{Reference Type}: Journal Article {Title}: Revision of linezolid disk diffusion quality control guidelines for testing Staphylococcus aureus ATCC 25923: an independent seven-laboratory trial. {Author}: Biedenbach DJ;Jones RN; {Journal}: Clin Microbiol Infect {Volume}: 9 {Issue}: 10 {Year}: Oct 2003 {Factor}: 13.31 {DOI}: 10.1046/j.1469-0691.2003.00712.x {Abstract}: The accuracy of the published disk diffusion quality control (QC) range for linezolid for Staphylococcus aureus ATCC 25923 has been reported to be problematic, prompting a multicenter study to determine if revision was necessary. Five out of seven laboratories reported values outside the current QC range. Three participants had significant (13.3-60.0%) 'unacceptable' results with zones smaller than the established range. Overall, a range of 21-32 mm was shown (82.4% within the NCCLS published range). A revision (25-31 mm; 95.7% of results) of the QC range is necessary when testing S. aureus ATCC 25923 against linezolid.