Glaucoma Drainage Implants

青光眼引流植入物
  • 文章类型: Journal Article
    背景:与传统小梁切除术相比,PreservFlo微分流术的使用由于其易于植入和减少了对术后干预的需求而越来越受欢迎。
    方法:然而,微分流暴露仍然是PreservFlo手术的严重并发症,特别是在有薄Tenon胶囊和结膜的患者中。然而,Tenon囊或结膜的实际厚度和强度只能在手术期间确认。
    方法:剥脱性青光眼与先前的几次青光眼手术有较薄的Tenon囊或结膜。
    方法:我们用手术技术进行了PreservFlo植入,通过在分流管下创建半厚度的矩形巩膜瓣,并将其覆盖在微分流管上,直到远端部分,以恢复薄的Tenon囊和结膜。类似于桥。
    结果:使用该技术,患者的眼压控制较好,外观呈阳性。
    结论:这项技术对于防止暴露和保持顶部都是有益的,除了改善化妆品外观。
    BACKGROUND: The use of the PreserFlo microshunt is gaining popularity owing to its ease of implantation and reduced need for postoperative intervention compared to conventional trabeculectomy.
    METHODS: However, microshunt exposure remains a severe complication of PreserFlo surgery, particularly in patients with a thin Tenon capsule and conjunctiva. However, the actual thickness and intensity of the Tenon capsule or conjunctiva can be confirmed only during surgery.
    METHODS: Exfoliation glaucoma with previous several glaucoma surgeries with thinner Tenon capsule or conjunctiva.
    METHODS: We performed PreserFlo implantation with a surgical technique to recover a thin Tenon capsule and conjunctiva by creating a half-thickness rectangular scleral flap under the shunt and covering it over the microshunt until the distal part, similar to the bridge.
    RESULTS: The patient had better intraocular pressure control with positive cosmetic appearance using this technique.
    CONCLUSIONS: This technique will be beneficial for both preventing exposure and holding down the top, in addition to improving cosmetic appearance.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    背景:为了研究XEN45植入物的长期有效性和安全性,单独或与超声乳化联合,在眼睛开角型青光眼(OAG)。
    方法:对2017年2月至2021年12月接受XEN45植入的连续OAG患者进行回顾性单中心研究。主要终点是平均眼内压(IOP)低于术前值。手术成功定义为IOP从术前值降低≥20%,IOP绝对值在6和13mmHg之间,没有(完全成功)或(合格成功)抗青光眼药物。
    结果:共纳入158只眼(XEN-solo34只(21.5%)眼和XEN+Phaco124只(78.5%)眼)。中位随访时间为28.5个月。在整个研究人群中,术前平均IOP从19.4±6.5mmHg显著降低至12.4±5.0mmHg.XEN-Solo组和XEN+Phaco组的术前平均眼压(95%置信区间)从21.3(19.3-23.2)mmHg和18.8(17.7-20.0)mmHg显著降低至12.0(10.4-13.6)mmHg和12.5(11.6-13.5)mmHg,分别(每个p<0.0001,分别)。在整个研究样本中,低眼压药物的平均数量显着减少(从3.4±0.9降至0.9±1.3,p<0.0001),XEN-Solo(从3.5±1.1到0.6±1.0,p<0.0001,和XEN+Phaco(从3.4±1.1到0.9±1.3,p<0.0001)组。八十四只(53.2%)眼睛被归类为成功,49(58.3%)被列为完全成功。81只(51.3%)眼接受了针刺,15只(9.5%)眼需要额外的外科手术。一只(0.6%)眼患有眼内炎。
    结论:XEN植入物,单独或联合白内障超声乳化术可显著降低IOP,并减少对降眼药物的需求,同时保持良好的安全性。
    BACKGROUND: To investigate the long-term effectiveness and safety of XEN45 implant, either alone or in combination with phacoemulsification, in eyes with open-angle glaucoma (OAG).
    METHODS: Retrospective and single center study conducted on consecutive OAG patients who underwent a XEN45 implant between February-2017 and December-2021. The primary endpoint was the mean intraocular pressure (IOP) lowering from preoperative values. Surgical success was defined as an IOP-lowering from preoperative values ≥ 20% and an IOP absolute value between 6 and 13 mm Hg, without (Complete-success) or with (Qualified-success) antiglaucoma medications.
