Noninvasive blood pressure

无创血压
  • 文章类型: Journal Article
    目的:俯卧位作为急性呼吸窘迫综合征和脊柱和脑部手术的治疗方法,最近获得了新的重要性。我们的研究旨在进行误差网格分析,以检查俯卧位的动脉血压(ABP)和无创血压(NIBP)之间的临床差异,并调查影响这些差异的危险因素。
    方法:对1389对100例连续俯卧位病例进行了误差网格分析。此分析将两种方法之间的差异分为五个临床相关区域,从“无风险”到“危险风险”。此外,进行多变量有序logistic回归分析以评估平均血压(MBP)的危险区域之间的关系,通过误差网格分析和感兴趣的协变量进行分类。
    结果:误差网格分析表明,收缩压危险区域A-E的测量对比例为96.8%,3.2%,0.1%,0%,0%,分别。相比之下,MBP比例为74.0%,25.1%,0.9%,0.1%,和0%。多变量有序logistic回归分析显示,手臂的位置(头部旁边)是一个重要因素(调整后的比值比:4.35,95%CI:2.38-8.33,P<0.001)。
    结论:误差网格分析显示,在俯卧位手术中,ABP和NIBP对于MBP存在临床上不可接受的差异。手臂靠近头部的位置与两种MBP测量方法之间的临床差异增加有关。
    OBJECTIVE: Prone position has recently gained renewed importance as a treatment for acute respiratory distress syndrome and spine and brain surgeries. Our study aimed to perform an error grid analysis to examine the clinical discrepancies between arterial blood pressure (ABP) and non-invasive blood pressure (NIBP) in the prone position and to investigate the risk factors influencing these differences.
    METHODS: Error grid analysis was performed retrospectively on 1389 pairs of 100 consecutive prone positioning cases. This analysis classifies the difference between the two methods into five clinically relevant zones, from \"no risk\" to \"dangerous risk\". Additionally, multivariable ordinal logistic regression analysis was conducted to evaluate the relationship between the risk zones of mean blood pressure (MBP), as classified by error grid analysis and the covariate of interest.
    RESULTS: Error grid analysis showed that the proportions of measurement pairs in risk zones A-E for systolic blood pressure were 96.8%, 3.2%, 0.1%, 0%, and 0%, respectively. In contrast, the MBP proportions were 74.0%, 25.1%, 0.9%, 0.1%, and 0%. Multivariable ordinal logistic regression analysis revealed that the position of arms (next to the head) was a significant factor (adjusted odds ratio: 4.35, 95% CI: 2.38-8.33, P < 0.001).
    CONCLUSIONS: Error grid analysis revealed a clinically unacceptable discrepancy between ABP and NIBP for MBP during prone positioning surgery. The position of the arms next to the head was associated with increased clinical discrepancy between the two MBP measurement methods.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    目的:使用多参数监测仪评估有创血压(IBP)和舌部示波血压(OBPton)之间的一致性。
    方法:非致盲,prospective,实验研究。
    方法:共12头雌性大白杂交猪。
    方法:招募进行需要动脉插管放置的实验程序的猪。将宽度最接近舌头周长40%的血压袖带置于舌系带的前端。收缩压,同时测量平均和舒张期IBP和OBPton,间隔5分钟.使用Bland-Altman分析检查配对测量之间的一致性。平均偏差,精度(平均偏差的标准偏差),95%的协议限制,计算了10和20mmHgIBP内的相关系数和测量值百分比.
    结果:收缩压记录的配对测量总数为124、126和124,平均和舒张压,分别。平均偏差,收缩期OBPton的精度和95%一致界限为11.5、11.5(-11.1至34.2),平均OBPton5.6,5.7(-5.7至16.8)和舒张OBPton7.6,10.1(-12.1至27.4)mmHg。仅平均OBPton的相关系数大于0.9。仅对于平均和舒张OBP,超过50%的测量值在IBP的10mmHg内,并且80%的测量值在IBP的20mmHg内。
    结论:作为示波血压测量的袖带部位的舌头是测量平均动脉的有用部位,但不是收缩压或舒张压麻醉的大型白色杂交猪。该技术符合美国兽医内科学院的标准,用于测量平均动脉压,而不是收缩压或舒张压。
    OBJECTIVE: To evaluate the agreement between measurements of invasive blood pressure (IBP) and oscillometric blood pressure from the tongue (OBPton) using a multiparameter monitor.
