MR thermometry

MR 测温
  • 文章类型: Journal Article
    磁共振(MR)测温法用于监测MR引导的微波消融(MWA),以及消融区域的术中评估。然而,温度映射的准确性可能会受到微波(MW)发生器发出的电磁干扰的影响。这项研究评估了使用改良的MW发电机在MWA期间改善磁共振成像(MRI)的不同设置。MWA在15个凝胶体模中进行,比较三个设置:MW发生器放置在MR扫描仪室之外,使用带有射频(RF)滤波器的穿透面板和7m同轴电缆(设置1)连接到MW施加器,或通过使用5m同轴电缆的波导(设置2)。设置3在MR扫描室内使用了MW发电机,由5米同轴电缆连接。设置2和3中的同轴电缆用定制屏蔽修改以减少干扰。将消融期间的设置(主动设置)与不存在MW系统的参考设置进行比较。比较了三种配置的测温和热剂量图(CEM43模型)。评估的主要终点是信噪比(SNR),温度精度,Sørensen-Dice-Coefficient(DSC),和射频噪声频谱。设置3在消融期间显示高度显著的电磁干扰,与参考成像相比,SNR降低-60.4%±13.5%(p<0.001)。对于设置1和设置2,信噪比没有显著下降,差异为-2.9%±9.8%(p=0.6)和-1.5%±12.8%(p=0.8)。分别。活动设置1和3之间的SNR差异显著,为-51.2%±16.1%(p<0.001),活动设置2和3之间的SNR差异显著,为-59.0%±15.5%(p<0.001),但活动设置1和2之间的SNR差异不显著,为19.0%±13.7%(p=0.09)。此外,在所有设置之间的温度精度或DSC没有显著差异,范围从0.33°C±0.04°C(设置1)到0.38°C±0.06°C(设置3)(p=0.6)和从87.0%±1.6%(设置3)到88.1%±1.6%(设置2)(p=0.58),分别。在MR扫描仪室外部使用MW发电机的两种设置(1和2)都有利于减少MWA期间的电磁干扰。此外,如果在设置2和3中使用屏蔽电缆,则所有配置在温度精度和DSC方面的差异可忽略不计,表明MW发电机的位置不会显著影响MWA期间测温的准确性。
    Magnetic Resonance (MR) thermometry is used for the monitoring of MR-guided microwave ablations (MWA), and for the intraoperative evaluation of ablation regions. Nevertheless, the accuracy of temperature mapping may be compromised by electromagnetic interference emanating from the microwave (MW) generator. This study evaluated different setups for improving magnetic resonance imaging (MRI) during MWA with a modified MW generator. MWA was performed in 15 gel phantoms comparing three setups: The MW generator was placed outside the MR scanner room, either connected to the MW applicator using a penetration panel with a radiofrequency (RF) filter and a 7 m coaxial cable (Setup 1), or through a waveguide using a 5 m coaxial cable (Setup 2). Setup 3 employed the MW generator within the MR scan room, connected by a 5 m coaxial cable. The coaxial cables in setups 2 and 3 were modified with custom shielding to reduce interference. The setups during ablation (active setup) were compared to a reference setup without the presence of the MW system. Thermometry and thermal dose maps (CEM43 model) were compared for the three configurations. Primary endpoints for assessment were signal-to-noise ratio (SNR), temperature precision, Sørensen-Dice-Coefficient (DSC), and RF-noise spectra. Setup 3 showed highly significant electromagnetic interference during ablation with a SNR decrease by -60.4%±13.5% (p<0.001) compared to reference imaging. For setup 1 and setup 2 no significant decrease in SNR was measured with differences of -2.9%±9.8% (p=0.6) and -1.5%±12.8% (p=0.8), respectively. SNR differences were significant between active setups 1 and 3 with -51.2%±16.1% (p<0.001) and between active setups 2 and 3 with -59.0%±15.5% (p<0.001) but not significant between active setups 1 and 2 with 19.0%±13.7% (p=0.09). Furthermore, no significant differences were seen in temperature precision or DSCs between all setups, ranging from 0.33 °C ± 0.04 °C (Setup 1) to 0.38 °C ± 0.06 °C (Setup 3) (p=0.6) and from 87.0%±1.6% (Setup 3) to 88.1%±1.6% (Setup 2) (p=0.58), respectively. Both setups (1 and 2) with the MW generator outside the MR scanner room were beneficial to reduce electromagnetic interference during MWA. Moreover, provided that a shielded cable is utilized in setups 2 and 3, all configurations displayed negligible differences in temperature precision and DSCs, indicating that the location of the MW generator does not significantly impact the accuracy of thermometry during MWA.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    目的:MRI引导的激光间质热治疗(MRgLITT)作为难治性癫痫的消融性立体定向手术最近引起了人们的兴趣,运动障碍,和脑肿瘤。传统上,LITT系统由激光发生器和冷却激光照射器组成,它是一个由护套包围的光纤芯,冷却流体通过护套被泵送。然而,这种占用空间会使系统笨重且不可移动,限制最大瞄准深度,并增加崩溃的机会。在这里,作者在猪模型中对非冷却MRgLITT系统进行了临床前评估.
