Sphygmomanometers

血压计
  • 文章类型: Journal Article
    目的:建议使用听诊式血压计测量血压(BP)以诊断儿童高血压。由于环境问题,水银血压计(MS)被禁止使用,确定无汞血压计的准确性对于更换它们至关重要。我们分析了这些设备的准确性,以指导国家调查的选择。
    方法:每次用MS测量三次BP,听诊装置(AD),使用国家调查数据,104名10-18岁的参与者中的示波装置(OD)。BP的差异定义为MS和其他装置之间的差异。BP的差异,相关性,并对影响因素进行了分析。还比较了高血压的发生频率。
    结果:MS和AD之间的收缩压BP(SBP)和舒张压BP(DBP)差异为0.88±3.36mmHg和0.63±3.95mmHg,MS和OD之间的值分别为0.43±5.83mmHg和4.57±6.89mmHg,分别。MS和OD之间DBP的绝对误差<10mmHg为76%。MS和AD的一致相关系数SBP为0.94,DBP为0.90。MS和OD分别为0.81和0.67。除MS和OD之间的SBP外,臂围与BP差异呈负相关。高血压的频率在MS和AD之间没有差异,但被OD低估了。
    结论:AD与MS有很好的相关性,而OD没有,尤其是DBP。在全国调查中,AD优于OD表明AD是MS的可能替代品。
    OBJECTIVE: Blood pressure (BP) measurement using an auscultatory sphygmomanometer is recommended for diagnosing hypertension in children. As mercury sphygmomanometers (MSs) are banned owing to environmental concerns, it is crucial to determine the accuracy of mercury-free sphygmomanometers to replace them. We analyzed the accuracy of these devices to guide the National Survey selection.
    METHODS: BP was measured thrice each with MS, auscultatory device (AD), and oscillometric device (OD) in 104 participants aged 10-18 using the National Survey data. The difference in BP was defined as the difference between MS and other devices. The BP differences, correlations, and influencing factors were analyzed. The frequencies of hypertension were also compared.
    RESULTS: Systolic BP (SBP) and diastolic BP (DBP) differences between MS and AD were 0.88±3.36 mmHg and 0.63±3.95 mmHg, and those between MS and OD were 0.43±5.83 mmHg and 4.57±6.89 mmHg, respectively. The absolute error of <10 mmHg for DBP between MS and OD was 76%. The concordance correlation coefficient between MS and AD was 0.94 for SBP and 0.90 for DBP, and 0.81 and 0.67, respectively for MS and OD. Arm circumference negatively correlated with BP differences except for SBP between the MS and OD. The frequency of hypertension was not different between MS and AD but was underestimated by OD.
    CONCLUSIONS: AD correlated well with MS, while OD did not, especially for DBP. The superiority of AD over OD suggests AD as a possible alternative for MS in the National Survey.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    这项研究评估了示波装置(OD),MicrolifeWatchBP办公室AFIB,和混合手动听诊装置(AD),Greenlight300TM,在无汞的情况下,为韩国国家健康和营养检查调查确定合适的血压(BP)测量设备。坚持2018年通用标准的建议共识,该研究涉及800名受试者(平均年龄51.2±17.5岁;44.3%为男性),按随机顺序休息5分钟后,进行了一式三份的BP测量(OD-优先:398名参与者;AD-优先:402名参与者)。BP差值计算为OD值减去AD值,结果按测量顺序分层。收缩压/舒张压(SBP/DBP)的总体BP差异和可容忍误差概率分别为-1.1±6.5/-2.6±4.9mmHg和89.2%/92.5%,分别。SBP/DBP的Lin一致相关系数为0.907/0.844(OD-first/AD-first:0.925/0.892,0.842/0.845对于DBP)。高血压(BP≥140和/或90mmHg)的总体一致性为0.71(p<0.0001),OD低估了整体高血压患病率5.1%。对AD-first数据的分析显示,与OD-first数据相比,一致性较低;但是,观察到的血压差符合2018年通用标准的标准1.微生命符合2018年通用标准的标准1,但低估了高血压的患病率。BP差异随着BP水平的提高而增加,男性,更小的AC。随着年龄的增长,SBP的差异减少,DBP的差异增加。
    This study evaluated an oscillometric device (OD), Microlife WatchBP Office AFIB, and a hybrid manual auscultatory device (AD), Greenlight 300TM, to determine a suitable blood pressure (BP) measurement device for the Korea National Health and Nutrition Examination Survey in a mercury-free context. Adhering to the 2018 Universal Standard\'s suggested consensus, the study involved 800 subjects (mean age 51.2 ± 17.5 years; 44.3% male), who underwent triplicate BP measurements following 5 min of rest in a randomized order (OD-first: 398 participants; AD-first: 402 participants). BP difference