{Reference Type}: Journal Article {Title}: Effectiveness of a standardized quality control management procedure for COVID-19 RT-PCR testing: a large-scale diagnostic accuracy study in China. {Author}: Liu M;Tian C;Chen Y;Zhu J;Zheng Y;Chen J;Li Z;Xu F;Wu L;Wang X;Xie L;Tan X;Cai Y; {Journal}: Diagn Microbiol Infect Dis {Volume}: 109 {Issue}: 2 {Year}: 2024 Jun 28 {Factor}: 2.983 {DOI}: 10.1016/j.diagmicrobio.2024.116287 {Abstract}: BACKGROUND: The study aimed to construct a standardized quality control management procedure (QCMP) and access its accuracy in the quality control of COVID-19 reverse transcriptase-polymerase chain reaction (RT-PCR).
METHODS: Considering the initial RT-PCR results without applying QCMP as the gold standard, a large-scale diagnostic accuracy study including 4,385,925 participants at three COVID-19 RT-PCR testing sites in China, Foshan (as a pilot test), Guangzhou and Shenyang (as validation sites), was conducted from May 21, 2021, to December 15, 2022.
RESULTS: In the pilot test, the RT-PCR with QCMP had a high accuracy of 99.18% with 100% specificity, 100% positive predictive value (PPV), and 99.17% negative predictive value (NPV). The rate of retesting was reduced from 1.98% to 1.16%. Its accuracy was then consistently validated in Guangzhou and Shenyang.
CONCLUSIONS: The RT-PCR with QCMP showed excellent accuracy in identifying true negative COVID-19 and relieved the labor and time spent on retesting.