{Reference Type}: Journal Article {Title}: Influence of Potentiostat Hardware on Electrochemical Measurements. {Author}: Krishnamurthy A;Žagar Soderžnik K; {Journal}: Sensors (Basel) {Volume}: 24 {Issue}: 15 {Year}: 2024 Jul 29 {Factor}: 3.847 {DOI}: 10.3390/s24154907 {Abstract}: We describe two operating modes for the same potentiostat, where the redox processes of hydroquinone in a hydrochloric acid medium are contrasted for cyclic voltammetry (CV) as functions of a digital/staircase scan and an analogue/linear scan. Although superficially there is not much to separate the two modes of operation as an end user, differences can be seen in the voltammograms while switching between the digital and analogue modes. The effects of quantization clearly have some impact on the measurements, with the outputs between the two modes being a function of the equivalent-circuit model of the electrochemical system under investigation. Increasing scan rates when using both modes produces higher peak redox currents, with the differences between the analogue and digital modes of operation being consistent as a function of the scan rate. Differences between the CV loops between the analogue and digital modes show key differences at certain points along the scans, which can be attributed to the nature of the electrolyte affecting the charging and discharging processes and consequently changing the peak currents of the redox processes. The faradaic processes were shown to be independent of the scan rates. Simulations of the equivalent-circuit behaviour show differences in the responses to different input signals, i.e., the step and ramp responses of the system. Both the voltage and current steps and ramp responses showed the time-domain behaviour of distinct elements of the equivalent electrochemical circuit model as an approximation of the applied digital and analogue CV input signals. Ultimately, it was concluded that similar parameters between the two modes of operation available with the potentiostat would lead to different output voltammograms and, despite advances in technology, digital systems can never fully emulate a true analogue system for electrochemical applications. These observations showcase the value of having hardware capable of true analogue characteristics over digital systems.