Dental Implant-Abutment Design

牙种植体 - 基牙设计
  • 文章类型: Journal Article
    背景:这项体外研究的目的是通过改善种植体-基台界面的表面粗糙度,研究抛光后处理工艺对机械循环试验前后选择性激光熔化(SLM)基台扭矩损失比和微间隙的影响。
    方法:制作了40个SLM桥台,有20人接受了轻微的反切,指定为抛光,在种植体-基台界面。基台分为三组:SLM基台(A组),原始基台(B组),和抛光的SLM桥台(C组),每个包含20个桥台。使用激光显微镜评价表面粗糙度。对种植体-基台标本进行机械循环试验,和拆卸扭矩值测量前后。扫描电子显微镜(SEM)用于测量样品纵向切片后的微孔。表面粗糙度之间的相关性,扭矩损失比,和微孔进行了评估。采用LSD检验和Tamhane的T2比较对数据进行分析(α=0.05)。
    结果:抛光的SLM基台(6.86±0.64μm)的Sz值与SLM基台(26.52±7.12μm)相比显着降低。抛光SLM桥台的扭矩损失率(24.16%)明显低于SLM桥台(58.26%),而原始基台无统计学差异(18.23%)。抛光SLM基台的种植体-基台微间隙(2.38±1.39μm)明显低于SLM基台(8.69±5.30μm),与原始基台(1.87±0.81μm)相比,差异无统计学意义。发现Sz值与循环试验后的扭矩损失比率之间存在显著正相关(r=0.903,P<0.01)。以及SLM桥台和抛光SLM桥台中所有试样的Sz值和微间隙(r=0.800,P<0.01)。
    结论:这项研究的结果表明,轻微背面切割的抛光步骤可以导致SLM基台界面粗糙度的显着改善,随后优化了种植体-基台的配合。可见,应用小反切法推进了SLM基台的临床应用。
    BACKGROUND: The purpose of this in vitro study was to investigate the effect of polishing post-treatment process on the torque loss ratio and microgap of Selective Laser Melting (SLM) abutments before and after mechanical cycling test through improving the surface roughness of the implant-abutment interface.
    METHODS: Forty SLM abutments were fabricated, with 20 underwent minor back-cutting, designated as polishing, in the implant-abutment interface. The abutments were divided into three groups: SLM abutments (group A), original abutments (group B), and polished SLM abutments (group C), each containing 20 abutments. Surface roughness was evaluated using a laser microscope. Implant-abutment specimens were subjected to mechanical cycling test, and disassembly torque values were measured before and after. Scanning electron microscope (SEM) was used to measure microgap after longitudinal sectioning of specimens. Correlation between surface roughness, torque loss ratio, and microgap were evaluated. LSD\'s test and Tamhane\'s T2 comparison were used to analyze the data (α = 0.05).
    RESULTS: The Sz value of polished SLM abutments (6.86 ± 0.64 μm) demonstrated a significant reduction compared to SLM abutments (26.52 ± 7.12 μm). The torque loss ratio of polished SLM abutments (24.16%) was significantly lower than SLM abutments (58.26%), while no statistically significant difference that original abutments (18.23%). The implant-abutment microgap of polished SLM abutments (2.38 ± 1.39 μm) was significantly lower than SLM abutments (8.69 ± 5.30 μm), and this difference was not statistically significant with original abutments (1.87 ± 0.81 μm). A significant positive correlation was identified between Sz values and the ratio of torque loss after cycling test (r = 0.903, P < 0.01), as well as Sz values and the microgap for all specimens in SLM abutments and polished SLM abutments (r = 0.800, P < 0.01).
    CONCLUSIONS: The findings of this study indicated that the polishing step of minor back-cutting can lead to a notable improvement in the roughness of SLM abutments interface, which subsequently optimized the implant-abutment fit. It can be seen that the application of minor back-cutting method has advanced the clinical use of SLM abutments.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    OBJECTIVE: This test aimed to investigate the factors affecting the locking force between the implant and abutment and the amount of abutment subsidence in pure Morse taper connection implant systems.