was calculated as OD value minus AD value, with results stratified by measurement sequence. The overall BP difference and tolerable error probability were -1.1 ± 6.5/-2.6 ± 4.9 mmHg and 89.2%/92.5% for systolic/diastolic BP (SBP/DBP), respectively. Lin\'s concordance correlation coefficient was 0.907/0.844 for SBP/DBP (OD-first/AD-first: 0.925/0.892 for SBP, 0.842/0.845 for DBP). The overall agreement for hypertension (BP ≥ 140 and/or 90 mmHg) was 0.71 (p < 0.0001), and the OD underestimated the overall hypertension prevalence by 5.1%. Analysis of the AD-first data revealed a lower level of agreement compared to the OD-first data; however, the observed blood pressure difference adhered to Criterion 1 of the 2018 Universal Standard. Microlife met the Criterion 1 of 2018 Universal Standard but underestimated the prevalence of hypertension. The BP discrepancy increased with higher BP levels, male sex, and smaller AC. With increasing age, the discrepancy decreased for SBP and increased for DBP.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    目的:根据欧洲高血压学会血压监测工作组2002(ESH-IP2002)制定的国际协议,报告对Riester大本钟台式无血血压计的验证。此传统出版物旨在向用户保证该设备满足当时的要求。
    方法:根据参与者的年龄,性别,臂围和入口SBP/DBP。对33名参与者进行了验证。根据ESH-IP评估血压计,与汞标准相比,其精度区域定义为≤5、≤10、≤15mmHg或更高。
    结果:平均(±SD)年龄为50.5±13.0岁,范围29-71年,输入SBP142.6±23.7mmHg,条目DBP89.0±17.8mmHg。设备通过了列出的所有要求和验证协议。里斯特大本钟方桌无液血压计稍微低估了观察者测量的SBP,但稍微高估了DBP。观察者与装置的分歧为-0.8±6.4mmHgSBP和0.6±4.0mmHgDBP。
    结论:这些数据表明,RisterBigBenSquare台式无液血压计符合ESH-IP2002对BP监测仪的验证要求。正是在此基础上,英国和爱尔兰高血压协会建议将其用于成人人群的临床使用。
    OBJECTIVE: To report a validation of the Riester Big Ben Square Desk Aneroid Sphygmomanometer according to the international protocol developed by the Working Group on Blood Pressure Monitoring of the European Society of Hypertension 2002 (ESH-IP 2002) in the interest of transparency. This legacy publication is intended to assure users that the device satisfied the requirements in place at that time.
    METHODS: Performance of the device was assessed by participants\' age, sex, arm circumference and entry SBP/DBP. Validation was performed in 33 participants. The sphygmomanometer was assessed according to the ESH-IP, which defines zones of accuracy compared to the mercury standard as ≤5, ≤10, ≤15 mmHg or more.
    RESULTS: The mean (± SD) age was 50.5 ± 13.0 years, range 29-71 years, entry SBP 142.6 ± 23.7 mmHg, entry DBP 89.0 ± 17.8 mmHg. The device passed all the requirements listed and the validation protocol. The Riester Big Ben Square Desk aneroid sphygmomanometer slightly underestimated the observer-measured SBP, yet slightly overestimated DBP. The observer-device disagreement was -0.8 ± 6.4 mmHg SBP and +0.6 ± 4.0 mmHg DBP.
    CONCLUSIONS: These data show that the Riester Big Ben Square Desk aneroid sphygmomanometer fulfilled the ESH-IP 2002 requirements for the validation of BP monitors. It was on this basis that the British and Irish Hypertension Society recommended it for clinical use in the adult population.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    目的:无袖带血压(BP)测量(BPM)设备的数量越来越多。尽管在比较单个测量结果时取得了有希望的结果,这些装置追踪与初始校正BP(CalibBP)相关的24小时内BP水平变化的能力尚不清楚.我们的目标是使用脉冲渡越时间在无袖口设备中分析这种能力。
    方法:我们前瞻性招募了166名参与者,同时进行无袖口(Somnotouch-NIBP)和袖口(Spacelabs90217A/IEMMobil-O-graph)24小时BPM。作为无袖口装置的CalibBP,首次使用基于袖带的BP.作为CalibBP后BP水平变化的替代,我们使用CalibBP和平均24小时之间的差值,这两个设备测量的清醒和睡眠血压。此外,我们分析了CalibBP与袖带BPM的差异与袖带BPM装置与无袖带BPM装置的差异之间的关系.
