{Reference Type}: Journal Article {Title}: Development and validation of high-performance liquid chromatography method for the simultaneous quantification of rivastigmine hydrogen tartrate and asiaticoside co-loaded in niosomes: A Box-Behnken design approach. {Author}: Hnin HM;Tun T;Jansook P; {Journal}: J Chromatogr B Analyt Technol Biomed Life Sci {Volume}: 1241 {Issue}: 0 {Year}: 2024 Jul 1 {Factor}: 3.318 {DOI}: 10.1016/j.jchromb.2024.124170 {Abstract}: Rivastigmine hydrogen tartrate (RHT), a reversible cholinesterase inhibitor, is considered as the first-line therapy for mild to moderate Alzheimer's disease. Asiaticoside (AS), a pentacyclic triterpenoid saponin, is well known as cognitive enhancer due to its antioxidant effect. Based on the hypothesis of their synergistic therapeutic potential, RHT and AS were co-encapsulated in niosomal formulation. A simple, precise, and accurate high-performance liquid chromatography method was developed for simultaneous quantitative analysis. The chromatographic parameters were optimized by Box-Behnken experimental design. The separation was performed on a reversed-phase Phenomenex C18 (150 mm × 4.6 mm, 5 μm) column at 30 °C under the UV detection of 210 nm. The optimized mobile phase consisted of a mixture of 20 mM potassium dihydrogen phosphate buffer (pH 2.6) and acetonitrile (72:28 % v/v) under the isocratic mode at the flow rate of 0.9 mL/min. The developed method was fully validated under the ICH guidelines and could be successfully applied for simultaneous quantitative analysis of RHT and AS in niosomal formulation.