    METHODS: With reference to the Bicon implant abutment connection design, different types of implant specimens and their corresponding types of abutments were fabricated. The implant-abutment locking taper was uniformly 1.5°. The locking depths were 1.0, 2.0, and 3.0 mm. The diameters of the locking column were 2.5, 3.0, and 3.5 mm. The thicknesses of the outer wall of the implant were 0.15 and 0.30 mm. The loading forces of the testing machine were 200, 300, and 400 N. At least 10 specimens of each group of implant-abutment were used. All specimens were loaded in the same manner using a universal testing machine (finger pressure + specified loading force, five times). The total height of the implant-abutment was measured before finger pressure, after finger pressure, and after the testing machine was loaded for five times to calculate the amount of sinking of the abutment. Finally, the implant and abutment were pulled apart using the universal testing machine, and the subluxation force was observed and recorded.
    RESULTS: The test loading force, locking depth, and locking post diameter had an effect on the implant-abutment locking force and abutment subsidence. The implant-abutment locking force increased with the increase in the test loading force, locking depth, and locking post diameter (R=0.963, 0.607, and 0.372, respectively), with the test loading force having the most significant effect. Abutment subsidence increased with the increase in test loading force (R=0.645) and decreased with the increase in locking depth and locking post diameter (R=-0.807 and -0.280, respectively), with locking depth having the most significant effect on abutment subsidence. No significant correlation was found between the thickness of the outer wall of the implant and the change in the magnitude of the implant-abutment locking force. However, an increase in the thickness of the outer wall of the implant decreased the amount of abutment subsidence, which was inversely correlated.
    CONCLUSIONS: The locking force of the implant-abutment can be increased by adjusting the design of the pure Morse taper connection implant⁃abutment connection, increasing the locking depth and locking post diameter, and increasing the amount and number of times the abutment is loaded during seating. Problems, such as loosening or detachment of the abutment, can be reduced. The recommended abutment to be loaded should be no less than five times during seating to prevent the abutment from sinking and causing changes in the occlusal relationship in the later stages. Preliminary occlusal adjustments should only be conducted in the early stages of the use of temporary restorations, and final restorations and occlusal adjustments are recommended to be performed after using the abutment for a period of time.
    目的: 探究纯莫氏锥度连接种植系统种植体与基台锁结力大小及基台下沉量的影响因素。方法: 参考Bicon种植体基台连接设计,制作不同型号的种植体试件及其对应型号的基台,种植体—基台锁结锥度统一为1.5°,锁结深度分别为1.0、2.0、3.0 mm,锁结柱直径分别为2.5、3.0、3.5 mm,种植体外壁厚度分别为0.15、0.30 mm,实验机加载力分别为200、300、400 N,每组至少10枚种植体—基台试件。所有试件均采用万能实验机进行同样的加载方式(指压+指定加载力5次),分别于指压前、指压后、实验机加压5次后测量种植体—基台总高度,计算基台下沉量,最后采用万能实验机拉开种植体与基台,观察并记录其锁结力。结果: 实验加载力、锁结深度、锁结柱直径对种植体—基台锁结力及基台下沉量均有影响,种植体—基台锁结力随实验加载力、锁结深度、锁结柱直径的增加而增加(R=0.963、R=0.607、R=0.372),其中实验加载力对种植体—基台锁结力的影响最为显著。基台下沉量随着实验加载力的增加而增加(R=0.645),随锁结深度、锁结柱直径的增加而减少(R=-0.807、R=-0.280),锁结深度对基台下沉量的影响最为明显。种植体外壁厚度与种植体—基台锁结力大小的变化无明显的相关性,但种植体外壁厚度的增加会减少基台的下沉量,两者呈反比关系。结论: 通过调整纯莫氏锥度连接种植体—基台连接设计,增加锁结深度及锁结柱直径,增加基台就位时的加载力大小及加载次数,可以增加种植体—基台锁结力,减少基台松动甚至脱落等问题。同时为防止基台下沉导致后期咬合关系的改变,建议基台就位时加载次数不少于5次,同时建议使用临时修复体及早期只行初步咬合调整,在使用一段时间后再行最终修复及最终的咬合调整。.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • DOI:
    文章类型: Journal Article
    目的:研究不同基台和牙冠对种植体支撑修复体颜色的影响。
    方法:氧化锆和二硅酸锂(例如max)制造具有A2阴影的圆盘以代表两个冠。植入物基台是未经处理的钛,不透明的钛,阳极氧化钛,A2阴影氧化锆和白色氧化锆。4.0mm厚的氧化锆和e.max样品分别用作参考。将牙冠放置在测试的基台上,其间有一滴透明的甘油,并使用数字分光光度计测量颜色。记录CIELab值以评估测试样本与参考之间的色差(ΔE)。
    结果:钛基台的色差高于氧化锆。未处理钛的ΔE值高于不透明钛的ΔE值。对于氧化锆冠,未处理的钛和阳极氧化的钛之间没有差异。氧化锆冠的ΔE值在阴影A2氧化锆和白色氧化锆基台之间没有显着差异;e.max冠显示出显着差异。对于所有钛和A2氧化锆基台,氧化锆冠ΔE值均低于e.max。二硅酸锂冠部和氧化锆基台可能更适合用于植入物支撑的修复体。当不能使用陶瓷基台时,不透明的钛基台可以改善美学区域的颜色。
