Fatigue testing

疲劳试验
  • 文章类型: Journal Article
    暂无摘要。
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    目的:分析体内回收的PEEK棒的特性,特别是它们的磨损和变形,生物降解性,组织相容性,和机械性能。
    方法:从翻修手术中取出6根PEEK棒以及假体周围组织。使用Micro-CT评估回收的PEEK棒的表面损伤和内部变化,而光学和电子显微镜用于确定假体周围组织的任何组织学变化。从病历中收集患者病史。两个完整的和回收的PEEK杆用于通过对脊柱构造的正弦载荷进行疲劳测试分析。
    结果:所有植入物都显示出螺杆-杆界面周围塑性变形的证据,而PEEK棒的内部结构没有变化,没有可见的空隙或裂纹。检查通过光学和电子显微镜捕获的图像表明,与螺旋杆界面相比,PEEK杆周围巨噬细胞的吞噬作用不那么严重。能谱分析的结果表明,PEEK棒周围组织元素的分布与正常组织没有显着差异。在疲劳测试期间,发现回收的PEEK棒经过136万次测试后破裂,而完整的PEEK棒完成了500万次疲劳测试,没有任何故障。
    结论:PEEK棒表现出令人满意的生物相容性,耐腐蚀性,化学稳定性,和机械性能。然而,据观察,在螺母和杆之间的连接处的压痕表现出相对较弱的强度,使其容易破损。作为预防措施,建议用反向扳手固定螺母,施加预设扭矩以防止过度拧紧。
    OBJECTIVE: To analyze the characteristics of PEEK rods retrieved in vivo, specifically their wear and deformation, biodegradability, histocompatibility, and mechanical properties.
    METHODS: Six PEEK rods were retrieved from revision surgeries along with periprosthetic tissue. The retrieved PEEK rods were evaluated for surface damage and internal changes using Micro-CT, while light and electron microscopy were utilized to determine any histological changes in periprosthetic tissues. Patient history was gathered from medical records. Two intact and retrieved PEEK rods were used for fatigue testing analysis by sinusoidal load to the spinal construct.
    RESULTS: All implants showed evidence of plastic deformation around the screw-rod interface, while the inner structure of PEEK rods appeared unchanged with no visible voids or cracks. Examining images captured through light and electron microscopy indicated that phagocytosis of macrophages around PEEK rods was less severe in comparison to the screw-rod interface. The results of an energy spectrum analysis suggested that the distribution of tissue elements around PEEK rods did not differ significantly from normal tissue. During fatigue testing, it was found that the retrieved PEEK rods cracked after 1.36 million tests, whereas the intact PEEK rods completed 5 million fatigue tests without any failure.
    CONCLUSIONS: PEEK rods demonstrate satisfactory biocompatibility, corrosion resistance, chemical stability, and mechanical properties. Nevertheless, it is observed that the indentation at the junction between the nut and the rod exhibits relatively weak strength, making it susceptible to breakage. As a precautionary measure, it is recommended to secure the nut with a counter wrench, applying the preset torque to prevent overtightening.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    目的:髂股静脉阻塞性疾病可导致显著,潜在的衰弱症状会对生活质量产生负面影响。与动脉疾病不同,深静脉疾病患者的中位年龄明显较低,因此,对长期支架通畅的需求变成了几十年而不是几年的问题。此外,髂股病变通常需要在腹股沟韧带上放置支架。此类支架会受到腿部运动产生的动态应力以及相关的设备疲劳问题。导致支架断裂。这项研究的目的是描述一种体外50年支架疲劳测试方法,旨在评估动态应力引起的设备断裂的耐久性。
    方法:通过文献综述,尸体研究,和计算机建模,经证实,最具挑战性的负荷是跨腹股沟韧带的髋关节屈曲.这发生在患者在就座和站立位置之间调整时。在这项体外实验研究中,在工作台上有效地模拟了腹股沟韧带的坐立髋关节屈曲。
    结果:在挑战参数下测试时,可靠地发现髋关节屈曲导致非静脉镍钛诺支架骨折.然而,专用的自膨胀镍钛诺静脉支架,为提高耐用性而设计,在工作台上进行了长达50年的模拟加载,其中15%(3/20)的支架在50年时发生骨折,与35%(14/40)的非静脉支架测试至1.4年的骨折相比;由于持续时间不匹配,因此未进行统计测试,目的是证明测试方法)。
