    OBJECTIVE: Abdominal organ motion seriously compromises the targeting accuracy for particle therapy in patients with pancreatic adenocarcinoma. This study compares three different abdominal corsets regarding their ability to reduce pancreatic motion and their potential usability in particle therapy.
    METHODS: A patient-individualized polyurethane (PU), a semi-individualized polyethylene (PE), and a patient-individualized three-dimensional-scan based polyethylene (3D-PE) corset were manufactured for one healthy volunteer. Time-resolved volumetric four-dimensional-magnetic resonance imaging (4D-MRI) and single-slice two-dimensional (2D) cine-MRI scans were acquired on two consecutive days to compare free-breathing motion patterns with and without corsets. The corset material properties, such as thickness variance, material homogeneity in Hounsfield units (HU) on computed tomography (CT) scans, and manufacturing features were compared. The water equivalent ratio (WER) of corset material samples was measured using a multi-layer ionization chamber for proton energies of 150 and 200 MeV.
    RESULTS: All corsets reduced the pancreatic motion on average by 9.6 mm in inferior-superior and by 3.2 mm in anterior-posterior direction. With corset, the breathing frequency was approximately doubled and the day-to-day motion variations were reduced. The WER measurements showed an average value of 0.993 and 0.956 for the PE and 3DPE corset, respectively, and of 0.298 for the PU corset. The PE and 3DPE corsets showed a constant thickness of 2.8 ± 0.2 and 3.8 ± 0.2 mm, respectively and a homogeneous material composition with a standard deviation (SD) of 31 and 32 HU, respectively. The PU corset showed a variable thickness of 4.2 - 25.6 mm and a heterogeneous structure with air inclusions with an SD of 113 HU.
    CONCLUSIONS: Abdominal corsets may be effective devices to reduce pancreatic motion. For particle therapy, PE-based corsets are preferred over PU-based corset due to their material homogeneity and constant thickness.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Sci-hub)

       PDF(Pubmed)

  • 文章类型: Case Reports
    据报道,一名50岁男子在左侧腹侧患有横纹肌肉瘤转移灶。为了局部控制病变,患者接受了50.4Gy放疗和伴随化疗。使用Tri-60-Co磁共振混合放射治疗装置进行治疗,并将呼吸门控协议应用于不同的呼吸阶段(自由呼吸,深呼吸保持和最终呼气保持)。计算了三个强度调制放射治疗(IMRT)计划,并且由于不存在PTV覆盖差异和较好的有风险的器官(即肾脏),最终选择了最终呼气呼吸保持计划。该病例报告表明,在MRI引导的呼吸门控放射治疗中避免处于危险中的器官是可行的,并且在保留处于危险中的器官具有最大的剂量学或临床重要性时尤其有利。
    The case of a 50-year-old man affected by a rhabdomiosarcoma metastatic lesion in the left flank Is reported. The patient was addressed to 50.4 Gy radiotherapy with concomitant chemotherapy in order to locally control the lesion. A Tri-60-Co magnetic resonance hybrid radiotherapy unit was used for treatment delivery and a respiratory gating protocol was applied for the different breathing phases (Free Breathing, Deep Inspiration Breath Hold and Final Expiration Breath Hold). Three intensity modulated radiation therapy (IMRT) plans were calculated and Final Expiration Breath Hold plan was finally selected due to the absence of PTV coverage differences and better organs at risk sparing (i.e. kidneys). This case report suggests that organs at risk avoidance with MRI-guided respiratory-gated Radiotherapy is feasible and particularly advantageous whenever sparing the organs at risk is of utmost dosimetric or clinical importance.