增加使用该系统制造的假牙的用户数量。尽管如此,关于微生物的粘附能力和光学,假牙的表面和机械性能,因此需要进一步调查。
    OBJECTIVE: Evaluate the efficacy of denture cleaners on the adhesion of Candida albicans and their effects on the surface, optical, and mechanical properties of resins for conventional, milled, and 3D-printed denture bases.
    METHODS: A total of 240 resin samples were made, 120 for testing Candida albicans adhesion, optical stabilities (ΔE00), roughness (Ra), hydrophilicity (°), surface free energy (Owens-Wendt) and 120 samples for testing Candida albicans adhesion, surface microhardness (Knoop), flexural strength and modulus of elasticity in a three-point test, in which they were divided into 3 groups of denture resin (n = 40) and subdivided into 5 cleaners of dentures (n = 8). Data were evaluated by two-way ANOVA and Tukey\'s test for multiple comparisons (α = 0.05).
    RESULTS: Denture cleaners with an alkaline solution and dilute acid composition were those that showed the greatest effectiveness in reducing Candida albicans (P < 0.001), however 1% NaOCl significantly affected the properties of the resins (P < 0.05). Denture 3D-printed showed that the surface microhardness was significantly lower for all cleansers (P < 0.05).
    CONCLUSIONS: Listerine demonstrated superior efficacy in reducing Candida albicans with minimal effect on denture properties, whereas 1% NaOCl had a significant negative impact on the properties. The mechanical properties were significantly lower in 3D-printed resin than in other resins for all denture cleansers.
    CONCLUSIONS: Denture base materials are being sold to adapt to the CAD/CAM system, increasing the number of users of dentures manufactured with this system. Despite this, there is little investigation into denture cleaners regarding the adhesion capacity of microorganisms and the optical, surface and mechanical properties of dentures, thus requiring further investigation.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    目的:UV-C波段的紫外线被称为杀菌辐射,广泛用于设备的灭菌和无菌环境的创建。该研究的目的是评估通过UV-C辐射消毒装置灭活沉积在具有各种纹理的表面上的微生物的有效性。
    方法:五种微生物(3种细菌,病毒,和真菌)沉积在金属上,塑料,用低压汞灯和紫外发光二极管(LED)发出的UV-C光照射具有光滑和粗糙纹理的玻璃表面,距离0.5米,1米,和1.5m以检查其暴露20分钟后的生存能力。
    结论:两种测试的UV-C源都能有效灭活微生物;然而,LED发射器在这方面比汞灯更有效。微生物的存活率取决于UV-C剂量,条件是与UV-C源的距离最高为0.5m,最低为1.5m。对于测试的微生物,UV-C照射后的最高存活率通常在玻璃和塑料表面上可见。在材料类型(制造技术设备的元件并可能受到特定活动污染)对于保持适当的卫生水平和避免不必要的和不受控制的微生物污染的所有环境中,都应考虑这一观察结果。
    OBJECTIVE: Ultraviolet light in the UV-C band is known as germicidal radiation and was widely used for both sterilization of the equipment and creation of a sterile environment. The aim of the study is to assess the effectiveness of inactivation of microorganisms deposited on surfaces with various textures by UV-C radiation disinfection devices.
    METHODS: Five microorganisms (3 bacteria, virus, and fungus) deposited on metal, plastic, and glass surfaces with smooth and rough textures were irradiated with UV-C light emitted by low-pressure mercury lamp and ultraviolet emitting diodes (LEDs), from a distance of 0.5 m, 1 m, and 1.5 m to check their survivability after 20-minute exposure.