    结论:提出的疲劳测试方法是评估打算用于静脉使用的支架的耐久性的合适方法。通过体外髋关节屈曲测试,静脉支架与非静脉支架相比表现出优异的抗疲劳性。
    OBJECTIVE: Iliofemoral venous obstructive disease can result in significant, potentially debilitating symptoms that can negatively affect quality of life. Unlike arterial disease, patients with deep venous disease have a significantly lower median age, therefore the need for long term stent patency becomes a matter of decades rather than years. Furthermore, iliofemoral lesions frequently require stent placement across the inguinal ligament. Such stents are subject to dynamic stress from leg movement and associated concerns for device fatigue, resulting in stent fracture. The aim of this study was to describe an in vitro 50 year stent fatigue test method designed to assess durability against dynamic stress induced device fracture.
    METHODS: Through literature review, cadaver studies, and computer modelling, the most challenging loading was confirmed to be hip flexion across the inguinal ligament. This occurs when the patient adjusts between a seated and standing position. Sit to stand hip flexion at the inguinal ligament was effectively simulated on the bench in this in vitro experimental study.
    RESULTS: When tested under challenge parameters, hip flexion was reliably found to cause fractures in non-venous nitinol stents. However, a dedicated self expanding nitinol venous stent, engineered for improved durability, underwent up to 50 years of simulated loading on the bench with 15% (3/20) of stents experiencing fractures at 50 years, compared with fractures in 35% (14/40) of non-venous stents tested to 1.4 years; no statistical testing was performed as durations do not match and the objective was to demonstrate the test method.
    CONCLUSIONS: The presented fatigue test method is a suitable approach for evaluating the durability of stents intended for venous use. Venous stents demonstrated superior fatigue resistance compared with non-venous stents via in vitro hip flexion testing.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    目的:预测二硅酸锂玻璃陶瓷制成的前冠的寿命和断裂概率(IPSe.maxCAD,LD,IvoclarVivadent,列支敦士登)和含氧化锆的硅酸锂玻璃陶瓷(CeltraDuo,ZLS,DentsplySirona,美国)在自行车载荷下。
    方法:进行三点弯曲测试以测量粘弹性参数。这些参数用于计算结晶后前冠的残余应力。在接下来的分析中,计算了前冠的循环载荷。基于这种组合应力状态(残余应力和由于外部循环载荷引起的应力状态),使用CARES/Life软件计算前冠的生命周期和骨折概率.最后,进行了疲劳实验,以比较和验证计算结果。
    结果:尽管可以使用此方法对两种材料的寿命进行可靠的定性比较,与使用实验确定的疲劳参数的疲劳测试结果相比,两种材料计算的前冠断裂概率非常低。为了达到与实验结果的良好对应,两种材料的SCG指数n应通过0.38的相关因子进行修改。
    结论:使用这种改进的计算策略,可以密切预测牙科玻璃陶瓷耗时的疲劳测试结果。该方法可以集成到新型玻璃陶瓷材料的开发过程中,以节省时间和成本。
    OBJECTIVE: To predict the lifetime and fracture probability of anterior crowns made of a lithium disilicate glass-ceramic (IPS e.max CAD, LD, Ivoclar Vivadent, Liechtenstein) and a zirconia-containing lithium silicate glass-ceramic (Celtra Duo, ZLS, Dentsply Sirona, USA) under cycling loading.