    结果:基于袖带或无袖带的血压装置的CalibBP与平均24hBP之间的平均(SD)差异为:收缩压为7.4(13.2)对1.8(8.3)mmHg(P<0.0001),舒张压为6.6(6.8)对1.6(5.8)mmHg(P<0.0001)。在CalibBP和基于袖带的BPM值之间的差异以及基于袖带的BPM装置和无袖带的BPM装置之间的差异之间观察到接近线性关系。
    结论:我们的数据表明,无袖口BPM设备在CalibBP后跟踪BP水平变化的能力较低。此外,无袖口装置的准确性与初始CalibBP后的BP水平变化有关-BP水平变化越大,设备之间的差异越大。
    背景:https://www.clinicaltrials.gov;唯一标识符:NCT03054688;NCT03975582.
    OBJECTIVE: There is an increasing number of cuffless blood pressure (BP) measurement (BPM) devices. Despite promising results when comparing single measurements, the ability of these devices to track changes in BP levels over 24 h related to an initial calibration BP (CalibBP) is unknown. Our aim was to analyse this ability in a cuffless device using pulse transit time.
    METHODS: We prospectively enrolled 166 participants for simultaneously performed cuffless (Somnotouch-NIBP) and cuff-based (Spacelabs 90217A/IEM Mobil-O-graph) 24 h BPM. As CalibBP for the cuffless device, first cuff-based BP was used. As surrogate for changes in BP levels after the CalibBP, we used the difference between the CalibBP and mean 24 h, awake and asleep BP measured by the two devices. In addition, we analysed the relationship between the difference of the CalibBP and the cuff-based BPM versus the difference between the cuff-based and the cuffless BPM devices.
    RESULTS: Mean(SD) difference between the CalibBP and mean 24hBP by the cuff-based or cuffless BP device were 7.4 (13.2) versus 1.8 (8.3) mmHg for systolic ( P  < 0.0001) and 6.6 (6.8) versus 1.6 (5.8) mmHg for diastolic ( P  < 0.0001). A near linear relationship was seen among the difference between the CalibBP and the cuff-based BPM values and the difference between the cuff-based and cuffless BPM device.
    CONCLUSIONS: Our data indicate a lower ability of the cuffless BPM device to track changes of BP levels after CalibBP. In addition, cuffless device accuracy was associated with the changes in BP levels after the initial CalibBP - the larger the BP level change, the larger the difference between the devices.
    BACKGROUND: https://www.clinicaltrials.gov ; Unique identifier: NCT03054688; NCT03975582.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    水银血压计(MS)现在已经越来越少地使用,并且没有制造新的设备(根据Minamata公约2013)。电子血压计(ES)在临床实践中的应用已变得越来越普遍。然而,在高海拔地区使用ES的可靠证据仍然很少。本研究的目的是验证ES在高海拔地区的适用性。
    在卢霍县,四川省,中国,海拔3400米,两名训练有素的医师使用水银血压计和ES测量参与者的血压(BP).Pearson相关分析和配对T检验,分别,用于比较由两个设备测量的BP值之间的相关性和差异。根据2018年医疗器械促进协会/欧洲高血压学会/国际标准化组织(AAMI/ESH/ISO)合作声明的验证标准,评估了ES在高海拔地区的适用性。
    在这项研究中,包括257名参与者。这两个设备测得的BP值之间存在很强的相关性,收缩压(SBP)和舒张压(DBP)的相关系数分别为0.97和0.93。与MS相比,ES倾向于测量受试者的DBP(76.21±13.29mmHg与76.53±14.07mmHg;P=0.557)准确,但高估了受试者的SBP(123.32±22.25mmHg与121.34±22.88mmHg;P<0.001)在一定程度上。两种设备在正常BP分类中的一致性,高血压前期,高血压占88.9%,80.7%,和89.2%,分别。
    一般来说,在3,400m高度使用ES成功达到AAMI/ESH/ISO合作声明的验证标准。建议在高海拔地区使用ES,包括高达3,400米。此外,因为ES倾向于高估SBP,我们推测它可能需要在高海拔地区进行校准。
    UNASSIGNED: Mercury sphygmomanometer (MS) has now been less and less used and no new devices have been manufactured (according to Minamata convention 2013). The application of the electronic sphygmomanometer (ES) in clinical practice has become increasingly common. However, reliable evidence for the use of the ES in high-altitude areas remains scarce. The purpose of this study was to validate the applicability of the ES in high altitude areas.
