关键词: Ambulatory blood pressure monitor Continual blood pressure monitoring Cuffless blood pressure Daytime blood pressure Hypertension Optical blood pressure monitor

Mesh : Humans Blood Pressure / physiology Blood Pressure Monitoring, Ambulatory / methods Blood Pressure Monitors Prospective Studies

来  源:   DOI:10.1038/s41440-023-01258-2   PDF(Pubmed)

Abstract:
In this preliminary study, we compared daytime blood pressure (BP) measurements performed by a commercially available cuffless-and continual-BP monitor (Aktiia monitor, Neuchâtel, Switzerland) and a traditional ambulatory BP monitor (ABPM; Dyasis 3, Novacor, Paris, France) from 52 patients enrolled in a 12-week cardiac rehabilitation (CR) program (Neuchâtel, Switzerland). Daytime (9am-9pm) systolic (SBP) and diastolic (DBP) BP from 7-day averaged data from Aktiia monitor were compared to 1-day averaged BP data from ABPM. No significant differences were found between the Aktiia monitor and the ABPM for SBP (μ ± σ [95% confidence interval]: 1.6 ± 10.5 [-1.5, 4.6] mmHg, P = 0.306; correlation [R2]: 0.70; ± 10/ ± 15 mmHg agreements: 60%, 84%). Marginally non-significant bias was found for DBP (-2.2 ± 8.0 [-4.5, 0.1] mmHg, P = 0.058; R2: 0.66; ±10/±15 mmHg agreements: 78%, 96%). These intermediate results show that daytime BP measurements using the Aktiia monitor generate data comparable to that of an ABPM monitor.
摘要:
在这项初步研究中,我们比较了市售的无袖口和连续BP监测仪(Aktiia监测仪,纳沙泰尔,瑞士)和传统的动态血压监测仪(ABPM;Dyasis3,Novacor,巴黎,法国)来自参加12周心脏康复(CR)计划的52名患者(纳沙泰尔,瑞士)。将来自Aktiia监测仪的7天平均数据的白天(上午9点至下午9点)收缩压(SBP)和舒张压(DBP)BP与来自ABPM的1天平均BP数据进行比较。Aktiia监测仪和ABPM的SBP之间没有发现显着差异(μ±σ[95%置信区间]:1.6±10.5[-1.5,4.6]mmHg,P=0.306;相关性[R2]:0.70;±10/±15mmHg协议:60%,84%)。DBP(-2.2±8.0[-4.5,0.1]mmHg,P=0.058;R2:0.66;±10/±15mmHg协议:78%,96%)。这些中间结果表明,使用Aktiia监测仪进行的白天BP测量可生成与ABPM监测仪相当的数据。
公众号