    METHODS: Unblinded, prospective, experimental study.
    METHODS: A total of 12 female Large White crossbreed pigs.
    METHODS: Pigs undergoing experimental procedures that required arterial cannula placement were recruited. A blood pressure cuff with the closest width to 40% of the circumference of the tongue was placed rostral to the lingual frenulum. Systolic, mean and diastolic IBP and OBPton were measured simultaneously at 5 minute intervals. Agreement between paired measurements was examined using Bland-Altman analysis. Mean bias, precision (standard deviation of mean bias), 95% limits of agreement, correlation coefficients and percentage of measurements within 10 and 20 mmHg of IBP were calculated.
    RESULTS: The total numbers of paired measurements recorded were 124, 126 and 124 for systolic, mean and diastolic blood pressures, respectively. The mean bias, precision and 95% limits of agreement for systolic OBPton were 11.5, 11.5 (-11.1 to 34.2), for mean OBPton 5.6, 5.7 (-5.7 to 16.8) and for diastolic OBPton 7.6, 10.1 (-12.1 to 27.4) mmHg. Correlation coefficients were greater than 0.9 for mean OBPton only. More than 50% of measurements were within 10 mmHg of IBP and 80% of measurements were within 20 mmHg of IBP for mean and diastolic OBPton only.
    CONCLUSIONS: The tongue as a cuff site for oscillometric blood pressure measurement is a useful site for measuring mean arterial, but not systolic or diastolic blood pressure in anaesthetized Large White crossbreed pigs. This technique fulfils the American College of Veterinary Internal Medicine criteria for measuring mean arterial pressure but not systolic or diastolic arterial pressure.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    背景:这项研究旨在评估踝关节血压测量与侧卧位有创血压相关的准确性。
    方法:这项前瞻性观察研究纳入了在侧卧位全身麻醉下择期非心脏手术的成年患者。使用GECarescapeB650监测仪在侧卧位记录成对的桡动脉有创和踝关节无创血压读数。主要结果是踝关节平均动脉压(MAP)检测低血压(MAP<70mmHg)的能力,使用受试者工作特征曲线下面积(AUC)分析。次要结果是踝关节收缩压(SBP)检测高血压(SBP>140mmHg)的能力以及偏倚(有创测量-无创测量),并使用Bland-Altman分析在两种方法之间达成一致。
    结果:我们分析了来自30例患者的415个配对读数。踝关节MAP检测低血压的AUC(95%置信区间[CI])为0.88(0.83-0.93)。踝关节MAP≤86mmHg的阴性和阳性预测值(95%CI)为99(97-100)%和21(15-29)%,分别,用于检测低血压。踝关节SBP检测高血压的AUC(95%CI)为0.83(0.79-0.86),阴性和阳性预测值(95%CI)为95(92-97)%和36(26-46)%,分别,截止值>144mmHg。两种方法之间的平均偏差为SBP的-12±17,3±12和-1±11mmHg,舒张压,还有MAP,分别。
    结论:在侧卧位全身麻醉的患者中,踝关节血压测量值与相应的侵入性测量值不可互换.然而,踝关节MAP>86mmHg可以排除低血压,准确率为99%,踝关节SBP<144mmHg可以排除高血压,准确率为95%。
    BACKGROUND: This study aimed to evaluate the accuracy of ankle blood pressure measurements in relation to invasive blood pressure in the lateral position.