    METHODS: Three-point bending tests were conducted to measure the viscoelastic parameters. These parameters are used to compute the residual stresses of the anterior crown after crystallization. In the next analysis, the cyclic loading on the anterior crown was calculated. Based on this combined stress state (residual stress and stress state due to external cyclic loading), the life cycle and fracture probability of the anterior crown was calculated using the CARES/Life software. Finally, fatigue experiments were carried out to compare and validate the results of the computations.
    RESULTS: Although a sound qualitative comparison of the lifetime of both materials can be done using this methodology, the calculated fracture probability of the anterior crown for both materials was very low in comparison with the fatigue test results using the fatigue parameters determined from the experiments. In order to achieve good correspondence with the experimental results, the SCG exponent n for both materials should be modified by a correlation factor of 0.38.
    CONCLUSIONS: Using this modified computational strategy, the results of the time-consuming fatigue tests for dental glass-ceramics can be closely predicted. This methodology can be integrated into the development process of new glass-ceramic materials in order to save time and costs.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    目的:种植体周围的骨吸收会影响种植体基台复合体(IAC)的阻力。本体外研究旨在评估表现出不同程度的骨损失和直径的植入物对静态疲劳的稳定性。
    方法:90个内部圆锥形连接的植入物,具有3种不同的植入物直径(3.3mm(I33),3.8mm(I38),和4.3毫米(I43))和3个模拟骨丢失设置(1.5毫米(I_15),3.0mm(I_30),和4.5毫米(I_45)(n=10))嵌入和标准桥台安装。所有标本均人工老化(1,200,000个周期,50N,同时进行热循环),并随后进行了断裂载荷测试。为了进行统计分析,科尔莫戈罗夫-斯米尔诺夫试验,Kruskal-Wallis测试,采用Mann-WhitneyU检验(p<0.05)。
    结果:所有试样都经受了人工老化而没有损坏。平均失效值为382.1(±59.2)N(I3315),347.0(±35.7)N(I3330),315.9N(±30.9)(I3345),531.4(±36.2)N(I3815),514.5(±40.8)N(I3830),477.9(±26.3)N(I3845),710.1(±38.2)N(I4315),697.9(±65.2)N(I4330),和662.2N(±45.9)(I4345)。当骨丢失倾向时,所有组的IAC的稳定性均降低。仅仅,I43亚组的失效载荷值无显著差异.
    结论:较大的植入物直径和植入物周围较小的圆形骨丢失导致IAC的稳定性更高。植入物直径越小,稳定性受圆周骨水平的影响越大。
    结论:保留牙髓骨水平对于确保具有锥形界面的植入物系统的生物力学可持续性很重要。在植入物计划过程和恢复性考虑过程中,考虑植入物周围最终骨质流失的影响似乎是明智的。
    OBJECTIVE: Bone resorption around implants could influence the resistance of the implant abutment complex (IAC). The present in vitro study aimed to assess the stability to static fatigue of implants presenting different levels of bone losses and diameters.
    METHODS: Ninety implants with an internal conical connection with 3 different implant diameters (3.3 mm (I33), 3.8 mm (I38), and 4.3 mm (I43)) and 3 simulated bone loss settings (1.5 mm (I_15), 3.0 mm (I_30), and 4.5 mm (I_45) (n = 10)) were embedded and standard abutments were mounted. All specimens were artificially aged (1,200,000 cycles, 50 N, simultaneous thermocycling) and underwent subsequently load-to-fracture test. For statistical analysis, Kolmogorov-Smirnov test, Kruskal-Wallis test, and Mann-Whitney U test (p < 0.05) were applied.
    RESULTS: All test specimens withstood the artificial aging without damage. The mean failure values were 382.1 (± 59.2) N (I3315), 347.0 (± 35.7) N (I3330), 315.9 N (± 30.9) (I3345), 531.4 (± 36.2) N (I3815), 514.5 (± 40.8) N (I3830), 477.9 (± 26.3) N (I3845), 710.1 (± 38.2) N (I4315), 697.9 (± 65.2) N (I4330), and 662.2 N (± 45.9) (I4345). The stability of the IACs decreased in all groups when bone loss inclined. Merely, the failure load values did not significantly differ among subgroups of I43.