    结论:二硅酸锂牙冠和氧化锆基台可能是在美学区域与植入物支撑的修复体实现优异颜色匹配的有效方法。
    OBJECTIVE: To investigate the effect of different abutments and crowns on the color of implant-supported restorations.
    METHODS: Zirconia and lithium disilicate (e.max) disks with A2 shade were fabricated to represent two crowns. The implant abutments were untreated titanium, opaqued titanium, anodized titanium, A2 shade zirconia and white zirconia. 4.0 mm-thickness zirconia and e.max specimens were used as references respectively. The crowns were placed on tested abutments with a drop of clear glycerin between them and the color was measured using a digital spectrophotometer. CIELab values were recorded to evaluate color differences (ΔE) between tested specimens and the references.
    RESULTS: Titanium abutments presented higher color differences than zirconia. The ΔE values with untreated titanium were higher than those with opaqued titanium. No differences were found between untreated titanium and anodized titanium for zirconia crowns. The ΔE values of zirconia crowns showed no significant differences between shade A2 zirconia and white zirconia abutments; e.max crowns showed a significant difference. The zirconia crown ΔE values were lower than those of e.max for all titanium and A2 zirconia abutments. Lithium disilicate crowns and zirconia abutments may be more suitable for implant-supported restorations. Opaqued titanium abutment may improve color in esthetic regions when a ceramic abutment cannot be used.
    CONCLUSIONS: Lithium disilicate crowns and zirconia abutments may be an effective method to achieve excellent color matching in esthetic regions with implant-supported restorations.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    Objective: To compare and evaluate the clinical esthetic effect of angle screw channel abutment and personalized zirconia adhesive abutment for single crown restoration in esthetic area. Methods: A total of 44 patients (21 males and 23 females), aged (37.4±13.5) years (18-67 years) who completed single crown restoration in the esthetic area of the Department of Oral Implantology, Dalian Stomatological Hospital from January 2018 to June 2022 were retrospectively selected. A total of 44 implants were inserted. According to the abutment selected for final restoration, the patients were divided into angle screw channel abutment group and personalized zirconia bonding abutment group, with 22 patients and 22 implants in each group. The implant survival rate, complication rate, pink and white esthetic score and marginal bone resorption were compared between the two groups. Results: Follow-up to 12 months after final restoration, implant survival rates were 100% (22/22) in both groups, and there was no statistically significant difference in the incidence of mechanical complications between the two groups [9% (2/22) in the angle screw channel abutment group and 0(0/22) in the personalized zirconia bonding abutment group, χ2=2.10, P=0.148]. In the follow-up appointment 12 months after final restoration, the pink esthetic score of the angle screw channel abutment group (12.95±1.05) was significantly better than that of the personalized zirconia bonding abutment group (11.45±2.02) (t=3.10, P=0.003). There was no significant difference in white esthetic scores between the two groups (t=1.27, P=0.212). There was no significant difference in the marginal bone resorption between the two groups (t=0.32, P=0.749). Conclusions: When a single implant supported restoration is delivered in the esthetic area of the anterior maxilla, high implant survival rate and stability of the marginal bone can be obtained by using the angle screw channel abutment or the personalized zirconia bonding abutment. The clinical efficacy of the angle screw channel system is reliable, and it will provide clinicians with a new treatment option.
