关键词: Delayed orbital hemorrhage fracture repair hematic cyst nylon foil orbital implant

Mesh : Anticoagulants / adverse effects Device Removal Diplopia / etiology Exophthalmos / etiology Fracture Fixation / instrumentation Humans Male Middle Aged Nylons Orbital Fractures / surgery Retrobulbar Hemorrhage / chemically induced Retrospective Studies Tomography, X-Ray Computed Warfarin / adverse effects

来  源:   DOI:10.1080/01676830.2019.1639771   PDF(Sci-hub)

Abstract:
The authors describe the first report in the literature of delayed orbital hemorrhage that may be partly caused by supratherapeutic anticoagulation. A 52-year-old man with supratherapeutic international normalized ratio (INR) presented with acute proptosis, orbital pain and diplopia 9 months after the floor and medial orbital wall fracture repair using nylon foil implant. He was found to have hemorrhaged into the capsule surrounding the orbital implant. Three weeks later, the patient underwent implant removal after warfarin was discontinued for 5 days and INR was normalized. His symptoms resolved postoperatively. This case describes a unique risk factor of delayed orbital hemorrhage in patients with previous orbital fracture repair, and highlights that coagulopathy should be investigated in patients presenting with acute proptosis with a history of orbital fracture repair. The authors also provide a comprehensive and up-to-date literature review on previously reported cases with delayed hemorrhagic complications from alloplastic orbital implants.
摘要:
暂无翻译