Scheimpflug

Scheimpflug
  • 文章类型: Journal Article
    背景:评估两种不同的眼生物计和Scheimpflug装置在圆锥角膜眼中的可重复性和一致性。
    方法:这种前瞻性,比较试验在大学医院进行,歌德大学,美因河畔法兰克福,德国。我们包括圆锥角膜的眼睛,每个病人一只眼睛,随机选择。用GalileiG6(Ziemer,瑞士)和PentacamAXL(Oculus,德国),每个连续测量三次。对模拟角膜曲率术(simK)的可重复性和一致性进行了评估,散光(sima),最大角膜曲率(KMax)及其轴,总角膜曲率测量(TCP),轴向长度(AL),前房深度(ACD),和最薄的测厚仪(TCT)。
    结果:两种设备均显示出出色的可重复性,所有参数的类内相关性(ICC)均>0.97。对于所有参数,95%的重复性极限(LoR95%)和一致性(LoA95%)是窄的。GalileiG6对TCT的LoAR95%较窄(2.1μm与4.6μm),但KMax的LoR95%(0.52Dvs.1.18D)。其他参数没有发现相关差异。设备之间的协议很好,适度,特别是对于simK和TCP。
    结论:两种设备均具有出色的可重复性,所有参数均具有窄的LoR95%和高的ICC。发现KMax和TCT的唯一相关差异分别有利于PentacamAXL和GalileiG6。协议是好到适度的,和大多数参数不应该被认为是可互换的。
    BACKGROUND: To evaluate the repeatability and agreement of two different ocular biometers and Scheimpflug devices in keratoconus eyes.
    METHODS: This prospective, comparative trial took place at the University hospital, Goethe University, Frankfurt am Main, Germany. We included eyes with keratoconus, one eye per patient, randomly selected. Measurements were taken with Galilei G6 (Ziemer, Switzerland) and Pentacam AXL (Oculus, Germany), three consecutive measurements each. Repeatability and agreement were evaluated for simulated keratometry (simK), astigmatism (simA), maximum keratometry (KMax) and its axis, total keratometry (TCP), axial length (AL), anterior chamber depth (ACD), and thinnest pachymetry (TCT).
    RESULTS: Both devices showed an excellent repeatability with intra class correlation (ICC) of > 0.97 for all parameters. The 95% limits of repeatability (LoR95%) and agreement (LoA95%) were narrow for all parameters. The Galilei G6 had a narrower LoAR95% for TCT (2.1 μm vs. 4.6 μm), but a wider LoR95% for KMax (0.52D vs. 1.18D). No relevant difference was found for the other parameters. Agreement between the devices was good to moderate, especially for simK and TCP.
    CONCLUSIONS: Both devices show excellent repeatability with narrow LoR95% and high ICC for all parameters. The only relevant difference was found for KMax and TCT in favor of Pentacam AXL and Galilei G6, respectively. Agreement was good to moderate, and most parameters should not be considered interchangeable.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    自1980年代以来,异维A酸一直是中度和重度寻常痤疮的最佳治疗选择。一些研究表明,用异维甲酸治疗的患者有亚临床眼前节受累的证据。
    本研究旨在使用Scheimpflug断层扫描的密度测定软件评估接受异维A酸治疗的患者的晶状体透明度,并与健康对照组进行比较。
    37例符合纳入标准的接受异维A酸治疗的痤疮患者(男性24例,女性13例,平均年龄22.94±4.21岁)和39名健康对照受试者被纳入研究。记录异维A酸和对照受试者的临床特征。使用Scheimpflug层析成像设备(PentacamHR,Oculus,Wetzlar,德国)。
    各组之间在年龄上没有统计学上的显着差异,性别分布,球形当量,或通过Pentacam系统测量的眼前节参数(所有p>0.05)。异维A酸组2区和3区的晶状体密度值显著较高(p=0.042,p<0.001),并且与累积异维A酸剂量呈正相关(2区:r=0.384,p=0.032;3区:r=0.384,p=0.005)。
    与健康对照相比,用异维甲酸治疗的患者的2区和3区晶状体密度更高。
    UNASSIGNED: Isotretinoin has been the best treatment option for moderate and severe acne vulgaris since the 1980s. Some studies have shown evidence of subclinical anterior segment involvement of the eye in patients treated with isotretinoin.
    UNASSIGNED: This study aimed to evaluate lens clarity with the densitometry software of Scheimpflug tomography in patients treated with isotretinoin and to compare with healthy control subjects.