    目的: 对比评价上颌前牙种植单冠修复时应用角度螺丝通道基台与个性化氧化锆粘接基台的临床美学效果,为临床提供参考。 方法: 回顾性选择2018年1月至2022年6月于大连市口腔医院种植科完成上颌前牙种植单冠修复的患者44例(男性21例,女性23例),年龄(37.4±13.5)岁(18~67岁),共植入44枚种植体,根据最终修复时选用的基台分为角度螺丝基台组(使用角度螺丝通道基台)和氧化锆基台组(使用个性化氧化锆粘接基台),每组各22例患者(22枚种植体)。比较两组种植体存留率、并发症发生率、粉白美学评分结果及边缘骨吸收量。 结果: 随访至修复完成后12个月,两组种植体存留率均为100%(22/22),两组机械并发症发生率差异无统计学意义[角度螺丝基台组9%(2/22),氧化锆基台组0(0/22)](χ2=2.10,P=0.148)。修复完成后12个月角度螺丝基台组粉色美学评分[(12.95±1.05)分]显著大于氧化锆基台组[(11.45±2.02)分](t=3.10,P=0.003),两组白色美学评分差异无统计学意义(t=1.27,P=0.212)。两组边缘骨吸收量差异无统计学意义(t=0.32,P=0.749)。 结论: 前牙美学区种植单冠修复时采用角度螺丝通道基台或个性化氧化锆粘接基台均能获得良好的种植体存留率,边缘骨水平稳定,角度螺丝通道系统的临床疗效可靠,为临床提供了新的治疗选择。.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    目的:评估两件式基牙工作流程对增强牙种植体周围牙槽骨和牙龈稳定性的影响,并确定患者的满意度。
    方法:共纳入48例后牙列缺损患者,分为两组:两件式基牙工作流程(TAW)和具有浸没式愈合流程的密封螺钉(SHW)。边缘骨水平(MBL),软组织指标,口腔卫生指标,在手术后0,3,6和12个月对患者满意度进行部分评估和记录.主要结果是MBL在不同时间段的变化。使用广义线性混合模型(GLMM)来考虑数据的相关性质,并调整组间差异内的潜在混杂因素。
    结果:在12个月的随访中,植入物和假体的存活率达到100%,TAW组MBL平均减少0.25mm(SD0.23mm),SHW组平均减少0.48mm(SD0.45mm)。6个月内通过GLMM分析,TAW组MBL变化(0.15±0.31mm)明显低于SHW组(0.41±0.41mm),而在12个月内没有发现显著性。此外,假体加载过程中牙龈疼痛和压迫较少,通过视觉模拟评分(VAS,p<0.05)。
    结论:在6个月内,两件式基台工作流程在保持边缘骨水平的完整性方面显示出优异的疗效。此外,它简化了治疗程序,减轻了不适,提高患者满意度。
    OBJECTIVE: To assess the impact of a two-piece abutment workflow on enhancing the stability of the alveolar bone and gingiva surrounding the dental implant, and to determine the level of patient satisfaction.