    方法:使用三特斯拉MRI指导非冷却激光施加器在猪大脑中的体内放置。该研究由存活臂和末端臂组成。在4-7W的功率下激活激光器≤180秒。使用生存臂中的MR测温软件ThermoGuide(n=5)或终端臂中的ThermoGuide软件和相邻插入的热探针(n=3)监测温度变化。通过软件使用43°C下累积等效分钟的温度-时间关系(CEM43)来确定热损伤。将软件计算的温度与温度探针记录的温度进行比较。在组织病理学分析中,将MR测温法给出的热损伤阈值(TDT;2-9、10-59、60-239,≥240CEM43等值线)的尺寸与不可逆损伤的尺寸进行了比较。
    结果:TherewasastrongcorrelationbetweentemperaturerecordingsbyThermalGuideandthosebythermalprobeattoth4mm(r=0.96)and8mm(r=0.80),4和8mm处的平均绝对误差为0.76°C±2.13°C和0.17°C±1.65°C,分别。2-9CEM43的面积大于在组织病理学分析上看到的不可逆损伤的面积。在组织病理学分析中,10和60CEM43的尺寸与病变的尺寸密切相关。在不可逆损伤区域和健康脑组织之间观察到明确的边界(≤1mm)。
    结论:这项临床前评估表明,非冷却LITT系统能够精确地达到目标,并在安全范围内形成明确定义的病变,没有任何不良影响。MR测温软件提供了精确的接近实时的脑组织温度,通过软件可视化的病变尺寸与组织病理学结果密切相关。测试该系统在人类受试者中的有效性和安全性的进一步研究正在进行中。
    OBJECTIVE: MRI-guided laser interstitial thermal therapy (MRgLITT) has recently gained interest as an ablative stereotactic procedure for intractable epilepsy, movement disorders, and brain tumors. Conventionally, a LITT system consists of a laser generator and cooled laser applicator, which is a fiber optic core surrounded by a sheath through which cooled fluid is pumped. However, this footprint can make the system bulky and nonmobile, limit the maximum depth of targeting, and increase the chances of breakdown. Herein, the authors conduct a preclinical assessment of a noncooled MRgLITT system in a porcine model.
    METHODS: Three-tesla MRI was used to guide the in vivo placement of noncooled laser applicators in the porcine brain. The study consisted of a survival arm and terminal arm. The laser was activated at a power of 4-7 W for ≤ 180 seconds. Temperature changes were monitored using the MR thermometry software ThermoGuide in the survival arm (n = 5) or both ThermoGuide software and adjacently inserted thermal probes in the terminal arm (n = 3). Thermal damage was determined by the software using the temperature-time relationship of cumulative equivalent minutes at 43°C (CEM43). Temperatures calculated by the software were compared with those recorded by the temperature probes. The dimensions of thermal damage thresholds (TDTs; 2-9, 10-59, 60-239, ≥ 240 CEM43 isolines) given by MR thermometry were compared with the dimensions of irreversible damage on histopathological analysis.