    METHODS: This prospective observational study included adult patients scheduled for elective non-cardiac surgery under general anesthesia in the lateral position. Paired radial artery invasive and ankle noninvasive blood pressure readings were recorded in the lateral position using GE Carescape B650 monitor. The primary outcome was the ability of ankle mean arterial pressure (MAP) to detect hypotension (MAP < 70 mmHg) using area under the receiver operating characteristic curve (AUC) analysis. The secondary outcomes were the ability of ankle systolic blood pressure (SBP) to detect hypertension (SBP > 140 mmHg) as well as bias (invasive measurement - noninvasive measurement), and agreement between the two methods using the Bland-Altman analysis.
    RESULTS: We analyzed 415 paired readings from 30 patients. The AUC (95% confidence interval [CI]) of ankle MAP for detecting hypotension was 0.88 (0.83-0.93). An ankle MAP of ≤ 86 mmHg had negative and positive predictive values (95% CI) of 99 (97-100)% and 21 (15-29)%, respectively, for detecting hypotension. The AUC (95% CI) of ankle SBP to detect hypertension was 0.83 (0.79-0.86) with negative and positive predictive values (95% CI) of 95 (92-97)% and 36 (26-46)%, respectively, at a cutoff value of > 144 mmHg. The mean bias between the two methods was - 12 ± 17, 3 ± 12, and - 1 ± 11 mmHg for the SBP, diastolic blood pressure, and MAP, respectively.
    CONCLUSIONS: In patients under general anesthesia in the lateral position, ankle blood pressure measurements are not interchangeable with the corresponding invasive measurements. However, an ankle MAP > 86 mmHg can exclude hypotension with 99% accuracy, and an ankle SBP < 144 mmHg can exclude hypertension with 95% accuracy.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    透析性低血压(IDH)是血液透析(HD)的严重并发症,对发病率和死亡率有重大影响。在这项研究中,我们使用可穿戴设备连续监测血液动力学生命指标,以检测HD期间的血液动力学变化,并尝试识别IDH.终末期肾病患者在开始治疗前15分钟持续监测,直到治疗结束后15分钟。测量心率(HR),无创无袖带收缩压和舒张压(SBP和DBP),每搏输出量(SV),心输出量(CO),和全身血管阻力(SVR)。对数据进行回顾性分析,包括比较可穿戴设备(每5s连续记录一次)和基于袖带的设备测得的BP。最终分析共包括98次透析,在22个疗程中发现了IDH(22.5%)。SBP和DBP在可穿戴设备和基于袖带的测量之间高度相关(r>0.62,p<0.001)。在持续监测的基础上,在HD治疗期间,IDH患者的SBP和DBP降低更早,更显著.此外,几乎所有的高级生命体征在组间都不同。应进行进一步研究,以充分了解无创高级连续监测在预测和预防IDH事件中的潜力。
    Intradialytic hypotension (IDH) is a severe complication of hemodialysis (HD) with a significant impact on morbidity and mortality. In this study, we used a wearable device for the continuous monitoring of hemodynamic vitals to detect hemodynamic changes during HD and attempted to identify IDH. End-stage kidney disease patients were continuously monitored 15 min before starting the session and until 15 min after completion of the session, measuring heart rate (HR), noninvasive cuffless systolic and diastolic blood pressure (SBP and DBP), stroke volume (SV), cardiac output (CO), and systemic vascular resistance (SVR). Data were analyzed retrospectively and included comparing BP measured by the wearable devices (recorded continuously every 5 s) and the cuff-based devices. A total of 98 dialysis sessions were included in the final analysis, and IDH was identified in 22 sessions (22.5%). Both SBP and DBP were highly correlated (r > 0.62, p < 0.001 for all) between the wearable device and the cuff-based measurements. Based on the continuous monitoring, patients with IDH had earlier and more profound reductions in SBP and DBP during the HD treatment. In addition, nearly all of the advanced vitals differed between groups. Further studies should be conducted in order to fully understand the potential of noninvasive advanced continuous monitoring in the prediction and prevention of IDH events.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    背景:无创连续血压监测具有改善医院环境中患者治疗的潜力。这样的非侵入性装置可以在治疗过程中更早地应用,以使护士和临床医生能够更快地对患者恶化作出反应,并具有消除与侵入性监测相关的风险的附加益处。然而,新兴技术必须能够为医疗决策再现当前的临床措施。