    METHODS: A total of 48 patients with dentition defect in the posterior region were included and divided into two groups: the two-piece abutment workflow (TAW) and the sealing screw with submerged healing workflow (SHW). Marginal bone level (MBL), soft tissue indicators, oral hygiene indicators, and patient satisfaction were assessed and recorded partially at 0, 3, 6, and 12 months after surgery. The primary outcome was the change of MBL in different time periods. A generalized linear mixed model (GLMM) was used to take into account the correlated nature of the data, and adjust for potential confounding factors within inter-group differences.
    RESULTS: The survival rate of implants and prosthesis reached 100% at 12-month follow-up, with an average decrease of 0.25 mm (SD 0.23 mm) of MBL in the TAW group and 0.48 mm (SD 0.45 mm) in the SHW group. The change of MBL in the TAW group (0.15 ± 0.31 mm) was significantly lower than the SHW group (0.41 ± 0.41 mm) through the analysis of GLMM within 6 months, while no significance was found in 12 months. Moreover, less gingival pain and oppression during prosthesis loading, and less time consumption overall duration were showed in the TAW group through Visual Analogue Scale (VAS, p < 0.05).
    CONCLUSIONS: Within a 6-month period, the two-piece abutment workflow showed superior efficacy in preserving the integrity of the marginal bone level. Furthermore, it streamlined treatment procedures and mitigated discomfort, hence increasing patient satisfaction.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    目的:本研究旨在研究钛基(钛基)基台宏观和微观几何形状对聚合物浸润陶瓷网(PICN)螺钉保留的种植体支撑单冠(iSC)的机械稳定性的影响。
    方法:每组12个标本,由PICNiSC修复的六种不同的种植体/基底基牙组合组成:Nb-T(牙龈高度[GH]:1.5mm,假肢高度[PH]:4.3mm),CC(GH:0.8mm,PH:4.3mm),CC-P(GH:0.8mm,PH:7mm),Nb-V(GH:1.5mm,PH:6mm),St(GH:1.5mm,PH:5.5mm),和Th(GH:0.5mm,pH值:9毫米)。样品经过热机械老化,和那些幸存下来的随后受到静态载荷,直到失败。使用单向ANOVA检验和随后的Tukey事后检验(α=.05)分析数据。
    结果:所有标本均在热机械老化中幸存,无并发症,即,可见裂缝,脱粘,或拧松螺钉。该组在所有组中表现出最高的强度值,与Nb-T相比有显著差异(p<.05),CC(p<.001),和St(p<.001)。此外,与CC(p<.05)和St(p<.05)相比,CC-P组表现出显著优异的断裂强度结果。
    结论:ti-base的选择,特别是假体高度,对PICNiSCs抗骨折能力有显著影响。然而,Ti-base的高度或几何特征在热机械载荷下对iSC/Ti-base组件的机械性能没有显着影响,因为所有标本都经受住了老化,没有并发症或失败。
    OBJECTIVE: This study aimed to investigate the influence of titanium base (ti-base) abutment macro- and micro-geometry on the mechanical stability of polymer-infiltrated ceramic network (PICN) screw-retained implant-supported single crowns (iSCs).
    METHODS: Twelve specimens per group were used, comprising six different implant/ti-base abutment combinations restored with PICN iSCs: Nb-T (gingival height [GH]: 1.5 mm, prosthetic height [PH]: 4.3 mm), CC (GH: 0.8 mm, PH: 4.3 mm), CC-P (GH: 0.8 mm, PH: 7 mm), Nb-V (GH: 1.5 mm, PH: 6 mm), St (GH: 1.5 mm, PH: 5.5 mm), and Th (GH: 0.5 mm, PH: 9 mm). The specimens underwent thermo-mechanical aging, and those that survived were subsequently subjected to static loading until failure. The data were analyzed using a one-way ANOVA test followed by Tukey post hoc test (α = .05).
