METHODS: Patients were recruited, and the ipsilateral sequential method was used for blood pressure measurement according to the standard. The validation results were assessed following the protocol and the Bland-Altman scatterplot was used to show the difference between the test device and reference invasive blood pressure results.
RESULTS: A total of 71 patients were included in the study, with 35 and 36 patients for each iFAST and iCUFS mode, respectively. The validation results showed an average device-reference difference of -3.27 ± 5.60 mmHg for SBP and -0.09 ± 6.10 mmHg for DBP for the iFAST mode, and -2.04 ± 5.55 mmHg for SBP and -0.79 ± 5.86 mmHg for DBP for the iCUFS mode, respectively, which passed the criteria of the ISO 81060-2 : 2018 in adults, adolescents or children population for both SBP and DBP.
CONCLUSIONS: The noninvasive blood pressure monitoring function of the EDAN elite V5 patient monitor passed all the requirements of ISO 81060-2:2018 and can be recommended for clinical use in adults, adolescents, or children.
方法:招募患者,按标准采用同侧序贯法测定血压。验证结果按照方案进行评估,Bland-Altman散点图用于显示测试设备和参考有创血压结果之间的差异。
结果:本研究共纳入71例患者,每种iFAST和iCUFS模式有35和36名患者,分别。验证结果显示,iFAST模式下SBP和DBP的平均设备参考差异为-3.27±5.60mmHg,为-0.09±6.10mmHg。对于iCUFS模式,SBP为-2.04±5.55mmHg,DBP为-0.79±5.86mmHg,分别,在成年人中通过了ISO81060-2:2018的标准,SBP和DBP的青少年或儿童人群。
结论:EDAN精英V5患者监护仪的无创血压监测功能通过了ISO81060-2:2018的所有要求,可推荐用于成人临床使用,青少年,或孩子。