lens dose

透镜剂量
  • 文章类型: Journal Article
    A short curtain that improves on the low versatility of existing long curtains was developed as a dedicated radiation protective device for the over-table tube fluorographic imaging units. The effect of this short curtain in preventing cataracts was then examined. First, the physician lens dose reduction rate was obtained at the position of the lens. Next, the reduction rate in the collective equivalent dose for the lens of the physician\'s eye was estimated. The results showed that lens dose reduction rates with the long curtain and the short curtain were 88.9% (literature-based value) and 17.6%, respectively, higher with the long curtain. In our hospital, the reduction rate in the collective equivalent dose for the lens of the physician\'s eye was 9.8% and 17.6% with a procedures mixture, using the long curtain where technically possible and no curtain in all other procedures, and the short curtain in all procedures, respectively, higher with the short curtain. Moreover, a best available for curtains raised the reduction rate in the collective equivalent dose for the lens of the physician\'s eye a maximum of 25.5%. By introducing the short curtain, it can be expected to have an effect in preventing cataracts in medical staff.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Sci-hub)

       PDF(Pubmed)

  • 文章类型: Journal Article
    OBJECTIVE: In cerebral angiography, for diagnosis and interventional neuroradiology, cone-beam computed tomography (CBCT) scan is frequently performed for evaluating brain parenchyma, cerebral hemorrhage, and cerebral infarction. However, the patient\'s eye lens is more frequently exposed to excessive doses in these scans than in the previous angiography and interventional neuroradiology (INR) procedures. Hence, radioprotection for the lenses is needed. This study selects the most suitable eye lens protection material for CBCT from among nine materials by evaluating the dose reduction rate and image quality.
    METHODS: To determine the dose reduction rate, the lens doses were measured using an anthropomorphic head phantom and a real-time dosimeter. For image quality assessment, the artifact index was calculated based on the pixel value and image noise within various regions of interest in a water phantom.
    RESULTS: The protective materials exhibited dose reduction; however, streak artifacts were observed near the materials. The dose reduction rate and the degree of the artifact varied significantly depending on the protective material. The dose reduction rates were 14.6%, 14.2%, and 26.0% when bismuth shield: normal (bismuth shield in the shape of an eye mask), bismuth shield: separate (two separate bismuth shields), and lead goggles were used, respectively. The \"separate\" bismuth shield was found to be effective in dose reduction without lowering the image quality.
    CONCLUSIONS: We found that bismuth shields and lead goggles are suitable protective devices for the optimal reduction of lens doses.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Sci-hub)

  • 文章类型: Journal Article
    OBJECTIVE: This study aimed to (a) develop a contact lens-type ocular in vivo dosimeter (CLOD) that can be worn directly on the eye and (b) assess its dosimetric characteristics and biological stability for radiation therapy.
    METHODS: The molder of a soft contact lens was directly used to create the dosimeter, which included a radiation-sensitive component - an active layer similar to a radiochromic film - to measure the delivered dose. A flatbed scanner with a reflection mode was used to measure the change in optical density due to irradiation. The sensitivity, energy, dose rate, and angular dependence were tested, and the uncertainty in determining the dose was calculated using error propagation analysis. Sequential biological stability tests, specifically, cytotoxicity and ocular irritation tests, were conducted to ensure the safe application of the CLOD to patients.
    RESULTS: The dosimeter demonstrated high sensitivity in the low dose region, and the sensitivity linearly decreased with the dose. The responses obtained for the 10 and 15 MV photon beams were 1.7% and 1.9% higher compared to the 6 MV photon beam. A strong dose rate dependence was not obtained for the CLOD. Angular dependence was observed from 90° to 180° with a difference in response from 1% to 2%. The total uncertainty in error propagation analysis decreased as a function of the dose in the red channel. For a dose range of 0 to 50 cGy, the total uncertainties for 5, 10, and 50 cGy were 14.2%, 8.9%, and 5%, respectively. Quantitative evaluation using the MTT (3-(4,5-dimethylthiazol-2-yl)-2,5-diphenyltetrazolium bromide) method presented no cytotoxicity. Further, no corneal opacity, iris reaction, or conjunctival inflammation was observed.
    CONCLUSIONS: The CLOD is the first dosimeter that can be worn close to the eye. The results of cytotoxicity and irritation tests indicate that it is a stable medical device. The evaluation of dose characteristics in open field conditions shows that the CLOD can be applied to an in vivo dosimeter in radiotherapy.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Sci-hub)