    UNASSIGNED: In Luhuo County, Sichuan Province, China, 3,400 m above the sea level, two trained physicians measured the blood pressure (BP) of participants using both the mercury sphygmomanometer and the ES. Pearson correlation analysis and paired T-test, respectively, were used to compare the correlation and the difference between the BP values measured by the two devices. The applicability of the ES in high-altitude areas was evaluated according to the validation standards of the 2018 Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Collaboration Statement.
    UNASSIGNED: In this study, 257 participants were included. There was a strong correlation between BP values measured by the two devices, with correlation coefficients for systolic blood pressure (SBP) and diastolic blood pressure (DBP) of 0.97 and 0.93, respectively. Compared with the MS, the ES tended to measure the subjects\' DBP (76.21 ± 13.29 mmHg vs. 76.53 ± 14.07 mmHg; P = 0.557) accurately, but overestimate the SBP of the subjects (123.32 ± 22.25 mmHg vs. 121.34 ± 22.88 mmHg; P < 0.001) to some extent. The consistency of the two devices in the classification of normal BP, prehypertension, and hypertension was 88.9%, 80.7%, and 89.2%, respectively.
    UNASSIGNED: In general, the utilization of ES at 3,400 m altitude successfully met the validation standards of the AAMI/ESH/ISO Collaboration Statement. The use of ES can be recommended at a high altitude, including up to 3,400 m. In addition, because the ES tended to overestimate SBP, we speculate that it may need to be calibrated in high-altitude areas.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Clinical Study
    根据AAMI/ESH/ISO通用标准(ISO81060-2:2018+Amd.1:2020),评估成人DBP-1333b上臂血压(BP)测量装置的准确性。在成年人群中招募受试者。测试装置是臂式电子血压计(DBP-1333b),参考装置是台式血压计(XJ11D)。以台式血压计测得的BP数据作为参考BP,对测试设备的非侵入性BP模块的准确性进行评估,以确定其是否符合要求。分析了来自90个人的数据。根据标准1,测试装置和参考装置之间的SBP的平均差为0.19mmHg,SD为7.45mmHg。DBP的平均差为-0.59mmHg,SD为6.47mmHg。SBP和DBP的平均差异均小于5mmHg,SD小于8mmHg,符合要求。根据标准2,SBP的SD为5.79mmHg,小于6.95mmHg,符合要求。DBP的SD为5.58mmHg,小于6.93mmHg,符合要求。结论是DBP-1333b符合AAMI/ESH/ISO通用标准(ISO81060-2:2018+Amd.1:2020),可以推荐成人使用。
    To evaluate the accuracy of the DBP-1333b upper-arm blood pressure (BP) measuring device in the adult population according to the AAMI/ESH/ISO universal standard (ISO 81060-2:2018+Amd.1:2020). Subjects were recruited in the adult population. The test device was an arm-type electronic sphygmomanometer (DBP-1333b) and the reference device was a desktop sphygmomanometer (XJ11D). Using the BP data measured by the desktop sphygmomanometer as reference BP, the accuracy of the non-invasive BP module of the test device was evaluated to determine whether it met the requirements. Data from 90 individuals were analysed. According to Criterion 1, the mean difference of SBP between the test and reference device was 0.19 mmHg and the SD was 7.45 mmHg. The mean difference of DBP was -0.59 mmHg and the SD was 6.47 mmHg. The mean difference of both SBP and DBP was less than 5 mmHg, and the SD was less than 8 mmHg, which met the requirements. According to Criterion 2, SD of SBP was 5.79 mmHg, which was less than 6.95 mmHg and met the requirements. The SD of DBP was 5.58 mmHg, which was less than 6.93 mmHg and met the requirements. It was concluded that the DBP-1333b complies with the AAMI/ESH/ISO universal standard (ISO 81060-2:2018+Amd.1:2020) and can be recommended for use by the adults.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    背景:自测血压(SMBP)监测越来越多地用于远程高血压管理,但是家用血压(BP)设备的实际性能尚不清楚。我们使用美国医学协会(AMA)的自我测量血压(SMBP)设备准确性测试工具检查了患者家用设备的BP测量值。
    方法:单个内科诊所的患者接受了多达5个座位,使用家用设备和自动BP设备(OmronHEM-907XL)的同臂BP读数。按照AMA的3步协议,我们第一次使用了病人的家用设备,第二,和第四次测量以及用于第三次和第五次(如果需要)测量的办公设备。设备协议失败定义为在两个确认步骤中的任一步骤中,家用和办公室设备之间收缩压>10mmHg的绝对差异。业绩按品牌进行了检查(欧姆龙vs非欧姆龙)。此外,我们通过对人口统计特征进行校正的logistic回归模型研究了与协议失败相关的患者因素.