    UNASSIGNED: Thirty-seven acnepatients treated with isotretinoin who met the inclusion criteria (24 males and 13 females, mean age 22.94 ± 4.21 years) and 39 healthy control subjects were included in the study. Clinical characteristics of the isotretinoin and control subjects were recorded. Lens density was evaluated with the densitometry software of the Scheimpflug tomography device (PentacamHR, Oculus, Wetzlar, Germany).
    UNASSIGNED: There was no statistically significant difference between the groups in age, gender distribution, spherical equivalent, or anterior segment parameters measured by the Pentacam system (p > 0.05 for all). Lens density values in zones 2 and 3 were significantly higher in the isotretinoin group (p = 0.042, p < 0.001) and positively correlated with cumulative isotretinoin dose (zone 2: r = 0.384, p = 0.032; zone 3: r = 0.384, p = 0.005).
    UNASSIGNED: Zone 2 and zone 3 lens density are higher in patients treated with isotretinoin when compared to healthy controls.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    本研究旨在探讨单纯性近视性屈光参差患者双眼角膜参数的差异及其与双眼生物参数不对称性的相互关系,从而阐明近视过程对各种角膜参数的影响。
    在这项横断面研究中,纳入单眼近视性屈光参差患者65例。他们被分为低屈光参差组:3.00D<Δ等效球形(SE)≤-1.00D(Δ代表两只眼睛之间的差异,即,近视数据减去正视数据)和高屈光参差组:ΔSE≤-3.00D。使用Pentacam测量角膜和眼部生物特征参数,科维斯ST,IOLMaster700.统计分析集中在双眼角膜参数不对称性上,使用对侧正视作为对照。
    参与者的平均年龄为18.5±1.3岁,近视和正视的平均SE为-2.93±1.09D和-0.16±0.41D,分别。中央角膜厚度(CCT),平面角膜曲率测量(Kf),角膜曲率测量散光(Ka),总角膜像差(6mm)(TOA),表面方差指数(ISV),垂直不对称指数(IVA),应力-应变指数(SSI),第一压平刚度参数(SPA1)和羊角关系厚度-水平(ARTh)在屈光参差的同眼之间显示出显着差异(p<0.05)。ΔIVA有显著差异,Δ下角膜和上角膜(I-S)的平均屈光力之差,Belin/Ambrósio增强扩张显示的Δ偏差值(BAD-D),两组的Δ变形幅度比max(2mm)(DAR)和Δ断层生物力学指数(TBI)(p<0.05)。角膜参数的不对称性与眼部生物特征参数的不对称性相关。屈光参差(ΔSE)与ΔIVA呈正相关(r=0.255,p=0.040),ΔBAD-D(r=0.360,p=0.006),在多元回归分析中,ΔSSI(r=0.276,p=0.039)与ΔDAR(r=-0.329,p=0.013)呈负相关。Δ平均角膜曲率(Km),Δ前房深度(ACD),和Δ生物力学校正的眼内压(bIOP)也与双眼角膜差异有关。
    与对侧正视相比,近视眼角膜较薄,角膜散光较小。近视角膜表现出相对更规则的表面形态,但更容易变形,并具有较差的生物力学特性。此外,屈光参差与角膜参数不对称性有一定的相关性,这将有助于预测近视的发展。
    UNASSIGNED: This study aims to investigate the differences in binocular corneal parameters and their interrelation with binocular biometric parameters asymmetry in patients with simple myopic anisometropia, thereby elucidating the influence of myopia process on various corneal parameters.
    UNASSIGNED: In this cross-sectional study, 65 patients with anisometropia in monocular myopia were included. They were divided into low anisometropia group: 3.00D<Δ spherical equivalent (SE)≤-1.00D (Δ represents the difference between the two eyes, i.e., myopic data minus emmetropic data) and high anisometropia group: ΔSE ≤ -3.00D. Corneal and ocular biometric parameters were measured using Pentacam, Corvis ST, and IOL Master 700. Statistical analyses focused on the binocular corneal parameters asymmetry, using the contralateral emmetropia as a control.
