{Reference Type}: Journal Article {Title}: Epithelial thickness map-adjusted transepithelial photorefractive keratectomy for treatment of myopic astigmatism: 12-month results. {Author}: Niknam S;Niknam J;Salouti R;Nowroozzadeh MH; {Journal}: Int Ophthalmol {Volume}: 44 {Issue}: 1 {Year}: 2024 Jun 22 {Factor}: 2.029 {DOI}: 10.1007/s10792-024-03184-9 {Abstract}: OBJECTIVE: To evaluate the refractive results of Transepithelial Photorefractive Keratectomy (t-PRK) with the Technolas Teneo2 Excimer laser platform.
METHODS: In this retrospective comparative interventional case series, a total of 199 patients with myopia ranging from - 1 to - 7 diopters were enrolled and separated into three groups based on their target refraction of zero (group 1), - 0.25 (group 2), and - 0.5 diopters (group 3), respectively. The main outcome measure was post-operative cycloplegic refraction. Trans-PRK was performed using the Technolas Teneo2 Excimer laser. To prevent any remaining epithelium during stromal ablation, we adopted the thickest epithelial point in the 7 mm central map as the reference for Phototherapeutic keratectomy (PTK) depth. Patients were examined three and twelve months after the procedure, and the results were analyzed.
RESULTS: At the 12-month follow-up, uncorrected distance visual acuity was 20/20 in all patients. However, there was a significant difference in cycloplegic spherical equivalent refraction between the three groups. The 12-month post-operative spherical equivalent refraction was 0.90 ± 0.33 D, 0.79 ± 0.26 D, and 0.60 ± 0.19 D in groups 1, 2, and 3, respectively (P < 0.001; Kruskal-Wallis test). The rates of spherical equivalent refraction of more than 0.75 D were 58.3%, 39.1%, and 9.1% in the 0 D, - 0.25 D, and -0.50 D groups, respectively (P < 0.001; Chi-squared test).
CONCLUSIONS: The t-PRK with Technolas Teneo2 Excimer laser and epithelial thickness map adjustment of PTK induce a significant amount of residual hyperopia (> 0.75 D) in a large proportion of eyes with a target refraction of 0 or - 0.25, which is significantly reduced by using a target refraction of - 0.5.