    RESULTS: A total of 158 eyes (34 (21.5%) eyes XEN-solo and 124 (78.5%) XEN + Phaco) were included. The median follow-up time was 28.5 months. In the overall study population, the mean preoperative IOP was significantly lowered from 19.4 ± 6.5 mm Hg to 12.4 ± 5.0 mm Hg. The mean preoperative (95% confidence interval) IOP was significantly lowered from 21.3 (19.3-23.2) mm Hg and 18.8 (17.7-20.0) mm Hg to 12.0 (10.4-13.6) mm Hg and 12.5 (11.6-13.5) mm Hg in the XEN-Solo and XEN + Phaco groups, respectively (p < 0.0001 each, respectively). The mean number of ocular-hypotensive medications was significantly reduced in the overall study sample (from 3.4 ± 0.9 to 0.9 ± 1.3, p < 0.0001), XEN-Solo (from 3.5 ± 1.1 to 0.6 ± 1.0, p < 0.0001, and XEN + Phaco (from 3.4 ± 1.1 to 0.9 ± 1.3, p < 0.0001) groups. Eighty-four (53.2%) eyes were categorized as success, with 49 (58.3%) classified as complete success. Eighty-one (51.3%) eyes underwent needling and 15 (9.5%) eyes required an additional surgical procedure. One (0.6%) eye had endophthalmitis.
    CONCLUSIONS: XEN implant, either alone or in combination with phacoemulsification significantly lowered IOP and reduced the need of ocular-hypotensive medication, while maintaining a good safety profile.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    目的:本研究旨在评估在新生血管性青光眼患者中注射或不注射抗VEGF的Ahmed青光眼阀(AGV)植入的疗效。
    方法:这项单中心回顾性研究评估了在注射或不注射抗VEGF的情况下接受AGV植入的NVG患者。人口统计学和临床数据,包括眼部发现,眼内压(IOP),视敏度,和青光眼药物计数,在术前和术后一天记录,一个月,和一年。该研究包括35名患者。第1组包括23名在AGV手术前接受抗VEGF注射的患者。第2组,有12名患者,手术前没有注射抗VEGF。成功的手术定义为IOP值在6和21mmHg之间。主要结果是IOP降低30%或更多。
    结果:两组在人口统计学或临床特征上没有显著差异(P>0.05)。手术前和手术后一年的视力在两组之间没有显着差异。然而,在一年的随访结束时,两组的IOP值均显着降低。两组之间在视力方面没有发现显着差异,IOP,或1年随访期间的用药次数(P>0.05)。第1组的成功率为95.7%,第2组的成功率为91.7%。两组并发症比较差异无统计学意义(P>0.05)。
    结论:AGV植入前注射抗VEGF并没有显著影响视力,IOP值,或一年随访期间的药物计数。
    OBJECTIVE: This study aimed to evaluate the efficacy of Ahmed glaucoma valve (AGV) implantation with or without anti-VEGF injections in neovascular glaucoma patients.
    METHODS: This single-center retrospective study assessed NVG patients who underwent AGV implantation with or without anti-VEGF injections. Demographic and clinical data, including ocular findings, intraocular pressure (IOP), visual acuity, and glaucoma medication count, were recorded preoperatively and postoperatively at one day, one month, and one year. The study included 35 patients. Group 1 consisted of 23 patients who received anti-VEGF injections before AGV surgery. Group 2, with 12 patients, had no anti-VEGF injections prior to surgery. Successful surgery was defined as IOP values between 6 and 21mmHg. The primary outcome was a 30% or more reduction in IOP.
    RESULTS: The groups displayed no significant difference in their demographic or clinical profiles (P>0.05). The visual acuity before and one year after surgery did not differ significantly between the groups. However, IOP values significantly decreased by the end of the one-year follow-up for both groups. No significant differences were found between the groups regarding visual acuity, IOP, or the number of medications during the one-year follow-up (P>0.05). Success rates were 95.7% for Group 1 and 91.7% for Group 2. No significant difference in complications between the groups was observed (P>0.05).