    方法:本研究旨在确定护士实施无创连续血压监测设备的可用性和意愿。次要目标直接比较收缩压,舒张压,和设备记录的平均动脉压值(VitalStream;CareTakerMedicalLLC,夏洛茨维尔,VA)在急诊科和重症监护病房设置中记录了“金标准”臂袖和动脉线措施。
    结果:VitalStream同样被急诊科和重症监护机构的护士接受,但最终得到了急诊护士的更大晋升。尽管测量方法之间有一些统计上的相似性,所有直接比较均不符合医疗器械促进协会2008和医疗器械促进协会/欧洲高血压学会/国际标准化组织2019年VitalStream和"金标准"临床测量之间可接受血压测量差异的共识声明标准.在所有情况下,Bland-Altman偏倚的标准偏差超过8mmHg,其中小于85%的成对差异落在\"金标准的10mmHg以内.\"
    结论:汇总,被测试的设备需要额外的后处理,以便在创伤或紧急护理中做出医疗决策。
    BACKGROUND: Noninvasive continuous blood pressure monitoring has the potential to improve patient treatment in the hospital setting. Such noninvasive devices can be applied earlier in the treatment process to empower nurses and clinicians to react more quickly to patient deterioration with the added benefit of eliminating the risks associated with invasive monitoring. However, emerging technologies must be capable of reproducing current clinical measures for medical decision making.
    METHODS: This study aimed to determine the usability and willingness of nurses to implement a noninvasive continuous blood pressure monitoring device. The secondary aim directly compared the systolic blood pressure, diastolic blood pressure, and mean arterial pressure values recorded by the device (VitalStream; CareTaker Medical LLC, Charlottesville, VA) with the \"gold standard\" brachial cuff and arterial line measures recorded in the emergency department and intensive care unit settings.
    RESULTS: VitalStream was similarly received by nurses in the emergency department and intensive care setting, but ultimately had greater promotion from emergency nurses. Despite some statistical similarity between measurement methodologies, all direct comparisons were found to not meet the Association for the Advancement of Medical Instrumentation 2008 and Association for the Advancement of Medical Instrumentation / European Society of Hypertension / International Organization for Standardization 2019 consensus statement criteria for acceptable blood pressure measure differences between the VitalStream and \"gold standard\" clinical measures. In all instances, the standard deviation of the Bland-Altman bias exceeded 8 mm Hg with less than 85% of paired differences falling within 10 mm Hg of the \"gold standard.\"
    CONCLUSIONS: Taken together, the tested device requires additional postprocessing for medical decision making in trauma or emergent care.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    目的:比较多普勒超声(DOP)和脉搏血氧仪体积描记术(POP)测量麻醉犬收缩期动脉压(SAP)到有创血压(IBP)的准确性。
    方法:40只客户拥有的健康狗>10公斤。
    方法:对狗进行背侧卧位的手术麻醉。从足背动脉测量有创血压。将DOP和POP装置探头放置在带有流量阻塞袖带的中位尾动脉上,以进行无创血压测量。通过DOP测量收缩压,脉搏血氧仪丢失体积描记器(POPL),将脉搏血氧定量仪(POPR)的返回与IBP测量的SAP进行比较。使用线性混合模型来确定相关性。进行Bland-Altman分析以确定偏倚,SD,和协议的限制。在不同密集状态下,将DOP和POP的准确性与IBP进行了比较。
    结果:DOP的条件R2值,POPL,POPR和IBP分别为0.92、0.85和0.87(均P<.001)。DOP的偏见,POPL,与IBP相比,POPR为7.6±13.1,3.9±14.4和8.6±15.2mmHg(偏差±SD),分别。协议限制(较低,上)是(-18.1,+33.3),(-24.3,+32.1),DOP为(-21.2,+38.4)mmHg,POPL,和POPR,分别。DOP和POP在低血压期间高估了SAP(SAP<90mmHg),DOP的幅度较小。
    结论:从尾中动脉测量的DOP对于背卧的SAP测量可能是可接受的,健康的麻醉狗>10公斤。POP被确定为不可接受的方法。
    OBJECTIVE: To compare the accuracy of doppler ultrasound (DOP) and pulse oximeter plethysmography (POP) in the measurement of systolic arterial pressure (SAP) to invasive blood pressure (IBP) in anesthetized dogs.