    CONCLUSIONS: Larger implant diameters and minor circular bone loss around the implant lead to a higher stability of the IAC. The smaller the implant diameter was, the more the stability was affected by the circumferential bone level.
    CONCLUSIONS: Preserving crestal bone level is important to ensure biomechanical sustainability at implant systems with a conical interface. It seems sensible to take the effect of eventual bone loss around implants into account during implant planning processes and restorative considerations.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    在牙科植入物可能发生的并发症中,任何植入物组件的骨折都是相对罕见但临床相关的问题。由于它们的机械特性,小直径植入物发生此类并发症的风险较高。这项实验室和有限元方法(FEM)研究的目的是比较在标准静态和动态条件下直径为2.9和3.3mm的植入物与圆锥形连接的力学行为,遵循国际标准化组织(ISO)14801:2017。进行了有限元分析,以比较在300-N下测试的植入物系统上的应力分布,30°倾斜力。使用2kN的测压元件进行静态测试;相对于种植体-基牙轴在30°处对实验样品施加力,一个5.5毫米的手臂。疲劳试验是在载荷减小的情况下进行的,在2Hz频率下,直到3个标本在200万次循环后没有任何损伤存活。在有限元分析中,桥台的出现轮廓产生了最大的应力区域,对于直径为2.9和3.3mm的植入物复合体,最大应力为5829和5480MPa,分别。平均最大载荷导致2.9mm直径的植入物为360N,3.3mm直径的植入物为370N。记录的疲劳极限为220和240N,分别。尽管3.3毫米直径的植入物效果更好,测试植入物之间的差异在临床上可以认为可以忽略不计。这可能是由于种植体-基牙连接的圆锥形设计,据报道,植入物颈部区域的应力值较低,从而增加了抗断裂性。
    Among the complications that can occur at dental implants, the fracture of any implant component is a relatively infrequent but clinically relevant problem. Because of their mechanical characteristics, small diameter implants are at higher risk of such complication. The aim of this laboratory and finite element method (FEM) study was to compare the mechanical behavior of a 2.9- and 3.3-mm-diameter implant with a conical connection under standard static and dynamic conditions, following the International Organization for Standardization (ISO) 14801:2017. Finite element analysis was performed to compare the stress distribution on the tested implant systems under a 300-N, 30° inclined force. Static tests were performed with a load cell of 2 kN; the force was applied on the experimental samples at 30° with respect to the implant-abutment axis, with an arm of 5.5 mm. Fatigue tests were performed with decreasing loads, at 2-Hz frequency, until 3 specimens survived without any damage after 2 million cycles. The emergence profile of the abutment resulted the most stressed area in finite element analysis, with a maximum stress of 5829 and 5480 MPa for 2.9- and 3.3-mm-diameter implant complex, respectively. The mean maximum load resulted in 360 N for 2.9-mm-diameter and 370 N for 3.3-mm-diameter implants. The fatigue limit was recorded to be 220 and 240 N, respectively. Despite the more favorable results of 3.3-mm-diameter implants, the difference between the tested implants could be considered clinically negligible. This is probably due to the conical design of the implant-abutment connection, which has been reported to present low stress values in the implant neck region, thus increasing the fracture resistance.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    目的:这项体外研究旨在研究长期性能,稳定性,和整体式混合基牙冠的断裂模式,以及不同材料对种植体-基台界面(IAI)的影响。