    UNASSIGNED: The mean age of participants was 18.5 ± 1.3 years, with the average SE for myopia and emmetropia being -2.93 ± 1.09D and -0.16 ± 0.41D, respectively. The central corneal thickness (CCT), flat keratometry (Kf), keratometry astigmatism (Ka), total corneal aberration (6 mm) (TOA), surface variance index (ISV), vertical asymmetry index (IVA), stress-strain index (SSI), and first applanation stiffness parameter (SPA1) and ambrosia relational thickness-horizontal (ARTh) showed significant differences between anisometropic fellow eyes (p < 0.05). There were significant differences in ΔIVA, Δ the difference between the mean refractive power of the inferior and superior corneas (I-S), Δ deviation value of Belin/Ambrósio enhanced ectasia display (BAD-D), Δ deformation amplitude ratio max (2 mm) (DAR)and Δ tomographic biomechanical index (TBI) (p < 0.05) in two groups. Asymmetry of corneal parameters was correlated with asymmetry of ocular biometric parameters. Anisometropia (ΔSE) was positively correlated with ΔIVA (r = 0.255, p = 0.040), ΔBAD-D (r = 0.360, p = 0.006), and ΔSSI (r = 0.276, p = 0.039) and negatively correlated with ΔDAR (r = -0.329, p = 0.013) in multiple regression analysis. Δ mean keratometry (Km), Δ anterior chamber depth (ACD), and Δ biomechanically corrected intraocular pressure (bIOP) were also associated with binocular corneal differences.
    UNASSIGNED: Compared to contralateral emmetropia, myopic eyes have thinner corneas and smaller corneal astigmatism. Myopic corneas exhibit relatively more regular surface morphology but are more susceptible to deformation and possess marginally inferior biomechanical properties. In addition, there is a certain correlation between anisometropia and corneal parameter asymmetry, which would be instrumental in predicting the development of myopia.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    使用CORVIS-ST装置在近视个体中表征角膜生物力学特性。
    这项前瞻性横断面研究包括近视患者。我们的研究包括154名年龄在18至40岁之间的近视患者的154只眼睛,有稳定的折射至少2年。使用PentacamHR设备进行全面的眼科检查和角膜断层扫描。使用CORVIS-ST装置评估角膜生物力学参数,中度,严重,和极端近视群体。
    在DA比率中观察到统计学上的显着差异(p=0.033),SP-A(p=0.009),CBI(p=0.041),SSI(p=0.000),和峰值距离(p=0.032)。与不同的CorvisST生物力学变量相关,发现SE与DA比率相关(r=-0.191,p=0.018),SP-A(r=0.199,p=0.013)和SSI(r=-0.336,p=0.000),而在多元回归分析中,发现SE与SSI和峰值距离独立相关(分别为p=0.036,0.038),而近视等级与SP-A独立相关(p=0.034)。
    SSI,峰值距离,和SP-A与SE和近视分级独立相关,证实了近视程度较高的眼睛更易变形且耐应力较小的假设。
    UNASSIGNED: To characterize corneal biomechanical properties using the CORVIS-ST device in myopic individuals.
    UNASSIGNED: This prospective cross-sectional study included patients with myopia. Our study included 154 eyes of 154 myopic patients aged between 18 and 40 years, with stable refraction for at least 2 years. A full ophthalmological examination and corneal tomography were performed using a Pentacam HR device. Corneal biomechanical parameters were assessed using the CORVIS-ST device in mild, moderate, severe, and extreme myopia groups.
    UNASSIGNED: Statistically significant differences were observed in the DA ratio (p = 0.033), SP-A (p=0.009), CBI (p=0.041), SSI (p=0.000), and Peak distance (p = 0.032). In correlation with different Corvis ST biomechanical variables, SE was found to be correlated with DA ratio(r=-0.191, p=0.018), SP-A(r=0.199, p=0.013) and SSI(r=-0.336, p=0.000), while in multiple regression analysis, SE was found to be independently correlated with SSI and peak distance(p=0.036,0.038 respectively) while the grade of myopia was found to be independently correlated with SP-A(p=0.034).
