{Reference Type}: Journal Article {Title}: Deep inspiratory breath-hold radiotherapy on a Helical Tomotherapy unit: Workflow and early outcomes in patients with left-sided breast cancer. {Author}: Nangia S;Burela N;Sawant M;Aishwarya G;Joshua P;Thiyagarajan V;Gaikwad U;Sharma DS; {Journal}: Tech Innov Patient Support Radiat Oncol {Volume}: 30 {Issue}: 0 {Year}: 2024 Jun 暂无{DOI}: 10.1016/j.tipsro.2024.100244 {Abstract}: UNASSIGNED: The clinical implementation of deep inspiratory breath-hold (DIBH) radiotherapy to reduce cardiac exposure in patients with left-sided breast cancer is challenging with helical tomotherapy(HT) and has received little attention. We describe our novel approach to DIBH irradiation in HT using a specially designed frame and manual gating, and compare cardiac substructure doses with the free-breathing (FB) technique.
UNASSIGNED: The workflow incorporates staggered junctions and a frame that provides tactile feedback to the patient and monitoring for manual cut-off. The treatment parameters and clinical outcome of 20 patients with left-sided breast cancer who have undergone DIBH radiotherapy as a part of an ongoing prospective registry are reported. All patients underwent CT scans in Free Breathing (FB) and DIBH using the in-house Respiframe, which incorporates a tactile feedback-based system with an indicator pencil. Plans compared target coverage, cardiac doses, synchronizing treatment with breath-hold and avoiding junction repetition. MVCT scans are used for patient alignment.
UNASSIGNED: The mean dose (Dmean) to the heart was reduced by an average of 34 % in DIBH-HT compared to FB-HT plans (3.8 Gy vs 5.7 Gy). Similarly, 32 % and 67.8 % dose reduction were noted in the maximum dose (D0.02 cc) of the left anterior descending artery, mean 12.3 Gy vs 18.1 Gy, and mean left ventricle V5Gy 13.2 % vs 41.1 %, respectively. The mean treatment duration was 451.5 sec with a median 8 breath-holds; 3 % junction locations between successive breath-holds were replicated. No locoregional or distant recurrences were observed in the 9-month median follow-up.
UNASSIGNED: Our workflow for DIBH with Helical-Tomotherapy addresses patient safety, treatment precision and challenges specific to this treatment unit. The workflow prevents junction issues by varying daily breath-hold durations and avoiding junction locations, providing a practical solution for left-sided breast cancer treatment with HT.