{Reference Type}: Journal Article {Title}: Distinguishing brain tumors by Label-free confocal micro-Raman spectroscopy. {Author}: Liu J;Wang P;Zhang H;Wu N; {Journal}: Photodiagnosis Photodyn Ther {Volume}: 45 {Issue}: 0 {Year}: 2024 Feb 7 {Factor}: 3.577 {DOI}: 10.1016/j.pdpdt.2024.104010 {Abstract}: BACKGROUND: Brain tumors have serious adverse effects on public health and social economy. Accurate detection of brain tumor types is critical for effective and proactive treatment, and thus improve the survival of patients.
METHODS: Four types of brain tumor tissue sections were detected by Raman spectroscopy. Principal component analysis (PCA) has been used to reduce the dimensionality of the Raman spectra data. Linear discriminant analysis (LDA) and quadratic discriminant analysis (QDA) methods were utilized to discriminate different types of brain tumors.
RESULTS: Raman spectra were collected from 40 brain tumors. Variations in intensity and shift were observed in the Raman spectra positioned at 721, 854, 1004, 1032, 1128, 1248, 1449 cm-1 for different brain tumor tissues. The PCA results indicated that glioma, pituitary adenoma, and meningioma are difficult to differentiate from each other, whereas acoustic neuroma is clearly distinguished from the other three tumors. Multivariate analysis including QDA and LDA methods showed the classification accuracy rate of the QDA model was 99.47 %, better than the rate of LDA model was 95.07 %.
CONCLUSIONS: Raman spectroscopy could be used to extract valuable fingerprint-type molecular and chemical information of biological samples. The demonstrated technique has the potential to be developed to a rapid, label-free, and intelligent approach to distinguish brain tumor types with high accuracy.