    METHODS: 40 client-owned healthy dogs > 10 kg.
    METHODS: Dogs were anesthetized for surgical procedures in dorsal recumbency. Invasive blood pressure was measured from a dorsal pedal artery. DOP and POP device probes were placed over the median caudal artery with a flow-occluding cuff for noninvasive blood pressure measurement. Systolic arterial pressure measured by DOP, loss of pulse oximeter plethysmograph (POPL), and return of pulse oximeter plethysmograph (POPR) were compared to SAP measured by IBP. A linear mixed model was used to determine correlation. Bland-Altman analyses were performed to determine bias, SD, and limits of agreement. The accuracy of DOP and POP was compared to IBP across different tensive states.
    RESULTS: Conditional R2 values for DOP, POPL, and POPR versus IBP were 0.92, 0.85, and 0.87, respectively (all P < .001). The biases for DOP, POPL, and POPR compared to IBP were +7.6 ± 13.1, +3.9 ± 14.4, and +8.6 ± 15.2 mm Hg (bias ± SD), respectively. Limits of agreement (lower, upper) were (-18.1, +33.3), (-24.3, +32.1), and (-21.2, +38.4) mm Hg for DOP, POPL, and POPR, respectively. DOP and POP overestimated SAP during hypotension (SAP < 90 mm Hg), DOP to a lesser magnitude.
    CONCLUSIONS: DOP measured from the median caudal artery may be acceptable for SAP measurement in dorsally recumbent, healthy anesthetized dogs > 10 kg. POP was determined an unacceptable method.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    背景:妊娠肥胖患者可能有无创血压(NIBP)测量受损。我们通过有创动脉监测(INV)和手臂NIBP评估了ClearSight™指套(FC)的准确性和趋势能力,在剖宫产的肥胖患者中。
    方法:参与者年龄≥18岁,妊娠≥34周,体重指数(BMI)≥40kgm-2。FC,INV,和NIBP测量值在5分钟间隔内获得。主要结果是FC测量值与参考标准INV的测量值一致,使用修改过的Bland-Altman地块.次要结果包括FC和NIBP之间的比较以及NIBP与INV之间的比较,使用四象限图报告不一致率并评估趋势能力。
    结果:23名参与者的中位数(IQR)BMI为45kgm-2(44-48)。当比较FC和INV时,平均偏差(SD,收缩压(SBP)的95%一致性界限为16mmHg(17,-17.3至49.3mmHg),对于舒张压(DBP)-0.2mmHg(10.5,-20.7至20.3),平均动脉压(MAP)5.2mmHg(11.1,-16.6至27.0mmHg)。SBP有54对(26%)不一致,DBP为41(23%),MAP为41(21.7%)。误差网格分析显示,A区(无风险区)SBP读数为92.1%。比较NIBP和INV时,SBP的平均偏倚(95%一致界限)为13.0mmHg(16.7,-19.7至29.3),对于DBP5.9mmHg(11.9,-17.4至42.0),MAP为8.2mmHg(11.9,-15.2至31.6)。SBP出现不一致(209人中有84人,40.2%),DBP(187人中的74人,39.6%),和MAP(191人中的63人,33.0%)。
    结论:FC和NIBP技术与INV没有足够的一致性。FC的趋势能力优于NIBP。临床上重要的差异可能发生在灌注依赖性胎儿的设置中。
    BACKGROUND: Pregnant patients with obesity may have compromised noninvasive blood pressure (NIBP) measurement. We assessed the accuracy and trending ability of the ClearSight™ finger cuff (FC) with invasive arterial monitoring (INV) and arm NIBP, in obese patients having cesarean delivery.