    UNASSIGNED: SSI, Peak distance, and SP-A were independently related to SE and myopia grade, confirming the hypothesis that eyes with higher myopia are more deformable and less stress resistant.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    目的:本研究评估了屈光结果与术后前房深度之间的关系(ACD,从角膜上皮到晶状体测量)通过扫频源光学相干断层扫描(SS-OCT)测量,光学低相干反射计(OLCR),以及未扩张瞳孔下的Scheimpflug装置。
    方法:在医院环境中接受白内障超声乳化人工晶状体(IOL)植入术的患者。术后ACD(postACD)用SS-OCT装置进行,OLCR设备,白内障手术后至少1个月使用Scheimpflug装置。将平均预测误差调整为0后,使用从3台设备测量的实际postACD和预测值,用Olsen公式计算屈光结果的差异。
    结果:总体而言,这项比较性病例研究包括69名患者的69只眼,并且使用所有3台设备成功进行ACD后测量。用SS-OCT测量的后ACD,OLCR,Scheimpflug装置分别为4.59±0.30、4.50±0.30和4.54±0.32mm,分别。在3种设备中发现了后ACD的统计学差异(P<0.001),与类内相关系数(ICCs)和Bland-Altman表现出良好的一致性。使用使用3种设备获得的实际postACD,使用Olsen公式没有发现中位数绝对误差的显着差异。65%的预测误差百分比在±0.50D内(OLCR),70%(Scheimpflug),和67%(SS-OCT)计算的实际后ACD与64%的预测值。
    结论:在从SS-OCT获得的ACD后测量中发现了基本一致,OLCR,和Scheimpflug装置,奥尔森公式的屈光结果具有可比性的趋势。同时,ACD后测量对于屈光结局的额外增强可能具有潜在优势.
    OBJECTIVE: This study evaluated the relationship between refractive outcomes and postoperative anterior chamber depth (ACD, measured from corneal epithelium to lens) measured by swept-source optical coherence tomography (SS-OCT), optical low-coherence reflectometry (OLCR), and Scheimpflug devices under the undilated pupil.
    METHODS: Patients undergoing cataract phacoemulsification with intraocular lens (IOL) implantation in a hospital setting were enrolled. Postoperative ACD (postACD) was performed with an SS-OCT device, an OLCR device, and a Scheimpflug device at least 1 month after cataract surgery. After adjusting the mean predicted error to 0, differences in refractive outcomes were calculated with the Olsen formula using actual postACD measured from 3 devices and predicted value.
    RESULTS: Overall, this comparative case study included 69 eyes of 69 patients, and postACD measurements were successfully taken using all 3 devices. The postACD measured with the SS-OCT, OLCR, and Scheimpflug devices was 4.59 ± 0.30, 4.50 ± 0.30, and 4.54 ± 0.32 mm, respectively. Statistically significant differences in postACD were found among 3 devices (P < 0.001), with intraclass correlation coefficients (ICCs) and Bland-Altman showing good agreement. No significant difference in median absolute error was found with the Olsen formula using actual postACD obtained with 3 devices. Percentage prediction errors were within ± 0.50 D in 65% (OLCR), 70% (Scheimpflug), and 67% (SS-OCT) calculated by actual postACD versus 64% by predicted value.
    CONCLUSIONS: Substantial agreement was found in postACD measurements obtained from the SS-OCT, OLCR, and Scheimpflug devices, with a trend toward comparable refractive outcomes in the Olsen formula. Meanwhile, postACD measurements may be potentially superior for the additional enhancement of refractive outcomes.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    根据Amsler-Krumeich(AK)的不同定义,重新测试Pentacam参数在亚临床圆锥角膜(KC)和轻度KC眼检测中的性能,圆锥角膜(CLEK)的协同纵向评估,和ABCD系统。
    这项基于大学的横断面研究包括24只患有亚临床KC的眼睛,144眼轻度KC(基于101眼的AK,CLEK在28眼,和15只眼睛中的ABCD),70个控件。最薄点(TP)测厚仪的诊断能力,KISA%索引,次-次-次不对称,角膜像差,Pentacam指数,前/后高程,测速进展指数,Ambrósio-关系厚度(ARTmax),和Belin/Ambrósio增强的Ectasia显示分数(Df,Db,Dp,Dt,Da,和D-final)进行评估。
    ARTmax(83.3%的敏感性/74.3%的特异性)在区分亚临床KC和正常KC方面具有最高的能力,其次是TP测厚仪,Dt,还有Da.D-final在基于AK(98%/100%)和CLEK(97.4%/100%)描述的轻度KC诊断中显示出出色的敏感性/特异性。在轻度KC-ABCD组中,垂直不对称指数准确检测到所有轻度KC的眼睛和97.1%的对照组。
    这项研究指出,由于术语和分级标准重叠,亚临床和轻度KC的眼睛检测中存在灰色地带。Pentacam参数似乎在亚临床KC检测中具有适度的能力,表明需要额外的诊断方式。然而,使用Pentacam参数可以高精度诊断轻度KC的眼睛,尽管最强的参数可能会根据“轻度KC”的定义而有所不同。然而,亚临床和临床KC分类的统一和明确标准是KC诊断和治疗共识所必需的.