    RESULTS: There was a strong correlation between temperature recordings by ThermoGuide and those by thermal probes at both 4 mm (r = 0.96) and 8 mm (r = 0.80), with a mean absolute error of 0.76°C ± 2.13°C and 0.17°C ± 1.65°C at 4 and 8 mm, respectively. The area of 2-9 CEM43 was larger than the area of irreversible damage seen on histopathological analysis. The dimensions of the 10 and 60 CEM43 correlated well with dimensions of the lesion on histopathological analysis. A well-defined border (≤ 1 mm) was observed between the area of irreversible damage and healthy brain tissue.
    CONCLUSIONS: This preclinical assessment showed that the noncooled LITT system was able to precisely reach the target and create well-defined lesions within a margin of safety, without any adverse effects. MR thermometry software provided an accurate near-real-time temperature of the brain tissue, and dimensions of the lesion as visualized by the software correlated well with histopathological findings. Further studies to test the system\'s efficacy and safety in human subjects are in progress.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    目的:通过图集图像增强对MR测温运动补偿算法进行了优化和验证。
    方法:在3TPhilipsMRI和ProfiedV1Sonalleve高强度聚焦超声(高强度聚焦超声系统。MR兼容的机器人被配置为在定制明胶体模上引起运动。在进行热疗的同时,在体模上引入了周期性和零星运动的试验。PCA-PDF算法用预测图集增强,以更好地补偿较大的零星运动。
    结果:在周期性运动期间,测温中的温度SD从1提高。1±0。1$1.1\\pm0.1$$到0。5±0。1○$0.5\\pm0。{1}^{\\circ}$$C与原始和增强PCA-PDF应用程序。对于大的零星运动,增强图集从8改进了原始PCA-PDF校正的运动补偿。8±0。5$8.8\\pm0.5$$到0。7±0。1○$0.7\\pm0。{1}^{\\circ}$$C.
    结论:PCA-PDF算法在周期性运动期间将温度精度提高到<1°C,但未能充分解决零星运动。通过增强PCA-PDF算法,大的零星运动过程中的温度SD也降低到<1°C,大大改进了原有的PCA-PDF算法。
    OBJECTIVE: A hybrid principal component analysis and projection onto dipole fields (PCA-PDF) MR thermometry motion compensation algorithm was optimized with atlas image augmentation and validated.
    METHODS: Experiments were conducted on a 3T Philips MRI and Profound V1 Sonalleve high intensity focused ultrasound (high intensity focused ultrasound system. An MR-compatible robot was configured to induce motion on custom gelatin phantoms. Trials with periodic and sporadic motion were introduced on phantoms while hyperthermia was administered. The PCA-PDF algorithm was augmented with a predictive atlas to better compensate for larger sporadic motion.
    RESULTS: During periodic motion, the temperature SD in the thermometry was improved from 1 . 1 ± 0 . 1 $$ 1.1\\pm 0.1 $$ to 0 . 5 ± 0 . 1 ∘ $$ 0.5\\pm 0.{1}^{\\circ } $$ C with both the original and augmented PCA-PDF application. For large sporadic motion, the augmented atlas improved the motion compensation from the original PCA-PDF correction from 8 . 8 ± 0 . 5 $$ 8.8\\pm 0.5 $$ to 0 . 7 ± 0 . 1 ∘ $$ 0.7\\pm 0.{1}^{\\circ } $$ C.
    CONCLUSIONS: The PCA-PDF algorithm improved temperature accuracy to <1°C during periodic motion, but was not able to adequately address sporadic motion. By augmenting the PCA-PDF algorithm, temperature SD during large sporadic motion was also reduced to <1°C, greatly improving the original PCA-PDF algorithm.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    射频导管消融是室性心动过速的既定治疗策略,但仍然与低成功率有关。MR指导室性心动过速显示有望提高这些手术的成功率,特别是由于它有可能使用心脏MR测温术提供有关病变形成的实时信息。现代低场MRI扫描仪(<1T)是MR引导消融的主要利益,因为潜在的好处包括更低的成本,通过减少设备引起的成像伪影和安全限制,增加了患者的访问和设备的兼容性。然而,低场心脏MR测温的可行性尚不清楚.在这项研究中,我们证明了在0.55T时心脏MR测温的可行性,并表征了其体内稳定性(即,精度)使用基于质子共振频率偏移方法的最先进技术。使用基于单次EPI成像的心脏MR测温协议扫描了9名健康志愿者(左心室3片,150个动态,TE=41ms)。重建管道包括图像配准以对齐所有图像,多基线方法(查找表长度=30)来校正呼吸引起的相位变化,和时间滤波来减少温度图中的噪声。测温的稳定性定义为温度随时间变化的逐像素标准偏差。在所有受试者中成功获得心脏MR测温,所有受试者的平均稳定性为1.8±1.0°C。没有多基线校正,总体稳定性为2.8±1.6°C。总之,心脏MR测温在0.55T时是可行的,并且有必要对低场MR引导导管消融进行进一步研究.