    CONCLUSIONS: Both tested UV-C sources were effective in inactivation of microorganisms; however, LED emitter was more efficient in this respect than the mercury lamp. The survival rate of microorganisms depended on the UV-C dose, conditioned by the distance from UV-C source being the highest at 0.5 m and the lowest at 1.5 m. For the tested microorganisms, the highest survival rate after UV-C irradiation was usually visible on glass and plastic surfaces. This observation should be considered in all environments where the type of material (from which the elements of technical equipment are manufactured and may be contaminated by specific activities) is important for maintaining the proper level of hygiene and avoiding the unwanted and uncontrolled spread of microbiological pollution.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    2022年,匈牙利成人重症监护病房(ICU)爆发了由粘质沙雷氏菌引起的严重血流感染。八个案例,其中5人死亡,被检测到。最初的控制措施无法阻止疫情爆发。我们进行了匹配的病例对照研究。在单变量分析中,这些病例更有可能位于ICU的一个水槽周围,并且比对照组有更多的医疗程序和药物,然而,多变量分析尚无定论.通过全基因组测序,从病例的血培养物中分离出的细菌与ICU环境密切相关,并且对ICU中使用的季铵化合物表面消毒剂具有抗性或耐受性。因此,尽管有定期的清洁和消毒,但美国粘质抗生素还是能够在环境中生存。医院用另一种消毒剂代替了消毒剂,加强了清洁协议,并加强了医护人员的手部卫生依从性。一起,这些控制措施已被证明有效地防止了新病例。我们的研究结果突出了多学科暴发调查的重要性,包括环境采样,分子分型和消毒剂抗性测试。
    In 2022, an outbreak with severe bloodstream infections caused by Serratia marcescens occurred in an adult intensive care unit (ICU) in Hungary. Eight cases, five of whom died, were detected. Initial control measures could not stop the outbreak. We conducted a matched case-control study. In univariable analysis, the cases were more likely to be located around one sink in the ICU and had more medical procedures and medications than the controls, however, the multivariable analysis was not conclusive. Isolates from blood cultures of the cases and the ICU environment were closely related by whole genome sequencing and resistant or tolerant against the quaternary ammonium compound surface disinfectant used in the ICU. Thus, S. marcescens was able to survive in the environment despite regular cleaning and disinfection. The hospital replaced the disinfectant with another one, tightened the cleaning protocol and strengthened hand hygiene compliance among the healthcare workers. Together, these control measures have proved effective to prevent new cases. Our results highlight the importance of multidisciplinary outbreak investigations, including environmental sampling, molecular typing and testing for disinfectant resistance.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    水传播病原体总是对公众健康构成相当大的威胁。群体感应(QS)系统有助于协调细菌的生长和代谢。然而,细菌通过群体感应对各种消毒技术的反应和调节机制尚不清楚。这项研究研究了氯化和臭氧化对铜绿假单胞菌QS信号缺乏突变体的生物膜和浮游细胞的灭活作用。细胞计数和活力评估表明,氯和臭氧的联合消毒对于在暴露后10分钟内灭活浮游铜绿假单胞菌最有效。此外,微流控芯片培养表明,喹诺酮类药物信号的分泌增强了生物膜的消毒抗性。消毒暴露显着改变了野生型菌株和QS信号缺乏突变体的基因表达。此外,QS系统触发了多层基因表达程序,作为对消毒剂暴露的响应性保护,包括氧化应激,核糖体合成,和细菌的营养吸收。这些见解扩大了我们对细菌QS对消毒的反应的理解,有希望的潜在策略,以实现有效的消毒过程。
    Waterborne pathogens invariably present considerable threats to public health. The quorum sensing (QS) system is instrumental in coordinating bacterial growth and metabolisms. However, the responses and regulatory mechanisms of bacteria to various disinfection technologies through quorum sensing are still unclear. This study examines the inactivation effect of chlorination and ozonation on biofilms and planktonic cells of QS signaling-deficient mutants of Pseudomonas aeruginosa. Cell counting and viability assessment revealed that the combined disinfection of chlorine and ozone was the most effective for inactivating planktonic P. aeruginosa within 10 min of exposure. Additionally, microfluidic chip culture demonstrated that the secretion of quinolone signals escalated biofilms\' disinfection resistance. Disinfection exposure significantly altered the gene expression of wild-type strains and QS signaling-deficient mutants. Moreover, the QS system triggered multilayered gene expression programs as a responsive protection to disinfectant exposure, including oxidative stress, ribosome synthesis, and the nutrient absorption of bacteria. These insights broaden our understanding of bacterial QS in response to disinfection, promising potential strategies toward efficient disinfection processes.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

公众号