    To retest the performance of Pentacam parameters in the detection of eyes with subclinical keratoconus (KC) and mild KC based on different definitions from the Amsler-Krumeich (AK), Collaborative Longitudinal Evaluation of Keratoconus (CLEK), and ABCD systems.
    This cross-sectional university-based study comprised 24 eyes with subclinical KC, 144 eyes with mild KC (based on AK in 101 eyes, CLEK in 28 eyes, and ABCD in 15 eyes), and 70 controls. Diagnostic ability of the thinnest point (TP) pachymetry, KISA% index, inferior-superior asymmetry, corneal aberrations, Pentacam indices, front/back elevations, pachymetric progression index, Ambrósio-Relational Thickness (ARTmax), and Belin/Ambrósio Enhanced Ectasia Display scores (Df, Db, Dp, Dt, Da, and D-final) were evaluated.
    ARTmax (83.3% sensitivity/74.3% specificity) had the highest ability in distinguishing subclinical KC from normal, followed by TP pachymetry, Dt, and Da. D-final showed excellent sensitivity/specificity in mild KC diagnosis based on AK (98%/100%) and CLEK (97.4%/100%) descriptions. In the mild KC-ABCD group, index of vertical asymmetry accurately detected all eyes with mild KC and 97.1% of the controls.
    This study points out the gray zone in the detection of eyes with subclinical and mild KC due to overlapping terminology and grading criteria. Pentacam parameters seem to have modest capability in subclinical KC detection, indicating the necessity for additional diagnostic modalities. However, eyes with mild KC can be diagnosed with high accuracy using Pentacam parameters, although the strongest parameters may vary according to the definition of “mild KC.” Nevertheless, uniform and definitive criteria for subclinical and clinical KC classification are required for a diagnostic and therapeutic consensus in KC.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    为了确定圆锥角膜疾病随时间的一致变化,有必要根据用于监测角膜的设备的会间期变异性建立进展截止值.这项研究的目的是使用Scheimpflug断层扫描和前段光学相干断层扫描在不同严重程度的健康和角膜圆锥眼睛中分析角膜参数的准确性,以确定指示实际进展的临界值。使用Pentacam和CasiaSS-1000设备记录健康(20只眼)和角膜圆锥眼(轻度=16,中度=25和重度=20)的每个角膜的三次重复测量。重复2-3周后。K1,K2,最大前后角膜曲率测量,收集最薄位置(TCT)的角膜厚度。这两种设备的准确性都很好;然而,与Pentacam相比,Casia装置在所有组的所有参数中具有更好的可重复性和可重复性.轻度阶段的Pentacam和Casia的截止值较低(K1=0.50和0.37D;K2=0.51和0.37D;Kmax-A=1.24和0.65D;Kmax-P=0.38和0.17D;TCT=19.64和11.19µm)比重度阶段(K1=1.09和0.88D;CT=1.41和0.87D;KP=2.82max-A=和这些结果表明,圆锥角膜严重程度越大,必须发生的变化越大,才能被认为是真实的。
    To determine consistent change over time in keratoconus disease, it is necessary to establish progression cut-off values based on intersession variability of the device used to monitor the cornea. The aim of this study was to analyze the accuracy of corneal parameters using Scheimpflug tomography and anterior segment optical coherence tomography in healthy and keratoconic eyes of varying severity to determine the cut-off values that indicate real progression. Three repeated measurements of each cornea of healthy (20 eyes) and keratoconic eyes (mild = 16, moderate = 25 and severe = 20) were recorded using Pentacam and Casia SS-1000 devices, which were repeated 2-3 weeks later. K1, K2, maximal anterior and posterior keratometry, and corneal thickness at the thinnest location (TCT) were collected. The accuracy was excellent with both devices; however, the Casia device presented better repeatability and reproducibility in all parameters in all groups compared to the Pentacam. The cut-off of the Pentacam and Casia in the mild stage were lower (K1 = 0.50 and 0.37 D; K2 = 0.51 and 0.37 D; Kmax-A = 1.24 and 0.65 D; Kmax-P = 0.38 and 0.17 D; TCT = 19.64 and 11.19 µm) than that of the severe stage (K1 = 1.09 and 0.88 D; K2 = 1.41 and 0.87 D; Kmax-A = 2.74 and 2.15 D; Kmax-P = 0.82 and 0.22 D; TCT = 28.68 and 14.83 µm). These results show that the greater the keratoconus severity, the greater the change that must occur for it to be considered real.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    目的:评估角膜密度测定对便携式压平(Perkins)和反弹(iCareic100和PRO)眼压的影响。次要目标是评估各种角膜特性与原发性开角型青光眼(POAG)的严重程度之间是否存在关系。