    Radiofrequency catheter ablation is an established treatment strategy for ventricular tachycardia, but remains associated with a low success rate. MR guidance of ventricular tachycardia shows promises to improve the success rate of these procedures, especially due to its potential to provide real-time information on lesion formation using cardiac MR thermometry. Modern low field MRI scanners (<1 T) are of major interest for MR-guided ablations as the potential benefits include lower costs, increased patient access and device compatibility through reduced device-induced imaging artefacts and safety constraints. However, the feasibility of cardiac MR thermometry at low field remains unknown. In this study, we demonstrate the feasibility of cardiac MR thermometry at 0.55 T and characterized its in vivo stability (i.e., precision) using state-of-the-art techniques based on the proton resonance frequency shift method. Nine healthy volunteers were scanned using a cardiac MR thermometry protocol based on single-shot EPI imaging (3 slices in the left ventricle, 150 dynamics, TE = 41 ms). The reconstruction pipeline included image registration to align all the images, multi-baseline approach (look-up-table length = 30) to correct for respiration-induced phase variations, and temporal filtering to reduce noise in temperature maps. The stability of thermometry was defined as the pixel-wise standard deviation of temperature changes over time. Cardiac MR thermometry was successfully acquired in all subjects and the stability averaged across all subjects was 1.8 ± 1.0°C. Without multi-baseline correction, the overall stability was 2.8 ± 1.6°C. In conclusion, cardiac MR thermometry is feasible at 0.55 T and further studies on MR-guided catheter ablations at low field are warranted.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    温度是影响几乎所有磁共振特性的标志参数(例如,T1,T2,质子密度,和扩散)。在临床前环境中,温度对动物生理学有很大影响(例如,呼吸频率,心率,新陈代谢,和氧化应激),并且需要仔细调节,特别是当动物处于麻醉状态并且体温调节中断时。我们提出了一种能够调节动物温度的开源加热和冷却系统。该系统是使用能够加热或冷却具有主动温度反馈的循环水浴的珀尔帖模块设计的。反馈是使用商用热敏电阻获得的,放在动物直肠里,并使用比例-积分-微分控制器来调节温度。它的操作在体模以及小鼠和大鼠动物模型中得到了证明,其中收敛时动物温度的标准偏差小于十分之一度。使用侵入性光学探针和非侵入性磁共振波谱测温测量来证明调节小鼠大脑温度的应用。
    Temperature is a hallmark parameter influencing almost all magnetic resonance properties (e.g., T1 , T2 , proton density, and diffusion). In the preclinical setting, temperature has a large influence on animal physiology (e.g., respiration rate, heart rate, metabolism, and oxidative stress) and needs to be carefully regulated, especially when the animal is under anesthesia and thermoregulation is disrupted. We present an open-source heating and cooling system capable of regulating the temperature of the animal. The system was designed using Peltier modules capable of heating or cooling a circulating water bath with active temperature feedback. Feedback was obtained using a commercial thermistor, placed in the animal rectum, and a proportional-integral-derivative controller was used to modulate the temperature. Its operation was demonstrated in a phantom as well as in mouse and rat animal models, where the standard deviation of the temperature of the animal upon convergence was less than a 10th of a degree. An application where brain temperature of a mouse was modulated was demonstrated using an invasive optical probe and noninvasive magnetic resonance spectroscopic thermometry measurements.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    目的:该研究旨在就使用单元件换能器进行经颅聚焦超声(tFUS)热应用的实用性提供见解。
