{Reference Type}: Journal Article {Title}: Use of a large-bore 088 intracranial access support catheter for delivery of large intracranial devices: case series with the TracStar LDP in 125 cases. {Author}: Campos JK;Zarrin DA;Meyer BM;Khan MW;Laghari FJ;Collard de Beaufort JC;Amin G;Beaty NB;Bender MT;Suzuki S;Colby GP;Lin LM;Coon AL; {Journal}: J Neurointerv Surg {Volume}: 0 {Issue}: 0 {Year}: 2024 Feb 28 {Factor}: 8.572 {DOI}: 10.1136/jnis-2023-021054 {Abstract}: BACKGROUND: The delivery of neuroendovascular devices requires a robust proximal access platform. This demand has previously been met with a 6Fr long sheath (8Fr guide) that is placed in the proximal internal carotid artery (ICA) or vertebral artery segments. We share our experience with the first 0.088 inch 8Fr guide catheter designed for direct intracranial access.
METHODS: We retrospectively reviewed a prospectively maintained IRB-approved institutional database of the senior authors to identify all cases where the TracStar Large Distal Platform (LDP) was positioned within the intracranial vasculature, defined as within or distal to the petrous ICA, vertebral artery (V3) segments, or transverse sinus. Technical success was defined as safe placement of the TracStar LDP within or distal to the described distal vessel segments with subsequent complication-free device implantation.
RESULTS: Over the 41-month study period from January 2020 to June 2023, 125 consecutive cases were identified in whom the TracStar LDP was navigated into the intracranial vasculature for triaxial delivery of large devices, 0.027 inch microcatheter and greater, for aneurysm treatment (n=108, 86%), intracranial angioplasty/stenting (n=15, 12%), and venous sinus stenting (n=2, 1.6%). All cases used a direct select catheter technique for initial guide placement (no exchange). Posterior circulation treatments occurred in 14.4% (n=18) of cases. Technical success was achieved in 100% of cases. No vessel dissections occurred in any cases.
CONCLUSIONS: The TracStar LDP is an 0.088 inch 8Fr guide catheter that can establish direct intracranial access with an acceptable safety profile. This can be achieved in a wide range of neurointerventional cases with a high rate of technical success.