    RESULTS: All specimens survived thermo-mechanical aging without complications, namely, visible cracks, debonding, or screw loosening. Th group demonstrated the highest strength values among all the groups, with significant differences compared to Nb-T (p < .05), CC (p < .001), and St (p < .001). Additionally, CC-P group exhibited significantly superior fracture strength results compared to CC (p < .05) and St (p < .05).
    CONCLUSIONS: The choice of ti-base, particularly prosthetic height, had a significant influence on fracture resistance of PICN iSCs. Nevertheless, the height or geometrical features of the ti-base did not exhibit a significant influence on the mechanical behavior of the iSC/ti-base assembly under thermomechanical loading, as all specimens withstood the aging without complication or failure.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Randomized Controlled Trial
    目的:比较临床,射线照相,一件式和两件式氧化锆基台之间的免疫学结果支持美学区域的单种植体冠。
    方法:这项研究遵循张口,双盲,和随机对照临床设计,为期3年。将22例符合条件的44个植入物的患者随机分为两组:第1组(带有氧化锆基底的一体式氧化锆基台,n=22)和第2组(带有钛基的两件式氧化锆基台,n=22)。主要结果是技术并发症发生率。此外,存活率,种植体周围液(PICF)中的细胞因子浓度,种植体周围条件,边缘性骨丢失,粉红/白色美学评分(PES/WES)作为次要结局.
    结果:22名患者中有12名参加了为期1年的随访(由于COVID大流行),19名患者参加了为期3年的检查。第1组的两个基牙在功能上10和12个月后断裂。此外,随访1年时,第1组出现1个螺钉松动.第1组的3年技术并发症发生率明显高于第2组(15.79%vs.0%,p<.001)。两组的3年植入物生存率均为100%。PICF中IFN-γ的浓度在第2组中显著上调(p=.018)。此外,IL-6浓度与BOP%呈正相关(p=0.020)。
    结论:两片式氧化锆基台在前部区域进行了3年的随访期间,与单件式设计相比,表现出优越的技术性能。然而,需要进一步的长期研究来验证两件式氧化锆基台的免疫学稳定性。
    OBJECTIVE: To compare the clinical, radiographic, and immunological outcomes between one-piece versus two-piece zirconia abutments supported single implant crowns in the esthetic region.
    METHODS: The study followed a split-mouth, double-blind, and randomized controlled clinical design for a duration of 3 years. Twenty-two eligible patients with 44 implants were randomly assigned to two groups: Group 1 (one-piece zirconia abutment with zirconia base, n = 22) and Group 2 (two-piece zirconia abutment with titanium base, n = 22). The primary outcome was the technical complication rate. Additionally, survival rates, cytokines concentrations in peri-implant crevicular fluid (PICF), peri-implant conditions, marginal bone loss, and pink/white esthetics score (PES/WES) were assessed as secondary outcomes.
    RESULTS: Twelve of 22 patients attended the 1-year follow-up (due to the COVID pandemic), and 19 patients attended the 3-year examination. Two abutments in Group 1 were fractured after 10 and 12 months in function. Additionally, one screw loosening occurred in Group 1 at 1-year follow-up. The 3-year technical complication rate was significantly higher in Group 1 than that in Group 2 (15.79% vs. 0%, p < .001). The 3-year implant survival rate was 100% in both groups. The concentration of IFN-γ in PICF was significantly upregulated in Group 2 (p = .018). Furthermore, the IL-6 concentration was positively correlated with BOP% (p = .020).
