{Reference Type}: Journal Article {Title}: Evaluation of the QMAC-dRAST System Version 2.5 for Rapid Antimicrobial Susceptibility Testing of Gram-Negative Bacteria From Positive Blood Culture Broth and Subcultured Colony Isolates. {Author}: Kim TY;Kang M;Shim HJ;Kang OK;Huh HJ;Lee NY; {Journal}: J Clin Lab Anal {Volume}: 38 {Issue}: 9 {Year}: 2024 May 28 {Factor}: 3.124 {DOI}: 10.1002/jcla.25043 {Abstract}: BACKGROUND: Rapid antimicrobial susceptibility testing (AST) for bloodstream infections (BSIs) facilitates the optimization of antimicrobial therapy, preventing antimicrobial resistance and improving patient outcomes. QMAC-dRAST (QuantaMatrix Inc., Korea) is a rapid AST platform based on microfluidic chip technology that performs AST directly using positive blood culture broth (PBCB). This study evaluated the performance of QMAC-dRAST for Gram-negative bacteria using PBCB and subcultured colony isolates, comparing it with that of VITEK 2 (bioMérieux, France) using broth microdilution (BMD) as the reference method.
METHODS: We included 141 Gram-negative blood culture isolates from patients with BSI and 12 carbapenemase-producing clinical isolates of Enterobacterales spiked into blood culture bottles. QMAC-dRAST performance was evaluated using PBCB and colony isolates, whereas VITEK 2 and BMD were tested only on colony isolates.
RESULTS: For PBCB, QMAC-dRAST achieved 92.1% categorical agreement (CA), 95.3% essential agreement (EA), with 1.8% very major errors (VMEs), 3.5% major errors (MEs), and 5.2% minor errors (mEs). With colony isolates, it exhibited 92.5% CA and 95.1% EA, with 2.0% VMEs, 3.2% MEs, and 4.8% mEs. VITEK 2 showed 94.1% CA and 96.0% EA, with 4.3% VMEs, 0.4% MEs, and 4.3% mEs. QMAC-dRAST yielded elevated error rates for specific antimicrobial agents, with high VMEs for carbapenems and aminoglycosides. The median time to result for QMAC-dRAST was 5.9 h for PBCB samples and 6.1 h for subcultured colony isolates.
CONCLUSIONS: The QMAC-dRAST system demonstrated considerable strengths and comparable performance to the VITEK 2 system; however, challenges were discerned with specific antimicrobial agents, underlining a necessity for improvement.