    方法:研究了75例原发性开角型青光眼患者的75只眼,按严重程度分为3组:25轻度,25个中度和25个高级。每个参与者用Perkins压平眼压计(PAT)测量眼压(IOP)3次,高盛压平眼压计(GAT)的手持式版本,iCarePRO和iCareic100。然后计算平均值。PentacamHR的角膜地形图(Oculus,Wetzlar,德国)也在所有个人中进行了表演。
    结果:所有组的平均年龄和性别相当,光密度测量值(P>0.05)。轻度青光眼组平均视野缺损(MD)为2.85(±1.23)dB,中度组8.26(±1.90)dB,晚期组15.66(±3.46)dB。进行了三个多变量回归分析。第一次和第二次计算评估了iCareic100和PAT获得的IOP的影响,作为年龄的因变量,性别,青光眼亚组和整体样本中的CCT和平均角膜曲率(Km)。进行了第三次分析,以评估作为因变量的角膜密度测定法与青光眼组中上述角膜参数之间的关系。在第一个多元回归分析中,在POAG全球样本中,ic100反弹眼压与CCT之间存在统计学上的显着相关性(coef。0.117;IC[-0.21-(-0.01)];P=0.025)。在亚组分析中没有发现统计学上显著的相关性。在第二个多变量分析中,PAT和CCT之间没有发现显著的相关性,Km,年龄或性别(P>0.05)。在第三个分析中,光密度测定与所有青光眼亚组的年龄相关(P<0.001),与中度青光眼亚组的CCT相关(coef。-0.037;IC[-0.67-(-0.01)];P=0.021)。眼压测量似乎受角膜密度测定的影响最小,与轻度正线性相关(R=0.03)。眼压值与3个眼压计相似:PAT16.07(±3.18)mmHg,PRO16.27(±3.42)mmHg和ic10015.17(±4.28)mmHg。有,然而,与PAT相比,ic100(-0.89mmHg)明显低估了IOP(P=0.007)。
    结论:角膜光密度测定在青光眼严重程度组之间没有显着差异。CCT与角膜密度测定和年龄均呈正相关。未发现与角膜曲率或青光眼严重程度相关。角膜密度测定对眼压测量的影响似乎很弱,几乎没有临床相关性。
    OBJECTIVE: To evaluate the influence of corneal densitometry on portable applanation (Perkins) and rebound (iCare ic100 and PRO) tonometry. A secondary goal was to assess if there was a relationship between various corneal properties and the severity of primary open angle glaucoma (POAG).
    METHODS: Seventy-five eyes of 75 patients with primary open angle glaucoma were studied, divided by severity into 3 groups: 25 mild, 25 moderate and 25 advanced. Intraocular pressure (IOP) was measured 3 times in each participant with a Perkins applanation tonometer (PAT), a handheld version of the Goldman applanation tonometer (GAT), an iCare PRO and an iCare ic100. Mean values were then calculated. Corneal topography with the Pentacam HR (Oculus, Wetzlar, Germany) was also performed in all individuals.
    RESULTS: Mean age and sex were comparable in all groups, as were densitometry values (P>0.05). The mean visual field defect (MD) was 2.85 (±1.23) dB in the mild glaucoma group, 8.26 (±1.90) dB in the moderate group and 15.66 (±3.46) dB in the advanced group. Three multivariate regression analyses were performed. The first and second calculations assessed the effect of IOP obtained with iCare ic100 and PAT as dependent variables with age, sex, CCT and mean keratometry (Km) within the glaucoma subgroups and the global sample. The third analysis was carried out to assess the relationship between corneal densitometry as the dependant variable and the aforementioned corneal parameters among the glaucoma groups. In the first multivariate regression analysis, a statistically significant correlation was found between ic100 rebound tonometry and CCT in the POAG global sample (coef. 0.117; IC [-0.21-(-0.01)]; P=0.025). No statistically significant correlation was found in the subgroup analyses. In the second multivariate analysis, no significant correlation was found between PAT and CCT, Km, age or sex (P>0.05). In the third analysis, densitometry was correlated with age in all glaucoma subgroups (P<0.001) and with CCT in the moderate glaucoma subgroup (coef. -0.037; IC [-0.67-(-0.01)]; P=0.021). Tonometry appeared to be minimally influenced by corneal densitometry, with a mild positive linear correlation seen (R=0.03). IOP values were similar with 3 of the tonometers: PAT 16.07 (±3.18) mmHg, PRO 16.27 (±3.42) mmHg and ic100 15.17 (±4.28) mmHg. There was, however, a significant underestimation of IOP with ic100 (-0.89mmHg) compared to PAT (P=0.007).