    CONCLUSIONS: Two-piece zirconia abutments exhibited superior technical performance compared to one-piece designs during a 3-year follow-up in the anterior region. However, further long-term research is necessary to verify the immunological stability of two-piece zirconia abutments.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    氧化锆材料由于其优异的物理和美学性能而越来越多地用于植入物康复。稳定的种植体周围上皮组织与经粘膜种植体基台的粘附可以显着增强种植体长期稳定性的功效。然而,由于氧化锆材料的强生物惰性,难以与种植体周围上皮组织形成稳定的化学或生物结合。在本研究中,我们研究了氧化锆的钙水热处理是否促进种植体周围上皮组织的密封。采用扫描电镜和能量色散光谱法进行体外实验,分析钙水热处理对氧化锆表面形貌和成分的影响。粘附蛋白的免疫荧光染色,即,在人牙龈成纤维细胞系(HGF-1)细胞中进行F-肌动蛋白和整联蛋白β1。在钙水热处理组中,这些贴壁蛋白的表达更高,HGF-1细胞增殖增加。通过提取大鼠的上颌右第一磨牙并用微型氧化锆基牙植入物代替它们进行了体内研究。钙水热处理组在氧化锆基台表面表现出更好的附着,在植入后2周抑制辣根过氧化物酶的渗透。这些结果表明,氧化锆的钙水热处理改善了种植体基台与周围上皮组织之间的密封,有可能增加植入物的长期稳定性。
    Zirconia materials have been increasingly used in implant rehabilitation due to their excellent physical and esthetic properties. Stable peri-implant epithelial tissue adhesion to the transmucosal implant abutment may significantly enhance the efficacy of implant long-term stability. However, it is difficult to form stable chemical or biological bindings with peri-implant epithelial tissue due to the strong biological inertia of zirconia materials. In the present study, we investigated whether calcium hydrothermal treatment of zirconia promotes sealing of peri-implant epithelial tissue. In vitro experiments were performed to analyze the effects of calcium hydrothermal treatment on zirconia surface morphology and composition by scanning electron microscopy and energy dispersive spectrometry. Immunofluorescence staining of adherent proteins, namely, F-actin and integrin β1, in human gingival fibroblast line (HGF-l) cells was performed. In the calcium hydrothermal treatment group, there was higher expression of these adherent proteins and increased HGF-l cell proliferation. An in vivo study was conducted by extracting the maxillary right first molars of rats and replacing them with mini-zirconia abutment implants. The calcium hydrothermal treatment group showed better attachment at the zirconia abutment surface, which inhibited horseradish peroxidase penetration at 2 weeks post-implantation. These results demonstrated that calcium hydrothermal treatment of zirconia improves the seal between the implant abutment and surrounding epithelial tissues, potentially increasing the long-term stability of the implant.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    研究证实,牙齿脱落与全身性疾病密切相关,比如肥胖,糖尿病,心血管疾病,某些类型的肿瘤,和老年痴呆症。在许多牙齿修复方法中,种植体修复是最常用的方法。植入后,植入物的长期稳定性不仅需要良好的骨结合,还需要植入物与周围软组织之间良好的软组织密封。氧化锆基牙用于临床种植修复治疗,但由于氧化锆的强烈生物惰性,很难与周围组织形成稳定的化学或生物键。在这项研究中,我们研究了通过水热法在氧化锆基台表面合成氧化锌(ZnO)纳米晶,使其更有利于软组织的早期封闭和分子机制。体外实验发现,不同的水热处理温度影响ZnO晶体的形成。ZnO的晶体直径在不同温度下由微米级变为纳米级,晶体形态也发生变化。体外,扫描电子显微镜,能量色散光谱法,和实时聚合酶链反应结果表明,ZnO纳米晶可以通过促进层粘连蛋白332与整合素β4的结合,调节PI3K/AKT通路,促进口腔上皮细胞在氧化锆表面的附着和增殖。在体内,ZnO纳米晶体最终促进软组织密封的形成。总的来说,ZnO纳米晶可以通过水热处理在氧化锆表面上合成。它可以帮助在植入物基台和周围软组织之间形成密封。该方法有利于植入物的长期稳定性,也可应用于其他医疗领域。
    Studies have confirmed that tooth loss is closely related to systemic diseases, such as obesity, diabetes, cardiovascular diseases, some types of tumors, and Alzheimer\'s disease. Among many methods for tooth restoration, implant restoration is the most commonly used method. After implantation, long-term stability of implants requires not only good bone bonding but also good soft tissue sealing between implants and surrounding soft tissues. The zirconia abutment is used in clinical implant restoration treatment, but due to the strong biological inertia of zirconia, it is difficult to form stable chemical or biological bonds with surrounding tissues. In this study, we investigated synthesized