    结果:我们评估了152名患者(平均年龄60±15岁,58%的女性,31%黑色)在2020年10月至2021年11月之间。设备协议失败发生在22.4%(95%CI:16.4%,29.7%)的测试设备,其中欧姆龙设备占19.1%,非欧姆龙设备占27.6%(P=0.23)。没有患者特征与协议失败相关。
    结论:根据AMASMBP设备准确性协议,超过五分之一的家庭设备不一致。这些发现证实了基于办公室的设备比较的重要性,以确保家庭BP监测的准确性。
    Self-measured blood pressure (SMBP) monitoring is increasingly used for remote hypertension management, but the real-world performance of home blood pressure (BP) devices is unknown. We examined BP measurements from patients\' home devices using the American Medical Association\'s (AMA) SMBP Device Accuracy Test tool.
    Patients at a single internal medicine clinic underwent up to five seated, same-arm BP readings using a home device and an automated BP device (Omron HEM-907XL). Following the AMA\'s three-step protocol, we used the patient\'s home device for the first, second, and fourth measurements and the office device for the third and fifth (if needed) measurements. Device agreement failure was defined as an absolute difference in systolic BP >10 mm Hg between the home and office devices in either of two confirmatory steps. Performance was examined by brand (Omron vs. non-Omron). Moreover, we examined patient factors associated with agreement failure via logistic regression models adjusted for demographic characteristics.
    We evaluated 152 patients (mean age 60 ± 15 years, 58% women, 31% Black) seen between October 2020 and November 2021. Device agreement failure occurred in 22.4% (95% CI: 16.4%, 29.7%) of devices tested, including 19.1% among Omron devices and 27.6% among non-Omron devices (P = 0.23). No patient characteristics were associated with agreement failure.
    Over one-fifth of home devices did not agree based on the AMA SMBP device accuracy protocol. These findings confirm the importance of office-based device comparisons to ensure the accuracy of home BP monitoring.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    为了正确诊断高血压,全世界都需要改善血压(BP)测量误差。心血管疾病每年导致1790万人死亡,对医疗保健造成巨大的货币压力。目前3mmHg的测量不确定度有待提高。缺乏或不存在动态压力测量标准。在这项研究中,我们提出了一种新颖的方法,用于在使用电容式麦克风进行BP测量时测量血压计管内的气压。我们设计了,已建成,并测试了使用射频(RF)调制方法将电容式麦克风的电容变化转换为压力信号的系统。我们测试了具有低频(LF)声源的RF麦克风,BP模拟器,并使用压阻式压力传感器作为参考。进行了必要的测试以评估系统的不确定性预算。RF麦克风原型在0至300mmHg的压力范围内具有0.5Hz至280Hz的工作频率范围。射频传声器的总扩展不确定度(k=2,p=95.5%)为4.32mmHg。所提出的方法可以建立BP测量设备对IEC61094-2中描述的声学标准的可追溯性,也可以用于形成动态BP标准。
    There is a worldwide need to improve blood pressure (BP) measurement error in order to correctly diagnose hypertension. Cardiovascular diseases cause 17.9 million deaths annually and are a substantial monetary strain on healthcare. The current measurement uncertainty of 3 mmHg should be improved upon. Dynamic pressure measurement standards are lacking or non-existing. In this study we propose a novel method of measuring air pressure inside the sphygmomanometer tubing during BP measurement using a condenser microphone. We designed, built, and tested a system that uses a radiofrequency (RF) modulation method to convert changes in capacitance of a condenser microphone into pressure signals. We tested the RF microphone with a low-frequency (LF) sound source, BP simulator and using a piezoresistive pressure sensor as a reference. Necessary tests were conducted to assess the uncertainty budget of the system. The RF microphone prototype has a working frequency range from 0.5 Hz to 280 Hz in the pressure range from 0 to 300 mmHg. The total expanded uncertainty (k = 2, p = 95.5%) of the RF microphone was 4.32 mmHg. The proposed method could establish traceability of BP measuring devices to acoustic standards described in IEC 61094-2 and could also be used in forming dynamic BP standards.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    准确的血压(BP)测量对于高血压的检测和管理至关重要。韩国国家健康和营养检查调查(KNHANES)使用代表性的横截面数据评估韩国人的健康状况。历史上,年龄≥10岁的参与者使用水银血压计进行BP测量。然而,KNHANES于2020年过渡到Greenlight300TM(无汞听诊设备),适用于年龄≥6岁的参与者,并在2021-2022年使用了双设备(MicrolifeWatchBPOfficeAFIB和Greenlight)。为了确保一致性,从2023年开始,KNHANES将采用Microlife作为统一的BP设备,与Greenlight一起进行设备验证。根据新协议,年龄≥6岁的参与者将在环境温度(20-25℃)和噪声≤65dB下休息5分钟后,每隔30秒测量3次血压.第二和第三读数的平均值将用作代表性BP值。质量控制(QC)计划涉及四名训练有素的审查员,每年通过“BP测量计划的质量控制和保证”三次,每年进行一次“每周校准过程视频监控”。此外,QC小组将每年在移动考试中心进行三次“BP测量现场评估”。还开发了用于BP设备的五步QC过程。本文档概述了KNHANES中的标准化BP测量协议和严格的QC计划,旨在确保准确可靠的BP数据,用于韩国的流行病学研究和公共卫生政策制定。