    CONCLUSIONS: Anti-VEGF injections prior to AGV implantation did not significantly impact visual acuity, IOP values, or medication count during the one-year follow-up.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    目的:探讨青光眼术后脉络膜脱离患者睫状体脱离的发生率及其对临床病程的影响。管理,和预后。
    方法:前瞻性观察性病例系列研究。纳入2018-2019年青光眼手术后脉络膜脱离的患者。所有患者均接受了完整的眼科检查和超声生物显微镜检查,以评估睫状体脱离的存在和程度。随访检查包括超声生物显微镜扫描在1周,1个月,3个月,和6个月。
    结果:纳入8例患者(8只眼),4男4女,平均年龄72岁(范围60-83)。5例患者行丝裂霉素C小梁切除术(0.02%),其中1例联合超声乳化白内障摘除术;2例接受了Ahmed青光眼瓣膜植入术,1例接受了丝裂霉素C(0.02%)的ab-internoXen45凝胶支架植入。术前平均眼压为26.0±7.65mmHg,术后第一天降至6.9±2.64mmHg。从手术到诊断脉络膜脱离的平均时间为11.6±5.73天。所有患者均通过超声生物显微镜检查发现睫状体脱离,介于一到四个象限之间。所有患者均接受局部类固醇和环麻痹药治疗;三人(37.5%)接受口服类固醇。未显示脉络膜或睫状体脱离的手术干预。
    结论:在这项现实世界的前瞻性研究中,在青光眼手术后出现脉络膜脱离的所有患者中均发现并发睫状体脱离.这一观察结果可能会加深我们对青光眼手术后常见的低眼压机制的理解。
    OBJECTIVE: To examine the rate of ciliary body detachment in patients with choroidal detachment following glaucoma surgery and its effect on the clinical course, management, and prognosis.
    METHODS: A prospective observational case-series study. Patients with choroidal detachment following glaucoma surgery in 2018-2019 were included. All underwent complete ophthalmological examination and ultrasound biomicroscopy for evaluation of the presence and extent of ciliary body detachment. Follow-up examinations including ultrasound biomicroscopy scans were performed at 1 week, 1 month, 3 months, and 6 months.
    RESULTS: Eight patients (8 eyes) were enrolled, 4 male and 4 female, of mean age 72 years (range 60-83). Five patients underwent trabeculectomy with mitomycin C (0.02%), which was combined with phacoemulsification cataract extraction in one; two underwent Ahmed glaucoma valve implantations, and one underwent ab-interno Xen45 gel stent implantation with mitomycin C (0.02%). The mean intraocular pressure was 26.0 ± 7.65 mmHg preoperatively, dropping to 6.9 ± 2.64 mmHg on first postoperative day one. Mean time from surgery to diagnosis of choroidal detachment was 11.6 ± 5.73 days. Ciliary body detachment was identified by ultrasound biomicroscopy in all patients, ranging between one and four quadrants. All patients were treated with topical steroids and cycloplegics; three (37.5%) received oral steroids. No surgical intervention for the choroidal or ciliary body detachments was indicated.
    CONCLUSIONS: In this real-world prospective study, concurrent ciliary body detachment was identified in all patients who presented with choroidal detachment following glaucoma surgery. This observation may deepen our understanding of the mechanism underlying the hypotony that is often seen after glaucoma surgery.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    目前临床应用的青光眼引流装置均为非降解材料。这些不可降解的引流装置通常会引发炎症反应和疤痕增生,可能导致手术失败.我们开发了一种可生物降解的材料羟基磷灰石涂层镁(HA-Mg)作为青光眼引流装置。12只新西兰大白兔随机分为3组:HA-Mg引流板组(6只右眼),小梁切除术组(6只右眼),对照组(12只左眼)。结果表明,所有HA-Mg引流板在术后约4个月均完全降解。术后第5个月,HA-Mg引流板组与对照组的角膜内皮密度无统计学差异(p=0.857)。HA-Mg引流板植入组的眼压(IOP)水平低于其他两组。HA-Mg引流板植入后5个月,锥虫蓝染料仍从前房引流到结膜下。HE染色显示引流板完全降解后,巩膜线状房水引流通道和前粘连,无明显炎性细胞浸润。这项研究表明HA-Mg青光眼引流板在植入兔前房后控制IOP的安全性和有效性。