    方法:使用1MHz单元素球形聚焦换能器,通过嵌入基于琼脂的组织模仿凝胶的颅骨体模进行FUS超声处理。头骨模型是用丙烯腈丁二烯苯乙烯(ABS)和树脂热塑性塑料3D打印的,具有健康志愿者的确切头骨几何形状。使用MR测温法在3T磁共振成像(MRI)扫描仪中监测加热期间和加热后的温度场分布。研究了颅骨厚度对颅内加热的影响。
    结果:在接近1580W/cm2的焦点声强度下在自由场中进行60秒的单FUS超声处理,将琼脂体模加热至达到约90°C(基线为37°C)的消融温度。ABS头骨强烈阻挡超声波,导致体模内零温度增加。通过树脂头骨实现了相当大的加热,但它仍然处于高温水平。相反,通过1毫米树脂头骨的tFUS显示增强的超声波穿透和加热,焦点温度达到70℃。
    结论:由于其更高的超声衰减和孔隙率,与树脂颅骨相比,ABS颅骨在tFUS方面表现较差。厚度为1毫米的薄树脂体模提供了一个有效的声学窗口,用于提供tFUS和加热深层体模区域。这些研究的结果对于加速建立更广泛的tFUS应用特别有用。
    OBJECTIVE: The study aims to provide insights on the practicality of using single-element transducers for transcranial Focused Ultrasound (tFUS) thermal applications.
    METHODS: FUS sonications were performed through skull phantoms embedding agar-based tissue mimicking gels using a 1 MHz single-element spherically focused transducer. The skull phantoms were 3D printed with Acrylonitrile Butadiene Styrene (ABS) and Resin thermoplastics having the exact skull bone geometry of a healthy volunteer. The temperature field distribution during and after heating was monitored in a 3 T Magnetic Resonance Imaging (MRI) scanner using MR thermometry. The effect of the skull\'s thickness on intracranial heating was investigated.
    RESULTS: A single FUS sonication at focal acoustic intensities close to 1580 W/cm2 for 60 s in free field heated up the agar phantom to ablative temperatures reaching about 90 °C (baseline of 37 °C). The ABS skull strongly blocked the ultrasonic waves, resulting in zero temperature increase within the phantom. Considerable heating was achieved through the Resin skull, but it remained at hyperthermia levels. Conversely, tFUS through a 1 mm Resin skull showed enhanced ultrasonic penetration and heating, with the focal temperature reaching 70 °C.
    CONCLUSIONS: The ABS skull demonstrated poorer performance in terms of tFUS compared to the Resin skull owing to its higher ultrasonic attenuation and porosity. The thin Resin phantom of 1 mm thickness provided an efficient acoustic window for delivering tFUS and heating up deep phantom areas. The results of such studies could be particularly useful for accelerating the establishment of a wider range of tFUS applications.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    背景:使用磁共振(MR)测温法监测微创热消融程序可以治疗甚至接近关键解剖结构的肿瘤。不幸的是,由于必要的准确性和实时性,术中监测仍然具有挑战性.其中一个原因是MR测量引入的统计误差,这导致消融区的预测变得不准确。
    目的:在这项工作中,我们基于热损伤模型CEM43推导了热消融过程中消融区预测的概率模型。通过将MR测量引起的统计误差集成到常规预测中,我们希望减少错误分类的体素的数量。
    方法:使用聚丙烯酰胺凝胶模型和三个体内猪肝对概率CEM43模型进行了经验评估。
    结果:结果显示,四个数据集中的三个数据具有更高的准确性,与常规模型相比,Sørensen-Dice系数的相对差异从-3.04%$-3.04\\%$到3.97%。此外,与传统模型相比,概率模型预测的消融区假阳性率相对降低11.89%-30.04%.
    结论:所提出的概率热剂量模型可能有助于防止消融区内体素的错误分类。这可能潜在地导致MR引导的热消融程序的成功率增加。未来的工作可能会在更现实的临床背景下解决其他错误源和后续研究。
    BACKGROUND: Monitoring minimally invasive thermo ablation procedures using magnetic resonance (MR) thermometry allows therapy of tumors even close to critical anatomical structures. Unfortunately, intraoperative monitoring remains challenging due to the necessary accuracy and real-time capability. One reason for this is the statistical error introduced by MR measurement, which causes the prediction of ablation zones to become inaccurate.
