关键词: drop dispensing electrospray ionization immediate drop on demand mass spectrometry microtiter plates noncontact dispensing open port sampling interface

Mesh : Isotope Labeling Molecular Weight Propranolol / chemistry Proteins / analysis Reproducibility of Results Reserpine / chemistry Spectrometry, Mass, Electrospray Ionization / instrumentation methods

来  源:   DOI:10.4155/bio-2017-0104

Abstract:
OBJECTIVE: The aim of this work was to demonstrate and evaluate the analytical performance of coupling the immediate drop on demand technology to a mass spectrometer via the recently introduced open port sampling interface and ESI. Methodology & results: A maximum sample analysis throughput of 5 s per sample was demonstrated. Signal reproducibility was 10% or better as demonstrated by the quantitative analysis of propranolol and its stable isotope-labeled internal standard propranolol-d7. The ability of the system to multiply charge and analyze macromolecules was demonstrated using the protein cytochrome c.
CONCLUSIONS: This immediate drop on demand technology/open port sampling interface/ESI-MS combination allowed for the quantitative analysis of relatively small mass analytes and was used for the identification of macromolecules like proteins.