    METHODS: Participants were aged ≥18 years, ≥34 weeks gestation, and body mass index (BMI) ≥ 40 kg m-2. FC, INV, and NIBP measurements were obtained across 5-min intervals. The primary outcome was agreement of FC measurements with those of the reference standard INV, using modified Bland-Altman plots. Secondary outcomes included comparisons between FC and NIBP and NIBP versus INV, with four-quadrant plots performed to report discordance rates and evaluate trending ability.
    RESULTS: Twenty-three participants had a median (IQR) BMI of 45 kg m-2 (44-48). When comparing FC and INV the mean bias (SD, 95% limits of agreement) for systolic blood pressure (SBP) was 16 mmHg (17, -17.3 to 49.3 mmHg), for diastolic blood pressure (DBP) -0.2 mmHg (10.5, -20.7 to 20.3), and for mean arterial pressure (MAP) 5.2 mmHg (11.1, -16.6 to 27.0 mmHg). Discordance occurred in 54 (26%) pairs for SBP, 41 (23%) for DBP, and 41 (21.7%) for MAP. Error grid analysis showed 92.1% of SBP readings in Zone A (no-risk zone). When comparing NIBP and INV, the mean bias (95% limits of agreement) for SBP was 13.0 mmHg (16.7, -19.7 to 29.3), for DBP 5.9 mmHg (11.9, -17.4 to 42.0), and for MAP 8.2 mmHg (11.9, -15.2 to 31.6). Discordance occurred in SBP (84 of 209, 40.2%), DBP (74 of 187, 39.6%), and MAP (63 of 191, 33.0%).
    CONCLUSIONS: The FC and NIBP techniques were not adequately in agreement with INV. Trending capability was better for FC than NIBP. Clinically important differences may occur in the setting of the perfusion-dependent fetus.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    目的:评估从舌头和胸肢测量的无创动脉血压(NIBP)与有创血压(IBP)之间的一致性。并比较麻醉马的舌头和胸肢的NIBP。
    方法:前瞻性临床研究。
    方法:一组八匹客户拥有的健康马匹进行了预定程序的麻醉,美国麻醉医师协会I-II分类,体重(平均值±标准偏差)498±91kg,年龄7.8±6.75岁。
    方法:用静脉内(IV)罗米替定(0.04-0.08mg/kg-1)和美沙酮(0.1mg/kg-1)对动物进行预用药。用氯胺酮(2.5mgkg-1)和咪达唑仑(0.05mgkg-1)诱导全身麻醉,并用异氟烷维持。将面动脉插入导管以进行IBP测量。收缩压(SAP),从NIBP舌记录平均(MAP)和舒张压(DAP)动脉压,手术期间每20分钟一次NIBP肢体和IBP。根据美国兽医内科学院标准和美国医疗器械协会标准,使用Bland-Altman方法评估了NIBP和IBP之间的协议。
    结果:从舌头测量的IBP和NIBP之间的平均偏差和精度均符合所有压力范围(<10mmHg和<15mmHg,分别)。来自舌头和胸肢的NIBP测量结果倾向于低估IBP测量结果。在低血压期间,MAPtongue和DAPtongue高估了IBP,但是精确度和准确度都符合标准。NIBPlimb的总体准确性和精密度均低于NIBPtongue。与IBP的差异<10mmHg的NIBP舌的百分比高于SAP的NIBP肢记录的百分比(46%对25%),MAP(77%对28%)和DAP(79%对19%)。
    结论:与胸肢相比,舌头是评估动脉血压的临床合适选择,并且可以可靠地检测健康麻醉马的低血压。
    OBJECTIVE: To evaluate the agreement between noninvasive arterial blood pressure (NIBP) measured from the tongue and thoracic limb with invasive blood pressure (IBP), and to compare NIBP measured from the tongue and thoracic limb in anaesthetized horses.
    METHODS: Prospective clinical study.
    METHODS: A group of eight client-owned healthy horses anaesthetized for scheduled procedures, American Society of Anesthesiologists classification I-II, weighing (mean ± standard deviation) 498 ± 91 kg and aged 7.8 ± 6.75 years.
