{Reference Type}: Journal Article {Title}: Direct immersion single-drop microextraction combined with fluorescence detection using an optical probe. Application for highly sensitive determination of rhodamine 6G. {Author}: Skok A;Bazel Y;Vishnikin A;Toth J; {Journal}: Talanta {Volume}: 269 {Issue}: 0 {Year}: 2024 Mar 1 {Factor}: 6.556 {DOI}: 10.1016/j.talanta.2023.125511 {Abstract}: The use of an optical probe for fluorescence detection combined with direct immersion single-drop microextraction has been demonstrated as an innovative approach. The optical probe served both as a drop holder for extractant and as a measuring device which made it possible to eliminate the use of cuvettes. A laser and a light emitting diode (LED) were tested as possible light sources. Both of them showed comparable results. However, given the much smaller half-band width of the laser radiation, its use has proven to be preferable since background correction can be eliminated. Direct immersion single-drop microextraction of an ionic association complex of rhodamine 6G with picric acid with subsequent fluorescent detection (λex was 532 nm and 525 nm for laser and LED, respectively; λem was 560 nm for both laser and LED) was used a model system to evaluate the new approach. The extractant phase was a 55 μL amyl acetate microdrop fixed in the optical part of the probe. LOD, LOQ and linear calibration range were found as 0.14, 0.48 and 0.5-10 nmol L-1, and 0.15, 0.50 and 0.5-5 nmol L-1 for laser and LED light sources, respectively. The accuracy of the method was assessed by analyzing real water samples.