zinc oxide (ZnO) nanocrystal on the zirconia abutment surface by the hydrothermal method to make it more beneficial for soft tissue early sealing and the molecular mechanism. In vitro experiments found that different hydrothermal treatment temperatures affect the formation of ZnO crystals. The crystal diameter of ZnO changes from micron to nanometer at different temperatures, and the crystal morphology also changes. In vitro, scanning electron microscopy, energy dispersive spectrometry, and real-time polymerase chain reaction results show that ZnO nanocrystal can promote the attachment and proliferation of oral epithelial cells on the surface of zirconia by promoting the binding of laminin 332 and integrin β4, regulating the PI3K/AKT pathway. In vivo, ZnO nanocrystal ultimately promotes the formation of soft tissue seals. Collectively, ZnO nanocrystal can be synthesized on a zirconia surface by hydrothermal treatment. It can help to form a seal between the implant abutment and surrounding soft tissue. This method is beneficial to the long-term stability of the implant and also can be applied to other medical fields.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • DOI:
    文章类型: English Abstract
    目的:研究钛基两种生产工艺对种植体单修复体的准确性,为合适的种植体单修复体工艺提供实验数据。
    方法:选取上海市普陀区眼病牙病防治所口腔种植科单颗后牙缺失患者30例。患者在临床上对两个装有植入物类似物的石膏模型进行了2种传统印象。然后根据不同的生产工艺将这些模型分为2组。实验组用安装在模型植入物模拟中的扫描体进行扫描,而对照组直接在安装在模型植入物类似物中的Ti基基基上进行扫描。沿颊舌侧切开两组种植体承载的单个修复体,并通过电子显微镜观察测量点到Ti-base基牙之间的距离。此外,观察了氧化锆冠的断裂极限,万能试验机用于加载直接力到表冠。采用SPSS22.0软件包进行数据分析。
    结果:实验组的种植体承载单修复体与Ti基基牙之间的间隙明显小于对照组。两组比较差异有统计学意义(P<0.05)。然而,通过测试氧化锆冠的断裂极限,差异无统计学意义(P>0.05)。
    结论:使用扫描体将种植体位置和钛基基牙数据信息传输到数字牙科设计软件比直接在模型模拟上扫描钛基更准确和可靠。建议使用扫描体来加工和制造植入式单修复体。
    OBJECTIVE: To study on the accuracy of implant-borne single restoration by two production processes with Ti-base to provide experimental data for proper processes of single implant-borne restoration.
    METHODS: Thirty patients were selected with single posterior teeth missing from the Department of Oral Implantology of Shanghai Putuo District Eye Disease and Dental Disease Prevention and Treatment Institute. The patients were taken 2 traditional impressions clinically for two plaster model equipped with implant analogue. These models were then divided into 2 groups according to different production processes. The experimental group was scanned with the scan body installed in the model implant analogue, while the control group was scanned directly on the Ti-base abutment installed in the model implant analogue. The implant-borne single restorations of the two groups were cut along the buccal-lingual side and the distance between the measuring point to the Ti-Base abutment was observed by electron microscopy. In addition, the breaking limit of zirconia crown was observed, universal test machine was used to load direct force to the crown. SPSS 22.0 software package was used for data analysis.
    RESULTS: The gap between the implant-borne single restoration to the Ti-base abutment of the experimental group was significantly smaller than that of the control group. The difference between the two groups was statistically significant (P<0.05). However, by testing the breaking limit of zirconia crown, there was no significant difference(P>0.05).
    CONCLUSIONS: Using scan body to transfer the implant position and Ti-base abutment data information to the digital dental design software is more accurate and reliable than directly scanning the Ti-base on the model analogue. Using scan body is recommended for processing and manufacture of implant-borne singe restoration.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

公众号