    METHODS: Animals were premedicated with intravenous (IV) romifidine (0.04-0.08 mg kg-1) and methadone (0.1 mg kg-1). General anaesthesia was induced IV with ketamine (2.5 mg kg-1) and midazolam (0.05 mg kg-1) and maintained with isoflurane. The facial artery was catheterized for IBP measurements. Systolic (SAP), mean (MAP) and diastolic (DAP) arterial pressures were recorded from the NIBPtongue, NIBPlimb and IBP every 20 minutes during the procedure. Agreement between NIBP and IBP was evaluated based on the American College of Veterinary Internal Medicine criteria and American Association for Medical Instrumentation criteria using the Bland-Altman method.
    RESULTS: The mean bias and precision between IBP and NIBP measured from the tongue met the standards for all pressure ranges (< 10 mmHg and < 15 mmHg, respectively). NIBP measurements from the tongue and thoracic limb tended to underestimate IBP measurements. During hypotension, MAPtongue and DAPtongue overestimated IBP, but both precision and accuracy met the criteria. The overall accuracy and precision of NIBPlimb was poorer than NIBPtongue. The percentage of NIBPtongue that differed from IBP by < 10 mmHg was higher than that recorded with NIBPlimb for SAP (46% versus 25%), MAP (77% versus 28%) and DAP (79% versus 19%).
    CONCLUSIONS: The tongue is a clinically suitable alternative for assessing arterial blood pressure compared with the thoracic limb and can reliably detect hypotension in healthy anaesthetized horses.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    连续,无创血压(CNIBP)监测提供了有价值的血流动力学信息,可检测心血管疾病的早期发作.安装在皮肤上的可穿戴式机电压力传感器具有出色的适应性,因此有望用于监测连续血压(BP)脉冲波形,简单的传感机制,和方便的信号采集。然而,获取高保真BP脉冲波形具有挑战性,因为它需要高度灵敏的传感器(灵敏度大于4×10-5kPa-1),这些传感器在较大的动态范围内对压力变化做出线性响应。覆盖典型的BP范围(5-25kPa)。在这里,这项工作引入了高保真,具有高灵敏度(4.82kPa-1)的基于离子的眼压传感器(ITS),良好的线性度(R2>0.995),和一个大的动态范围(高达180%的输出变化)在一个宽的工作范围(0至38千帕)。此外,ITS在40Pa时显示出低检测限,快速负载响应时间(35ms)和释放时间(35ms),以及每个释放周期超过5000个负载的稳定响应,为人机界面交互中的潜在应用铺平了道路,电子皮肤,和机器人触觉。本文进一步探讨了ITS在实时监控中的应用,通过测量肱动脉和桡动脉脉搏波形来达到血压。这项工作提供了一种高灵敏度的可穿戴压力传感器的合理设计,良好的线性度,和实时监测CNIBP的大动态范围。
    Continuous, noninvasive blood pressure (CNIBP) monitoring provides valuable hemodynamic information that renders detection of the early onset of cardiovascular diseases. Wearable mechano-electric pressure sensors that mount on the skin are promising candidates for monitoring continuous blood pressure (BP) pulse waveforms due to their excellent conformability, simple sensing mechanisms, and convenient signal acquisition. However, it is challenging to acquire high-fidelity BP pulse waveforms since it requires highly sensitive sensors (sensitivity larger than 4 × 10-5 kPa-1 ) that respond linearly with pressure change over a large dynamic range, covering the typical BP range (5-25 kPa). Herein, this work introduces a high-fidelity, iontronic-based tonometric sensor (ITS) with high sensitivity (4.82 kPa-1 ), good linearity (R2 > 0.995), and a large dynamic range (up to 180% output change) over a broad working range (0 to 38 kPa). Additionally, the ITS demonstrates a low limit of detection at 40 Pa, a fast load response time (35 ms) and release time (35 ms), as well as a stable response over 5000 load per release cycles, paving ways for potential applications in human-interface interaction, electronic skins, and robotic haptics. This work further explores the application of the ITS in monitoring real-time, beat-to-beat BP by measuring the brachial and radial pulse waveforms. This work provides a rational design of a wearable pressure sensor with high sensitivity, good linearity, and a large dynamic range for real-time CNIBP monitoring.