    方法:用3Y-TZP氧化锆(ZY3)制造了八十个整体混合基牙冠,“梯度技术”氧化锆(ZY35),5Y-TZP氧化锆(ZY5),二硅酸锂陶瓷(LDS),氧化锆增强硅酸锂陶瓷(ZLS),聚合物渗透陶瓷网(MHY),聚甲基丙烯酸甲酯(PMA),和3D打印混合复合材料(PHC)(每种材料n=10)。八十种植入物(CamlogProgressive-Line,直径:3.8mm)根据ISO标准14801嵌入,并安装了牙冠。人工老化后(1.2×106次循环,50N,热循环),将完整的样品在万能试验机中30°离轴加载直至失效。
    结果:PHC组中的七个标本在人工老化过程中失败,其他人都活了下来.基于单因素方差分析和平均骨折载荷值的Dunnett检验(p<0.05),有两个亚组。第一个包括Z3Y,ZY35,Z5Y,LDS,平均断裂载荷在499.4和529.7N之间,而第二个包括ZLS,MHY,和PMA,值在346.2-416.0N范围内。ZY3,ZY35,ZY5和LDS在骨折负荷测试后表现出可见的植入物肩部变形,尺寸不同。
    结论:由LDS制成的王冠,ZLS,MHY,和PMA可能作为潜在的应力破坏者,并防止IAIs可能变形。进一步的临床研究需要评估这些材料是否也在体内承受相关负荷。
    Purpose This in vitro study aimed to investigate the long-term performance, stability, and fracture mode of monolithic hybrid abutment crowns, and the effect of different materials on the implant-abutment interface (IAI).Methods Eighty monolithic hybrid abutment crowns luted on titanium bases were manufactured from 3Y-TZP zirconia (ZY3), \"Gradient Technology\" zirconia (ZY35), 5Y-TZP zirconia (ZY5), lithium disilicate ceramic (LDS), zirconia-reinforced lithium silicate ceramic (ZLS), polymer-infiltrated ceramic network (MHY), polymethylmethacrylate (PMA), and 3D-printed hybrid composite (PHC) (n = 10 for each material). Eighty implants (Camlog Progressive-Line, diameter: 3.8 mm) were embedded in accordance with ISO standard 14801, and crowns were mounted. After artificial aging (1.2 × 106 cycles, 50 N, thermocycling), intact specimens were loaded 30° off-axis in a universal testing machine until failure.Results Seven specimens in the PHC group failed during artificial aging, and all the others survived. There were two subgroups based on the one-way analysis of variance and Dunnett\'s test (P < 0.05) of the mean fracture load values. The first comprised Z3Y, ZY35, Z5Y, and LDS, with mean fracture loads between 499.4 and 529.7 N, while the second included ZLS, MHY, and PMA, with values in the 346.2-416.0 N range. ZY3, ZY35, ZY5, and LDS exhibited irreversible, visible deformations of the implant shoulders with varying dimensions after load-to-fracture tests.Conclusions Crowns made of LDS, ZLS, MHY, and PMA may act as potential stress breakers, and prevent possible deformation at IAIs. Further clinical studies need to assess if these materials also withstand relevant loads in-vivo.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Randomized Controlled Trial
    这项研究调查了一种称为缺血调节(IC)的新疗法是否可以改善间歇性跛行的外周血管疾病患者的步行能力和下肢肌肉表现。
    纳入43例跛行患者,接受IC或ICSham治疗2周,控制,双盲,前瞻性研究。IC会话涉及五个周期,将血压袖带交替5分钟充气至225mmHg(ICSham为25mmHg)和5分钟放气,在受影响的下肢的大腿周围。
    跛行发作时间的变化没有差异(Δ=114±212sIC与104±173sICSham;p=0.67)或峰值步行时间(Δ=42±139sIC与IC和ICSham组之间为12±148sICSham;p=0.35)。在膝盖的水平,IC组的参与者完成了更多的工作(Δ=3,029±4,999JIC与345±2,863JICSham;p=0.03),并显示出更长时间的肌肉疲劳(Δ=147±221sIC与-27±236sICSham;p=0.01)。
    在跛行患者中,IC改善了膝盖的总功和疲劳时间,但没有改变步行参数。
    This study investigated whether a novel therapy called ischemic conditioning (IC) improves walking capacity and lower extremity muscle performance in patients with peripheral vascular disease who experience intermittent claudication.
