{Reference Type}: Journal Article {Title}: Use of radar plots to compare the analytical performance of five blood glucose monitoring systems. {Author}: Pardo S;Dunne N;Simmons DA; {Journal}: Ther Deliv {Volume}: 10 {Issue}: 12 {Year}: 12 2019 暂无{DOI}: 10.4155/tde-2019-0046 {Abstract}: Aim: The radar plot is a relatively new way of communicating blood glucose monitoring system (BGMS) accuracy and precision: data points positioned within concentric circles represent the magnitude (increasing with distance from center) and direction (relative to horizontal) of BGMS-error (center = equivalency with reference instrument measurement). This manuscript aims to demonstrate the utility of radar plots as visual tools for interpretation of BGMS analytical performance. Results & methodology: Radar plots were constructed for five BGMSs, to compare BGMS blood glucose results with reference instrument measurements. Conclusion: Radar plots are a useful tool for the visualization of BGMS analytical performance, communicating accuracy, precision and the satisfaction of certain regulatory criteria at a glance.