    CONCLUSIONS: Corneal densitometry did not show significant differences between glaucoma severity groups. A positive correlation was seen with CCT and both corneal densitometry and age. No correlation was found with keratometry or severity of glaucoma. The influence of corneal densitometry on IOP measurements appears weak, with little clinical relevance identified.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    目的:探讨角膜前后高阶像差(HOAs)在圆锥角膜(KC)和可疑圆锥角膜(SKC)检测中的应用价值。
    方法:回顾性研究,病例对照研究评估非直肠(正常)眼睛,SKC的眼睛,KC眼睛SiriusScheimpfug(CSO,意大利)分析用于测量角膜前后表面的HOA。灵敏度,特异性,计算受试者工作特征曲线下面积(AUC)。
    结果:分析中包括二百二十只眼(正常n=108,SKCn=42,KCn=70)。受试者工作特征(ROC)曲线分析显示,前角膜HOA参数具有很高的预测能力:均方根(RMS)总角膜HOA,RMS三叶草,和RMS昏迷以检测圆锥角膜(所有AUC&gt;0.9)。来自角膜前表面的RMSComa(3,±1)是区分可疑圆锥角膜和正常眼睛的能力最高的参数(AUC=0.922;截止值&gt;0.2)。所有后角膜HOA参数在SKC和正常眼之间的区别均不令人满意(AUC&lt;0.8)。然而,除RMS球面像差(AUC=0.846)外,它们检测KC的能力均优异,AUC为0.9。
    结论:角膜前后高阶像差可以区分圆锥角膜和正常眼,具有高度的确定性。在怀疑圆锥角膜疾病中,然而,只有前角膜HOAs,特别是类似昏迷的像差,是有价值的。角膜像差法在该病患病率高的人群中筛查圆锥角膜可能具有价值。
    To investigate the application of anterior and posterior corneal higher-order aberrations (HOAs) in detecting keratoconus (KC) and suspect keratoconus (SKC).
    A retrospective, case-control study evaluating non-ectatic (normal) eyes, SKC eyes, and KC eyes. The Sirius Scheimpfug (CSO, Italy) analyses was used to measure HOAs of the anterior and posterior corneal surfaces. Sensitivity, specificity, and area under the receiver operating characteristic curve (AUC) were calculated.
    Two-hundred and twenty eyes were included in the analysis (normal n = 108, SKC n = 42, KC n = 70). Receiver operating characteristic (ROC) curve analysis revealed a high predictive ability for anterior corneal HOAs parameters: the root mean square (RMS) total corneal HOAs, RMS trefoil, and RMS coma to detect keratoconus (AUC > 0.9 for all). RMS Coma (3, ±1) derived from the anterior corneal surface was the parameter with the highest ability to discriminate between suspect keratoconus and normal eyes (AUC = 0.922; cut-off > 0.2). All posterior corneal HOAs parameters were unsatisfactory in discriminating between SKC and normal eyes (AUC < 0.8 for all). However, their ability to detect KC was excellent with AUC of >0.9 for all except RMS spherical aberrations (AUC = 0.846).