    The current clinical application of glaucoma drainage devices is made of non-degradable materials. These non-degradable drainage devices often trigger inflammatory responses and scar proliferation, possibly leading to surgical failure. We developed a biodegradable material hydroxyapatite-coated magnesium (HA-Mg) as a glaucoma drainage device. Twelve New Zealand white rabbits were randomly assigned to three groups: HA-Mg drainage plate group (6 right eyes), trabeculectomy group (6 right eyes), and control group (12 left eyes). Results showed that all HA-Mg drainage plates were completely degraded ~4 months postoperatively. At the 5th month postoperatively, there was no statistical difference in the corneal endothelium density between the HA-Mg drainage plate group and the control group (p = 0.857). The intraocular pressure (IOP) level in the HA-Mg drainage plate implantation group was lower than in the other two groups. The trypan blue dye still drained from the anterior chamber to the subconjunctiva 5 months after HA-Mg drainage plate implantation. HE staining revealed the scleral linear aqueous humor drainage channel and anterior synechia were observed after drainage plate completely degraded, with no obvious infiltration with the inflammatory cells. This study showed the safety and efficacy of HA-Mg glaucoma drainage plate in controlling IOP after implantation into the anterior chamber of rabbit eyes.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    目的:描述使用闭合性结膜外途径双重植入Xen45凝胶支架(Xen)的结果。
    方法:回顾性单中心病例系列原发性开角型青光眼患者,在没有结膜开放的情况下,通过abexterno技术在同一只眼睛植入第二个Xen后,进行至少六个月的随访。
    结果:纳入8例患者的8只假晶状体眼。眼压(IOP)从术前的30±2.6mmHg下降到第一次Xen植入后一个月的22.4±2.3mmHg(平均差:-7.6mmHg[95%置信区间:-9.4,-5.9mmHg],p=0.0092)。然后植入第二个Xen以达到目标IOP。术中、术后无明显并发症。第二次植入后6个月,眼压降至16.1±2.7mmHg(平均差:-6.3mmHg[95%置信区间:-7.2,-5.3mmHg],p=0.0183);然而,3例患者需要药物治疗以进一步降低IOP至目标值。
    结论:在结膜闭合的情况下,使用abexterno方法序贯植入两个Xen45凝胶支架似乎是一个有希望的手术,在这个小病例系列中显示出良好的安全性和有效性。该试验数据可能为进一步研究以评估该程序的安全性和有效性铺平道路。
    OBJECTIVE: To describe the outcomes of double implantation of Xen 45 Gel Stent (Xen) using an ab externo approach with closed conjunctiva.
    METHODS: Retrospective single-centre case series of primary open-angle glaucoma patients with at least six months of follow-up after implantation of a second Xen in the same eye via ab externo technique without conjunctival opening.
    RESULTS: Eight pseudophakic eyes of 8 patients were included. Intraocular pressure (IOP) dropped from 30 ± 2.6 mmHg pre-operatively to 22.4 ± 2.3 mmHg one month after the first Xen implant (mean difference: -7.6 mmHg [95% confidence interval: -9.4, -5.9 mmHg], p = 0.0092). A second Xen was then implanted to achieve the target IOP. The procedure showed no significant intraoperative or postoperative complications. The IOP dropped to 16.1 ± 2.7 mmHg six months following this second implant (mean difference: -6.3 mmHg [95% confidence interval: -7.2, -5.3 mmHg], p = 0.0183); however, 3 patients needed medical therapy to further reduce the IOP towards the target value.
    CONCLUSIONS: Sequential implantation of two Xen 45 Gel Stents using an ab externo approach with closed conjunctiva appears a promising procedure that showed a favorable safety and efficacy profile in this small case series. This pilot data might pave the way for further studies to evaluate the safety and efficacy of the procedure.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    背景:使用XEN-45凝胶支架的微创气泡手术尚未建立用于治疗正常眼压性青光眼(NTG)。这项研究的主要目的是评估XEN-45在非受控NTG眼中的长期治疗效果和安全性。
    方法:回顾性分析2016年至2021年在Tuebingen大学医院接受XEN-45凝胶支架植入术的NTG患者。主要结果指标是手术成功三年后定义为眼内压(IOP)降低≥20%,目标IOP在6和15mmHg之间。无论使用局部抗青光眼药物,成功都是完全的,并且是合格的。需要进一步的青光眼手术,除了针刺,被视为失败。次要结果指标包括平均IOP的变化,抗青光眼药物的数量,针刺和并发症的发生率。
    结果:23例患者的28只眼纳入最终分析。三年后,完全和合格的成功率分别为56.5%和75%,分别。术后平均眼压±标准差在三年后从基线时的19.3±2.0mmHg显著下降至13.7±4.2mmHg(n=22;p<0.0001)。抗青光眼药物的中位数量在三年后从2(范围0-4)降至0(范围0-3;p<0.0001)。16只眼睛(57%)需要中位数为1(范围1-3)的针刺程序。一只眼睛需要进一步的青光眼手术。未观察到危及视力的并发症。
    结论:XEN-45支架对于NTG的长期治疗是有效和安全的。然而,经常需要针刺来改善结果.
