{Reference Type}: Journal Article {Title}: Comparison of Alignment Accuracy and Clinical Outcomes between a CT-Based, Saw Cutting Robotic System and a CT-Free, Jig-guided Robotic System for Total Knee Arthroplasty. {Author}: Zhou G;Wang X;Geng X;Li Z;Tian H; {Journal}: Orthop Surg {Volume}: 16 {Issue}: 5 {Year}: 2024 May 7 {Factor}: 2.279 {DOI}: 10.1111/os.14055 {Abstract}: OBJECTIVE: The different cutting mode of robot-assisted TKAs may influence the accuracy of alignment. The purpose of this study was to compare alignment accuracy and early clinical outcomes between a CT-based, saw cutting robotic system (MAKO) and a CT-free, jig-guided robotic system (ROSA) for total knee arthroplasty (TKA).
METHODS: A total of 20 MAKO TKAs and 20 ROSA TKAs from June 2021 to June 2022 were retrospectively analyzed. Differences in the postoperative hip-knee-ankle (HKA) angle, lateral distal femoral angle (LDFA), medial proximal tibial angle (MPTA), posterior tibial slope (PTS) and 3° outlier frequency of the HKA, LDFA, MPTA and PTS were studied at 3 months and 1 year of follow-up. The operative time and total blood loss (TBL) were compared between these two groups. Clinical outcomes at 1 year after surgery, including range of motion (ROM), Western Ontario McMaster University Osteoarthritis Index (WOMAC) score, and Knee Society Score-2011 (KSS-2011), were also compared between these two groups.
RESULTS: The baseline characteristics of the two groups were comparable. There were no significant differences in the mean deviations of postoperative HKA, LDFA, MPTA or PTS between the two groups at 3 months or 1 year (all ps > 0.05). Moreover, there was no significant difference in the percentage of 3° outliers for HKA, LDFA, MPTA, or PTS between the two groups at 3-month or 1-year follow-up (all ps > 0.05). The mean operation time of MAKO was longer than that of ROSA (112.7 ± 12.8 min vs 94.8 ± 23.0 min, p = 0.001), but the mean TBL (1356.7 ± 648.5 mL vs 1384.5 ± 676.3 mL) and transfusion rate (15.0% vs 5.0%) were not significantly different between the two groups (all ps > 0.05). No significant differences were found in postoperative ROM, WOMAC score or KSS score at 1 year (all ps > 0.05).
CONCLUSIONS: The MAKO and ROSA had similar accuracy and precision in TKA alignment. The clinical outcomes at 1 year after surgery were also comparable.