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    目的:评估从舌头测量的示波血压(OBP)与有创血压(IBP)之间的一致性,并将从舌头测量的OBP与从骨盆肢体和尾巴测量的OBP进行比较。
    方法:前瞻性实验研究。
    方法:共8只成年比格犬,体重11.1±1.2kg。
    方法:用静脉内(IV)乙酰丙嗪(0.005mgkg-1)对动物进行预用药。用阿法沙松(3mgkg-1)IV诱导麻醉,并用异氟烷维持。将足背动脉插入导管以进行IBP测量。收缩压(SAP),从舌头同时测量舒张压(DAP)和平均动脉压(MAP),骨盆肢体和尾巴。基于侵入性SAP,高血压(>140mmHg),通过控制呼气末异氟醚浓度和/或多巴酚丁胺/多巴胺给药可诱导血压正常(90-140mmHg)和低血压(<90mmHg).使用用于小动物和人类的无创血压设备的参考标准分析了配对的IBP和OBP测量之间的一致性。
    结果:无论袖带放置如何,IBP和OBP之间的平均偏差±标准偏差符合MAP和DAP的兽医标准(≤10±15mmHg)和人类标准(<5±8mmHg)。OBP装置提供的SAP测量结果与IBP的一致性不可接受,方法之间的偏差在较高的血压下增加,无论袖口部位。在低血压期间,舌OBP显示与SAP的IBP相关的绝对差异<10mmHg的最大百分比(90%),MAP(97%),和DAP(93%),与骨盆肢体相比(60%,97%和82%,分别)和尾部OBP(54%,92%和77%,分别)。
    结论:舌头是测量麻醉比格犬OBP的临床有用部位,提供可靠的MAP和DAP估计。舌头可以代替其他袖带放置部位,并且可以是用于评估低血压的相对合适的部位。
    OBJECTIVE: To evaluate the agreement between oscillometric blood pressure (OBP) measured from the tongue and invasive blood pressure (IBP), and to compare OBPs measured from the tongue with OBPs measured from the pelvic limb and tail.
    METHODS: Prospective experimental study.
    METHODS: A total of eight adult Beagle dogs weighing 11.1 ± 1.2 kg.
    METHODS: Animals were premedicated with intravenous (IV) acepromazine (0.005 mg kg-1). Anesthesia was induced with alfaxalone (3 mg kg-1) IV and maintained with isoflurane. The dorsal pedal artery was catheterized for IBP measurements. Systolic (SAP), diastolic (DAP) and mean (MAP) arterial pressure were simultaneously measured from the tongue, pelvic limb and tail. Based on invasive SAP, hypertension (>140 mmHg), normotension (90-140 mmHg) and hypotension (<90 mmHg) were induced by controlling end-tidal isoflurane concentrations and/or dobutamine/dopamine administration. Agreement between paired IBP and OBP measurements was analyzed with reference standards for noninvasive blood pressure devices used in small animals and humans.
    RESULTS: Regardless of cuff placement, the mean bias ± standard deviation between IBP and OBP met veterinary (≤10 ± 15 mmHg) and human (<5 ± 8 mmHg) standards for MAP and DAP. SAP measurements provided by the OBP device showed unacceptable agreement with IBP, and the bias between methods increased at higher blood pressures, regardless of cuff site. During hypotension, tongue OBP showed the largest percentage of absolute difference <10 mmHg in relation to IBP for SAP (90%), MAP (97%), and DAP (93%), compared with pelvic limb (60%, 97% and 82%, respectively) and tail OBP (54%, 92% and 77%, respectively).
    CONCLUSIONS: The tongue is a clinically useful site for measuring OBP in anesthetized Beagle dogs, providing reliable estimates of MAP and DAP. The tongue could replace other cuff placement sites and may be a relatively suitable site for assessing hypotension.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

公众号