    BACKGROUND: Minimally invasive bleb surgery using the XEN-45 gel stent has not been established for the treatment of normal-tension glaucoma (NTG). The main objective of this study was to evaluate the long-term treatment efficacy and safety of XEN-45 in eyes with uncontrolled NTG.
    METHODS: A retrospective analysis of patients with NTG who underwent XEN-45 gel stent implantation at university hospital Tuebingen between 2016 and 2021. The primary outcome measure was surgical success after three years defined as lowering of intraocular pressure (IOP) of ≥ 20%, with target IOP between 6 and 15 mmHg. Success was complete without and qualified irrespective of topical antiglaucoma medication use. The need for further glaucoma surgery, except for needling, was regarded as a failure. The secondary outcome measures included changes in mean IOP, number of antiglaucoma medications, and needling and complication rates.
    RESULTS: Twenty-eight eyes from 23 patients were included in the final analysis. Complete and qualified success rates were 56.5% and 75% after three years, respectively. Mean postoperative IOP ± standard deviation decreased significantly after three years from 19.3 ± 2.0 mmHg at baseline to 13.7 ± 4.2 mmHg (n = 22; p < 0.0001). The median number of antiglaucoma medications decreased from 2 (range 0-4) to 0 after three years (range 0-3; p < 0.0001). Sixteen eyes (57%) required a median of 1 (range 1-3) needling procedures. One eye required further glaucoma surgery. No sight-threatening complications were observed.
    CONCLUSIONS: The XEN-45 stent is effective and safe for the long-term treatment of NTG. However, needling was frequently required to improve outcomes.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    背景:Ahmed瓣膜植入在青光眼手术中的比例越来越高,但是预测目标眼压的成功维持仍然是一项具有挑战性的任务。这项研究旨在评估机器学习(ML)在预测Ahmed瓣膜植入后手术结果方面的表现,并评估与手术失败相关的潜在风险因素,以提高成功率。
    方法:本研究使用了2017年至2021年在Ajou大学医院接受Ahmed瓣膜植入的患者的术前数据。这些数据集包括人口统计学和眼科参数(数据集A),不包括精神病记录的系统医疗记录(数据集B),和精神病药物(数据集C)。Logistic回归,极端梯度提升(XGBoost),首先仅使用数据集A对支持向量机进行评估。根据受试者工作特征曲线下面积(AUROC)选择性能最佳的算法。最后,使用最佳性能算法开发了另外三个预测模型,合并多个数据集的组合以预测1年的手术结果。
    结果:在133例患者的153只眼中,131只(85.6%)和22只(14.4%)眼睛被归类为成功和失败组,分别。XGBoost显示为具有0.684的AUROC值的最佳性能模型,仅使用数据集A。基于使用XGBoost模型的多个数据集的组合来开发最后三个进一步的预测模型。所有数据集组合在AUROC方面表现出最佳性能(数据集A+B:0.782;A+C:0.773;A+B+C:0.801)。此外,年龄增长是手术失败发生率较高的危险因素.
    结论:ML在预测Ahmed瓣膜植入1年后的结果方面提供了一定的预测价值。ML评估显示,年龄增长是手术失败的常见风险因素。
    BACKGROUND: Ahmed valve implantation demonstrated an increasing proportion in glaucoma surgery, but predicting the successful maintenance of target intraocular pressure remains a challenging task. This study aimed to evaluate the performance of machine learning (ML) in predicting surgical outcomes after Ahmed valve implantation and to assess potential risk factors associated with surgical failure to contribute to improving the success rate.