  • 文章类型: Journal Article
    Cataracts are the leading cause of blindness and visual disability worldwide. Of the known contributing factors to this condition, ionising radiation is considered the primary concern in a radiological context given the particular radiosensitivity of the lens of the eye. In light of the substantially increased application of computed tomography in brain imaging, an investigation of the relevent literature is warranted to assess thresholds, lens radiation doses and dose reduction techniques in respect to the cataractogenic risk of such examinations. The value and very existence of a lens dose threshold is debatable given different considerations of radiation dose, latency, opacity classifications and historical sample populations, though ICRP guidelines suggest a threshold of 0.5 Gy. Documented CT-specific radiation doses to the eye following scans of the brain are highly variable between studies (2-130 mGy), primarily owing to discrepancies in scanning technique. These findings, when coupled with the relative ambiguity of known threshold values, present difficulties in assessing the overall risk of cataracts following serial CT examinations to the head. In the absence of definitive risk evaluations, a cautionary approach is advised. The implementation of gantry tilt along the supraorbital margin is recommended as standard practice on account of its highly effective radiation dose reduction outcomes. Organ-based tube modulation and reductions in tube current may also be considered beneficial. Bismuth eye shielding is only advised where gantry tilting is unachievable, and in such cases, ensure careful adherence to appropriate shield placement and infection control measures.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Sci-hub)

  • 文章类型: Journal Article
    The purpose of this study is to evaluate the potential of an automated kilo-voltage selection software for the reduction of lens dose in pediatric CT scans.
    Two metal oxide semiconductor field effect transistor (MOSFET) detectors measured the lens dose in two anthropomorphic 1- and 5-year-old phantoms. These phantoms were scanned using a clinical pediatric brain protocol at 120 kVp as a control with the MDCT scanner. Scans were then repeated using automated kilovoltage software. The automated kilovoltage was set to operate at tube potentials of 120, 110, and 100 kVp. Dose savings were compared with the average lens dose of both eyes between automated kilovoltage and the control setting. Image quality was studied by contrast-to-noise ratios (CNRs) for each setting.
    The mean (± SD) lens dose from the routine brain scan without automated kilovoltage was 0.92 ± 0.03 cGy and 0.81 ± 0.03 cGy for the 1- and 5-year-old phantoms, respectively. Use of the automated kilovoltage software at 120 kVp, 110 kVp, and 100 kVp resulted in dose reductions of 9.8%, 17.4%, and 19.6%, respectively, for the 1-year-old phantom and 1.2%, 8.6%, and 17.3%, respectively, for the 5-year-old phantom. The CNR for all automated kilovoltage scans was within 11% of the control scans for the 1-year-old and within 6% for the 5-year-old phantom.
    Our results show that automated kilovoltage software is effective for reducing the radiation dose to the lens of the eye in pediatric patients. Furthermore, the image quality by CNR remained acceptable within 11% of the baseline for all kilovoltage settings used.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Sci-hub)

  • 文章类型: Journal Article
    The aim of this work is to develop a method to calculate lens dose for fluoroscopically-guided neuro-interventional procedures and for CBCT scans of the head. EGSnrc Monte Carlo software is used to determine the dose to the lens of the eye for the projection geometry and exposure parameters used in these procedures. This information is provided by a digital CAN bus on the Toshiba Infinix C-Arm system which is saved in a log file by the real-time skin-dose tracking system (DTS) we previously developed. The x-ray beam spectra on this machine were simulated using BEAMnrc. These spectra were compared to those determined by SpekCalc and validated through measured percent-depth-dose (PDD) curves and half-value-layer (HVL) measurements. We simulated CBCT procedures in DOSXYZnrc for a CTDI head phantom and compared the surface dose distribution with that measured with Gafchromic film, and also for an SK150 head phantom and compared the lens dose with that measured with an ionization chamber. Both methods demonstrated good agreement. Organ dose calculated for a simulated neuro-interventional-procedure using DOSXYZnrc with the Zubal CT voxel phantom agreed within 10% with that calculated by PCXMC code for most organs. To calculate the lens dose in a neuro-interventional procedure, we developed a library of normalized lens dose values for different projection angles and kVp\'s. The total lens dose is then calculated by summing the values over all beam projections and can be included on the DTS report at the end of the procedure.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Sci-hub)

  • 文章类型: Comparative Study
    OBJECTIVE: The purpose of this study was to study different optimization methods for reducing eye lens dose in head CT.
    METHODS: Two anthropomorphic phantoms were scanned with a routine head CT protocol for evaluation of the brain that included bismuth shielding, gantry tilting, organ-based tube current modulation, or combinations of these techniques. Highsensitivity metal oxide semiconductor field effect transistor dosimeters were used to measure local equivalent doses in the head region. The relative changes in image noise and contrast were determined by ROI analysis.
    RESULTS: The mean absorbed lens doses varied from 4.9 to 19.7 mGy and from 10.8 to 16.9 mGy in the two phantoms. The most efficient method for reducing lens dose was gantry tilting, which left the lenses outside the primary radiation beam, resulting in an approximately 75% decrease in lens dose. Image noise decreased, especially in the anterior part of the brain. The use of organ-based tube current modulation resulted in an approximately 30% decrease in lens dose. However, image noise increased as much as 30% in the posterior and central parts of the brain. With bismuth shields, it was possible to reduce lens dose as much as 25%.
    CONCLUSIONS: Our results indicate that gantry tilt, when possible, is an effective method for reducing exposure of the eye lenses in CT of the brain without compromising image quality. Measurements in two different phantoms showed how patient geometry affects the optimization. When lenses can only partially be cropped outside the primary beam, organ-based tube current modulation or bismuth shields can be useful in lens dose reduction.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Sci-hub)

公众号