    Accurate blood pressure (BP) measurement is crucial for hypertension detection and management. The Korea National Health and Nutrition Examination Survey (KNHANES) assesses the health of Koreans using representative cross-sectional data. BP measurements were historically done with mercury sphygmomanometers for participants aged ≥10 years. However, KNHANES transitioned to Greenlight 300TM (mercury-free auscultatory device) in 2020 for participants aged ≥6 years and used dual devices (Microlife WatchBP Office AFIB and Greenlight) in 2021-2022. To ensure consistency, KNHANES will adopt Microlife as the unified BP device with Greenlight for device validation from 2023. Under the new protocol, participants aged ≥6 years will have their BP measured three times at 30-second intervals after a 5-minute rest under ambient temperature (20-25℃) and noise ≤65 dB. The average of the 2nd and 3rd readings will be used as the representative BP value. The quality control (QC) program involves four trained examiners passing the \"quality control and assurance of BP measurement program\" three times annually, and undergoing \"video monitoring of weekly calibration process\" once a year. Additionally, the QC team will conduct \"on-site evaluations of BP measurement\" at mobile examination centers three times a year. A Five-Step QC process for BP devices was also developed. This document outlines the standardized BP measurement protocol and rigorous QC program in KNHANES, aiming to ensure accurate and reliable BP data for epidemiological research and public health policymaking in South Korea.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

  • 文章类型: Journal Article
    目的:临床以外的血压(BP)波动在心血管疾病发展中的作用越来越被认识到。晕厥,和过早死亡,并作为定制高血压治疗的有希望的目标。然而,当前基于袖带的BP设备,包括家用和移动设备,无法捕捉到人类活动中BP变异性的广度,经验,和上下文。
    结果:无袖口,可穿戴BP设备提供了节拍的承诺,连续,在清醒和睡眠期间进行血压的无创测量,使患者的不便最小。重要的是,无袖口BP装置可以表征BP变异性,允许识别家庭环境中BP激增的患者特定触发因素。不幸的是,证据的速度,regulation,和验证测试已经落后于创新和直接消费者营销的步伐。我们概述了用于无袖口BP监测的可用技术和设备,校准和验证这些设备的注意事项,以及无袖扣BP范式的希望和陷阱。
    Blood pressure (BP) fluctuations outside of clinic are increasingly recognized for their role in the development of cardiovascular disease, syncope, and premature death and as a promising target for tailored hypertension treatment. However, current cuff-based BP devices, including home and ambulatory devices, are unable to capture the breadth of BP variability across human activities, experiences, and contexts.
    Cuffless, wearable BP devices offer the promise of beat-to-beat, continuous, noninvasive measurement of BP during both awake and sleep periods with minimal patient inconvenience. Importantly, cuffless BP devices can characterize BP variability, allowing for the identification of patient-specific triggers of BP surges in the home environment. Unfortunately, the pace of evidence, regulation, and validation testing has lagged behind the pace of innovation and direct consumer marketing. We provide an overview of the available technologies and devices for cuffless BP monitoring, considerations for the calibration and validation of these devices, and the promise and pitfalls of the cuffless BP paradigm.
    导出

    更多引用

    收藏

    翻译标题摘要

    我要上传

       PDF(Pubmed)

公众号