    Forty-three patients with claudication were enrolled and received either IC or IC Sham for 2 weeks in this randomized, controlled, double-blinded, prospective study. IC sessions involved five cycles of alternating 5-min inflations of a blood pressure cuff to 225 mm Hg (25 mm Hg for IC Sham) and 5-min deflations, around the thigh of the affected lower extremity.
    There was no difference in the change in claudication onset time (Δ = 114 ± 212 s IC vs. 104 ± 173 s IC Sham; p = 0.67) or peak walking time (Δ = 42 ± 139 s IC vs. 12 ± 148 s IC Sham; p = 0.35) between the IC and IC Sham groups. At the level of the knee, participants in the IC group performed more work (Δ = 3,029 ± 4,999 J IC vs. 345 ± 2,863 J IC Sham; p = 0.03) and displayed a greater time to muscle fatigue (Δ = 147 ± 221 s IC vs. -27 ± 236 s IC Sham; p = 0.01).
    In patients with claudication, IC improved total work performed and time to fatigue at the knee but did not change walking parameters.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

    求助全文

  • 文章类型: Journal Article
    使用交替的正交缝线和环形旋缝技术可以增强缝线固定。
    有人假设这种新颖的多平面,与标准环状旋针(WS)和Krackow针法(KS)相比,垂直环状旋针(MP)技术将改善生物力学特性。
    对照实验室研究。
    将30例胫骨前腱随机分为3组,每组10例。肌腱被固定在一个定制的夹子上,另一端用三种技术中的一种缝合:KS,WS,或新颖的MP。MP以从右到左开始的交替正交投掷进行,然后从前到后,从左到右,回到前面。每种技术都使用了4遍编号。2FiberWire间隔5mm,与钢筋束末端相距10mm。肌腱被预加载到5N,以100mm/min的速度预张至50N,持续3个周期,回到5N1分钟,以200mm/min从5到100N循环100次,然后以20mm/min的速度加载到失效状态。在预张紧和循环后记录伸长率,并在整个缝合线-肌腱界面以及从缝合线-肌腱界面的底部到缝合线上的标记(结构伸长率)进行测量。进行了单向方差分析,在适当的时候使用Bonferroni事后分析。
    3种技术在横截面积或刚度上没有差异。与MP(270.76±39.36N)和KS(298.90±25.94N)(两者的P≤0.001)相比,WS(183.33±57.44N)的极限载荷均较小。与WS和MP相比,KS的构造伸长率较小,从预张紧到循环结束时测得(P<.001)。在测试过程中,所有3种技术在缝合线-肌腱界面处的长度(缩短)均减少。与KS相比,WS的缝合肌腱界面缩短更多(P=.006)。
    KS看起来优越,因为它最大限度地提高强度,同时最大限度地减少构造伸长或接枝缩短。MP技术的极限载荷大于标准技术的极限载荷,但与KS技术的极限载荷没有显着差异。
    优选KS。如果使用WS,多平面,应考虑垂直通过。
    UNASSIGNED: Using alternating orthogonal suture throws with the looped whipstitch technique may allow enhanced suture fixation.
    UNASSIGNED: It was hypothesized that this novel multiplanar, perpendicular looped whipstitch (MP) technique would have improved biomechanical properties compared with the standard looped whipstitch (WS) and Krackow stitch (KS).
    UNASSIGNED: Controlled laboratory study.
