{Reference Type}: Journal Article {Title}: Impact of MRI RF coil design on the RF-induced heating of medical implants: fixedB1+rmsexposure versus normal operating mode. {Author}: Yao A;Li Z;Ma M; {Journal}: Phys Med Biol {Volume}: 69 {Issue}: 5 {Year}: 2024 Feb 26 {Factor}: 4.174 {DOI}: 10.1088/1361-6560/ad2714 {Abstract}: A direct comparison of the impact of RF coil design under specific absorption rate andB1+rmslimitations are investigated and quantified using RF coils of different geometries and topologies at 64 MHz and 128 MHz. The RF-inducedin vivoelectric field and power deposition of a 50 cm long pacemaker and 55 cm long deep brain stimulator (DBS) are evaluated within two anatomical models exposed with these RF coils. The associated uncertainty is quantified and analyzed under a fixedB1+rmsincident and normal operating mode. For a fixedB1+rmsincident, thein vivoincident field shows a much higher uncertainty (>5.6 dB) to the RF coil diameter compared to other design parameters (e.g. <2.2 dB for coil length and topology), while the associated uncertainty reduced greatly (e.g. <1.5 dB) under normal operating mode exposure. Similar uncertainties are observed in the power deposition near the pacemaker and DBS electrode. Compared to the normal operating mode, applying a fixedB1+rmsfield to the untested implant will lead to a large variation in the induced incident and power deposition of the implant, as a result, a larger safe margin when different coil designs (e.g. coil diameter) are considered.