    OBJECTIVE: In this work, we derive a probabilistic model for the prediction of ablation zones during thermal ablation procedures based on the thermal damage model CEM43 . By integrating the statistical error caused by MR measurement into the conventional prediction, we hope to reduce the amount of falsely classified voxels.
    METHODS: The probabilistic CEM43 model is empirically evaluated using a polyacrilamide gel phantom and three in-vivo pig livers.
    RESULTS: The results show a higher accuracy in three out of four data sets, with a relative difference in Sørensen-Dice coefficient from - 3.04 % $-3.04\\%$ to 3.97% compared to the conventional model. Furthermore, the ablation zones predicted by the probabilistic model show a false positive rate with a relative decrease of 11.89%-30.04% compared to the conventional model.
    CONCLUSIONS: The presented probabilistic thermal dose model might help to prevent false classification of voxels within ablation zones. This could potentially result in an increased success rate for MR-guided thermal ablation procedures. Future work may address additional error sources and a follow-up study in a more realistic clinical context.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Clinical Trial
    目的:了解MRg-LITT质子共振频率-(PRF-)位移测温图像中无法解释的信号伪影是否是由气泡或出血引起的,并表征它们对温度测量的影响。
    方法:检查了IRB批准的颅内MRg-LITT临床试验的回顾性图像数据,检查了消融期间相位数据中观察到的不对称畸变,以前报道过的很可能是出血.总共选择了8例患者:7例发生伪影,1例没有伪影。实施了气泡或出血的数学图像模型,以估计解释临床观察到的相位伪影所需的气泡或出血的大小。相关性和Bland-Altman分析用于确定气泡模型或出血模型是否与临床数据更好地相关。该模型用于将气泡注入干净的PRF相位数据中,而没有伪影,以检查温度分布失真如何随切片方向变化。将模拟的气泡注入数据与包含伪影的临床数据进行比较,以检查气泡对温度和热损伤估计的影响。
    结果:该模型表明,直径约1厘米的气泡可以解释临床观察到的相位伪影。气泡模型预测,出血必须是气泡的2.2倍,以解释临床数据中观察到的相同程度的相位失真。气泡与临床PRF阶段数据的相关性比出血高16%,即使在重新调整出血阶段以更好地匹配数据之后。气泡模型还解释了相位伪影如何导致较大的正温度误差和较大的负温度误差,高达±100°C,这可能会导致几毫米的损伤估计误差。
    结论:结果表明,伪影可能是由气泡而不是出血引起的,可在加热前引入或在加热过程中出现。依赖于PRF偏移测温的设备的制造商和用户应该意识到这些来自气泡伪影的相位失真可能导致大的温度误差。
    To understand if unexplained signal artifacts in MRg-LITT proton resonance frequency- (PRF-) shift thermometry images are caused by air bubbles or hemorrhages, and to characterize their effects on temperature measurements.
    Retrospective image data from an IRB-approved clinical trial of intracranial MRg-LITT were inspected for asymmetric distortions observed in phase data during ablations, which have been previously reported as likely hemorrhages. A total of eight patient cases were selected: seven with artifact occurrence and one without. Mathematical image models for air bubbles or hemorrhages were implemented to estimate the size of the air bubble or hemorrhage needed to explain the clinically observed phase artifacts. Correlations and Bland-Altman analyses were used to determine if an air bubble model or a hemorrhage model was better correlated to the clinical data. The model was used to inject bubbles into clean PRF phase data without artifacts to examine how temperature profile distortions change with slice orientation. The simulated air-bubble injected data were compared to clinical data containing artifacts to examine the bubbles\' effects on temperature and thermal damage estimates.
    The model demonstrated that air bubbles up to approximately 1 cm in diameter could explain the clinically observed phase artifacts. The bubble model predicts that a hemorrhage would have to be 2.2 times as large as an air bubble in order to explain the same extent of phase distortion observed in clinical data. Air bubbles had 16% percent higher correlations to the clinical PRF phase data than hemorrhages, even after rescaling the hemorrhage phases to better match the data. The air bubble model also explains how the phase artifacts lead to both large positive and large negative temperature errors, up to ±100 °C, which could cascade to damage estimate errors of several millimeters.
