关键词: (71)Ga(n,γ)(72)Ga reaction AB-BNCT device Epithermal neutron flux detector Quality assurance

Mesh : Humans Boron Neutron Capture Therapy / methods Neutrons Radiotherapy Dosage Gamma Rays Monte Carlo Method

来  源:   DOI:10.1016/j.apradiso.2024.111249

Abstract:
The 71Ga(n,γ)72Ga reaction-based epithermal neutron flux detectors are novel instruments developed to measure the epithermal neutron flux of boron neutron capture therapy (BNCT) treatment beams. In this study, a spherical epithermal neutron flux detector using 71Ga(n,γ)72Ga reaction was prototyped. The performance of the detector was experimentally evaluated at an accelerator-based BNCT (AB-BNCT) device developed by Lanzhou University, China. Based on the experimental results and related analysis, we demonstrated that the detector is a reliable tool for the quality assurance of BNCT treatment beams.
摘要:
71Ga(n,γ)基于72Ga反应的超热中子通量探测器是开发用于测量硼中子俘获治疗(BNCT)治疗束的超热中子通量的新型仪器。在这项研究中,使用71Ga(n,γ)72Ga反应为原型。在兰州大学研制的基于加速器的BNCT(AB-BNCT)装置上对探测器的性能进行了实验评估,中国。根据实验结果和相关分析,我们证明了检测器是BNCT治疗束质量保证的可靠工具。