    UNASSIGNED: A total of 30 cadaveric tibialis anterior tendons were randomly assigned into 3 groups of 10. Tendons were secured to a custom clamp, and the other end was sutured using 1 of 3 techniques: the KS, WS, or novel MP. The MP was performed with alternating orthogonal throws starting right to left, then front to back, left to right, and back to front. Each technique used 4 passes of No. 2 FiberWire spaced 5 mm apart and ending 10 mm from the tendon end. Tendons were preloaded to 5 N, pretensioned to 50 N at 100 mm/min for 3 cycles, returned to 5 N for 1 minute, cycled from 5 to 100 N at 200 mm/min for 100 cycles, and then loaded to failure at 20 mm/min. Elongation was recorded after pretensioning and cycling and was measured both across the suture-tendon interface and from the base of the suture-tendon interface to markings on the suture limbs (construct elongation). One-way analyses of variance were performed, with Bonferroni post hoc analysis when appropriate.
    UNASSIGNED: There were no differences in cross-sectional area or stiffness among the 3 techniques. The ultimate load for WS (183.33 ± 57.44 N) was less compared with both MP (270.76 ± 39.36 N) and KS (298.90 ± 25.94 N) (P ≤ .001 for both). There was less construct elongation for KS compared with WS and MP for total displacement, measured from pretensioning to the end of cycling (P < .001). All 3 techniques saw a decrease in length (shortening) at the suture-tendon interface during testing. There was more shortening at the suture-tendon interface for WS compared with KS (P = .006).
    UNASSIGNED: The KS appears superior, as it maximized strength while minimizing construct elongation or graft shortening. The ultimate load of the MP technique was greater than that of the standard technique but not significantly different from that of the KS technique.
    UNASSIGNED: The KS is preferred. If using a WS, multiplanar, perpendicular passes should be considered.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    关节反复负荷引起的软结缔组织的机械磨损是退行性关节病的主要因素,因此需要准确表征关节结构中的磨损的方法。这里,我们评估了使用结构光3D光学扫描系统和建模软件量化和可视化体外关节负荷下整个人体半月板体积损失的准确性。使用具有已知磨损体积的3D打印弯月面复制品,我们确定这种新颖的成像方法的平均准确度约为13mm3,当测量半月板0.2cm3(豌豆大小)的体积变化时,平均误差小于7%。然后将成像方法应用于在单个尸体膝盖经受一百万次受控关节负荷循环的四个时间点测量整个人半月板的体外磨损。内侧和外侧半月板达到每百万周期0.72cm3和0.34cm3的稳态体积减少,分别。线性磨损深度的比色法图显示,内侧和外侧半月板的后部区域都有很高的磨损和变形。第一次,这项研究开发了一种方法,以准确地表征体积损失在整个半月板受到体外关节负荷。这种3D扫描方法为研究人员提供了一种新的研究工具来研究半月板的机械磨损和关节退化,和其他软结缔组织。
    The mechanical wear and tear of soft connective tissue from repetitive joint loading is a primary factor in degenerative joint disease, and therefore methods are needed to accurately characterize wear in joint structures. Here, we evaluate the accuracy of using a structured light 3D optical scanning system and modeling software to quantify and visualize volume loss in whole human meniscus subjected to in vitro joint loading. Using 3D printed meniscus replicas with known wear volumes, we determined that this novel imaging method has a mean accuracy of approximately 13 mm3, corresponding to a mean error of less than 7% when measuring meniscal volumetric changes of 0.2 cm3 (size of a pea). The imaging method was then applied to measure the in vitro wear of whole human menisci at four time points when a single cadaveric knee was subjected to one million cycles of controlled joint loading. The medial and lateral menisci reached steady state volumetric reductions of 0.72 cm3 and 0.34 cm3 per million cycles, respectively. Colorimetric maps of linear wear depth revealed high wear and deformation in the posterior regions of both the medial and lateral menisci. For the first time, this study has developed a method to accurately characterize volume loss in whole meniscus subjected to in vitro joint loading. This 3D scanning method offers researchers a new investigative tool to study mechanical wear and joint degeneration in meniscus, and other soft connective tissues.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

公众号