    Anterior and posterior corneal higher-order aberrations can differentiate between keratoconus and normal eyes, with a high level of certainty. In suspect keratoconus disease, however, only anterior corneal HOAs, and in particular coma-like aberrations, are of value. Corneal aberrometry may be of value in screening for keratoconus in populations with a high prevalence of the disease.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    未经证实:70%的Fuchs内皮角膜营养不良(FECD)病例是由转录因子4基因(TCF4)的内含子三核苷酸重复扩增引起的。这项研究的目的是表征有(RE)和没有三核苷酸重复扩增(RE-)的FECD患者的角膜基底下神经丛和角膜雾霾,并评估这些参数与疾病严重程度的相关性。
    未经评估:横截面,单中心研究。
    UNASSIGNED:该研究包括29名受试者的52只眼,其FECD严重程度的Krachmer等级从1到6。29名受试者中有15名携带扩展的TCF4等位基因长度≥40个胞嘧啶-胸腺嘧啶-鸟嘌呤重复序列(RE)。
    UNASSIGNED:体内共聚焦显微镜评估角膜神经纤维长度(CNFL),角膜神经分支密度,角膜神经纤维密度(CNFD),和前角膜基质反向散射(haze);Scheimpflug断层摄影术密度测定法测量前的haze,中央,和后角膜层。
    未经评估:使用共聚焦显微镜,我们在RE+受试者的眼睛中检测到FECD严重程度与CNFL和CNFD之间呈负相关(分别为Spearmanρ=-0.45,P=0.029和ρ=-0.62,P=0.0015),但在RE-受试者的眼睛中没有。此外,在RE+受试者中,CNFD与扩增等位基因的重复长度呈负相关(Spearmanρ=-0.42,P=0.038)。我们发现RE和RE-组的前基质背向散射与严重程度之间呈正相关(分别为ρ=0.60,P=0.0023和ρ=0.44,P=0.024)。前部,中央,和后Scheimpflug密度测定测量值也与RE组和RE-组的严重程度呈正相关(分别为P=5.5×10-5、2.5×10-4和2.9×10-4,在合并分析中调整扩展状态后。然而,对于患有严重FECD(克拉赫默5级和6级)的患者,RE+组的后路光密度测量值高于RE-组(P<0.05)。
    UNASSIGNED:FECD角膜神经丢失支持将TCF4三核苷酸重复扩增障碍分类为神经退行性疾病。前面的阴霾,中央,后角膜与严重程度相关,无论基因型。角膜神经和角膜雾霾的定量评估可能有助于评估和监测RE患者的FECD疾病严重程度。
    UNASSIGNED: Seventy percent of Fuchs\' endothelial corneal dystrophy (FECD) cases are caused by an intronic trinucleotide repeat expansion in the transcription factor 4 gene (TCF4). The objective of this study was to characterize the corneal subbasal nerve plexus and corneal haze in patients with FECD with (RE+) and without the trinucleotide repeat expansion (RE-) and to assess the correlation of these parameters with disease severity.
    UNASSIGNED: Cross-sectional, single-center study.
    UNASSIGNED: Fifty-two eyes of 29 subjects with a modified Krachmer grade of FECD severity from 1 to 6 were included in the study. Fifteen of the 29 subjects carried an expanded TCF4 allele length of ≥ 40 cytosine-thymine-guanine repeats (RE+).
    UNASSIGNED: In vivo confocal microscopy assessments of corneal nerve fiber length (CNFL), corneal nerve branch density, corneal nerve fiber density (CNFD), and anterior corneal stromal backscatter (haze); Scheimpflug tomography densitometry measurements of haze in anterior, central, and posterior corneal layers.
    UNASSIGNED: Using confocal microscopy, we detected a negative correlation between FECD severity and both CNFL and CNFD in the eyes of RE+ subjects (Spearman ρ = -0.45, P = 0.029 and ρ = -0.62, P = 0.0015, respectively) but not in the eyes of RE- subjects. Additionally, CNFD negatively correlated with the repeat length of the expanded allele in the RE+ subjects (Spearman ρ = -0.42, P = 0.038). We found a positive correlation between anterior stromal backscatter and severity in both the RE+ and RE- groups (ρ = 0.60, P = 0.0023 and ρ = 0.44, P = 0.024, respectively). The anterior, central, and posterior Scheimpflug densitometry measurements also positively correlated with severity in both the RE+ and RE- groups (P = 5.5 × 10-5, 2.5 × 10-4, and 2.9 × 10-4, respectively, after adjusting for the expansion status in a pooled analysis. However, for patients with severe FECD (Krachmer grades 5 and 6), the posterior densitometry measurements were higher in the RE+ group than in the RE- group (P < 0.05).
    UNASSIGNED: Loss of corneal nerves in FECD supports the classification of the TCF4 trinucleotide repeat expansion disorder as a neurodegenerative disease. Haze in the anterior, central, and posterior cornea correlate with severity, irrespective of the genotype. Quantitative assessments of corneal nerves and corneal haze may be useful to gauge and monitor FECD disease severity in RE+ patients.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

公众号