    Results showed that the artifacts are likely caused by air bubbles rather than hemorrhages, which may be introduced before heating or appear during heating. Manufacturers and users of devices that rely upon PRF-shift thermometry should be aware these phase distortions from bubble artifacts can result in large temperature errors.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    随着乳腺癌发病率和死亡率的逐渐上升,迫切需要改善患者的预后和美容,磁共振成像(MRI)引导下的射频消融(RFA)治疗作为一种新的乳腺癌治疗方法受到了广泛的关注。MRI-RFA导致更高的完全消融率和极低的复发率和并发症率。因此,它可以用作乳腺癌的独立治疗或保乳手术的辅助治疗,以减少乳房切除的程度.此外,在MRI指导下,可以实现RFA的精确控制,乳腺癌治疗可以进入微创的新阶段,安全,和综合治疗。随着MR测温技术的进步,MRI的应用有望拓宽。
    With a gradual increase in breast cancer incidence and mortality rates and an urgent need to improve patient prognosis and cosmetology, magnetic resonance imaging (MRI)-guided radiofrequency ablation (RFA) therapy has attracted wide attention as a new treatment method for breast cancer. MRI-RFA results in a higher complete ablation rate and extremely low recurrence and complication rates. Thus, it may be used as an independent treatment for breast cancer or adjuvant to breast-conserving surgery to reduce the extent of breast resection. Furthermore, with MRI guidance, accurate control of RFA can be achieved, and breast cancer treatment can enter a new stage of minimally invasive, safe, and comprehensive therapy. With progress in MR thermometry technology, the applications of MRI are expected to broaden.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    目的:这项工作的目的是开发一种测温方法,以测量体内温度的升高,具有足以验证热模拟的精度和准确性。这种MR测温模型将是获得关于MR成像中的主要安全问题之一的指示的有价值的工具:由于射频(RF)暴露而发生的组织加热。为了防止温度过度升高,射频功率沉积,表示为比吸收率,不能超过预定义的阈值。使用这些阈值,MRI已经证明了安全使用的广泛历史。然而,MR测温将是解决射频安全评估领域一些未满足需求的宝贵工具。例如比吸收率和热模拟的验证,在扫描过程中调查局部峰值温度,或基于温度的安全指南。
    方法:提出了基于谐波初始化模型的多回波方法。该方法将先前发布的基于模型的多回波水/脂肪分离方法与先前发布的近谐波2D重建方法相结合。该方法在7T的多发射阵列下在人体大腿上进行了测试,在三名志愿者中,和几个RF垫片。
    结果:与以前的脂肪参考方法相比,精度和准确性大大提高(精度:0.09vs.0.19°C)。测量的温升分布与特定对象的模拟对应物的比较显示了多个RF垫片设置的良好相对一致性。
    结论:高精度显示了用于验证目的和其他RF安全性应用的潜力。
    The aim of this work is the development of a thermometry method to measure temperature increases in vivo, with a precision and accuracy sufficient for validation against thermal simulations. Such an MR thermometry model would be a valuable tool to get an indication on one of the major safety concerns in MR imaging: the tissue heating occurring due to radiofrequency (RF) exposure. To prevent excessive temperature rise, RF power deposition, expressed as specific absorption rate, cannot exceed predefined thresholds. Using these thresholds, MRI has demonstrated an extensive history of safe usage. Nevertheless, MR thermometry would be a valuable tool to address some of the unmet needs in the area of RF safety assessment, such as validation of specific absorption rate and thermal simulations, investigation of local peak temperatures during scanning, or temperature-based safety guidelines.
    The harmonic initialized model-based multi-echo approach is proposed. The method combines a previously published model-based multi-echo water/fat separated approach with an also previously published near-harmonic 2D reconstruction method. The method is tested on the human thigh with a multi-transmit array at 7 T, in three volunteers, and for several RF shims.
    Precision and accuracy are improved considerably compared to a previous fat-referenced method (precision: 0.09 vs. 0.19°C). Comparison of measured temperature rise distributions to subject-specific simulated counterparts show good relative agreement for multiple RF shim settings.
    The high precision shows promising potential for validation purposes and other RF safety applications.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

公众号