    METHODS: This study used preoperative data of patients who underwent Ahmed valve implantation from 2017 to 2021 at Ajou University Hospital. These datasets included demographic and ophthalmic parameters (dataset A), systemic medical records excluding psychiatric records (dataset B), and psychiatric medications (dataset C). Logistic regression, extreme gradient boosting (XGBoost), and support vector machines were first evaluated using only dataset A. The algorithm with the best performance was selected based on the area under the receiver operating characteristics curve (AUROC). Finally, three additional prediction models were developed using the best performance algorithm, incorporating combinations of multiple datasets to predict surgical outcomes at 1 year.
    RESULTS: Among 153 eyes of 133 patients, 131 (85.6%) and 22 (14.4%) eyes were categorized as the success and failure groups, respectively. The XGBoost was shown as the best-performance model with an AUROC value of 0.684, using only dataset A. The final three further prediction models were developed based on the combination of multiple datasets using the XGBoost model. All datasets combinations demonstrated the best performances in terms of AUROC (dataset A + B: 0.782; A + C: 0.773; A + B + C: 0.801). Furthermore, advancing age was a risk factor associated with a higher surgical failure incidence.
    CONCLUSIONS: ML provides some predictive value in predicting the outcomes of Ahmed valve implantation at 1 year. ML evaluation revealed advancing age as a common risk factor for surgical failure.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Case Reports
    暂无摘要。
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    这项研究旨在比较独立PreservfloMicroShunt和iStent第一代植入联合超声乳化术在12个月随访期的高加索患者中降低IOP的有效性和安全性。这项回顾性研究分析了在眼科接受抗青光眼手术的患者的病史,比亚韦斯托克医科大学,2019年1月至2022年1月。主要结局指标是成功率(完整:IOP<18mmHg(标准A)和<15mmHg(标准B)的眼睛比例或在没有任何青光眼药物的情况下IOP降低20%;合格:IOP<18mmHg和<15mmHg的眼睛比例或在有或没有药物的情况下IOP从基线降低20%),平均眼压下降(%),药物负担,并发症的数量和额外的干预措施。在这两组中,与基线相比,在6个月和12个月时观察到IOP和药物负担显著降低.12个月时,在Preservfo和iStent组中,有67.4%和85.7%的患者在标准A中记录了合格的手术成功率,分别(p=0.045)。在12个月时,标准B的手术成功率占Prserflo组患者的61.4%,iStent组患者的32.7%(p=0.04)。12个月时手术失败的患者分别为30.2%和6.3%,分别(p=0.003)。12个月后,两组之间的眼压降低百分比存在显着差异。在Preservflo组中观察到更大的减少,MD=-8.41CI95[-15.88;-0.95],p=0.028,(-33.49%±21.59vs.iStent组-25.07%±14.15)。两种方法均有效降低了青光眼患者的眼压和术后抗青光眼药物的使用。
    This study aimed to compare the IOP-lowering effectiveness and safety of standalone Preserflo MicroShunt and iStent 1st generation implantation combined with phacoemulsification in Caucasian patients with a 12-month follow-up period. This retrospective study analyzed the medical histories of patients undergoing antiglaucoma surgery at the Department of Ophthalmology, Medical University of Bialystok, between January 2019 and January 2022. The main outcome measures were success rates (complete: proportion of eyes with IOP < 18 mmHg (criterion A) and < 15 mmHg (criterion B) or 20% reduction in IOP without any glaucoma medication; qualified: proportion of eyes achieving IOPs < 18 mmHg and < 15 mmHg or 20% reduction in IOP from baseline with or without medications), mean reduction (%) in IOP, medication burden, number of complications and additional interventions. In both groups, a significant decrease in IOP and medication burden were observed at 6 and 12 months when compared with baseline. At 12 months, qualified surgical success in criterion A was recorded in 67.4% and 85.7% of patients in the Preserfo and iStent groups, respectively (p = 0.045). Complete surgical success in criterion B at 12 months accounted to 61.4% of patients from Prserflo group and 32.7% patients in iStent group (p = 0.04). Surgical failure at 12 months was documented in 30.2% and 6.3% of patients, respectively (p = 0.003). There was a significant difference between groups in %IOP reduction after 12 months. Greater reduction was observed in Preserflo group, MD = - 8.41 CI95 [- 15.88; - 0.95], p = 0.028, (- 33.49% ± 21.59 vs - 25.07% ± 14.15 in iStent group). Both procedures effectively reduced IOP and postoperative use of antiglaucoma medications in glaucoma patients.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)