%0 Journal Article %T Influence of scanning pattern on accuracy, time, and number of photograms of complete-arch implant scans: A clinical study. %A Gómez-Polo M %A Cascos R %A Ortega R %A Barmak AB %A Kois JC %A Pérez-Barquero JA %A Revilla-León M %J J Dent %V 150 %N 0 %D 2024 Aug 15 %M 39153542 %F 4.991 %R 10.1016/j.jdent.2024.105310 %X OBJECTIVE: To measure the influence of scanning pattern on the accuracy, time, and number of photograms of complete-arch intraoral implant scans.
METHODS: A maxillary edentulous patient with 7 implants was selected. The reference implant cast was obtained using conventional methods (7Series Scanner). Four groups were created based on the scanning pattern used to acquire the complete-arch implant scans by using an intraoral scanner (IOS) (Trios4): manufacturer's recommended (Occlusal-Buccal-Lingual (OBL)), zig-zag (Zig-zag), circumferential (Circumf), and novel pattern that included locking an initial occlusal scan (O-Lock group) (n = 15). Scanning time and number of photograms were recorded. The linear and angular measurements were used to assess scanning accuracy. One-way ANOVA and Tukey tests were used to analyze trueness, scanning time, and number of photograms. The Levene test was selected to assess precision (α=0.05).
RESULTS: Statistically significant differences in trueness were detected among OBL, Zig-zag, Circumf, and O-Lock regarding linear discrepancy (P<0.01), angular discrepancy (P<0.01), scanning time (P<0.01), and number of photograms (P<0.01). The O-Lock (63 ± 20 µm) showed the best linear trueness with statistically significant differences (P < 0.01) with Circumferential (86 ± 16 µm) and OBL (87 ± 19 µm) groups. The O-Lock (93.5 ± 13.4 s, 1080 ± 104 photograms) and Circumf groups (102.9 ± 15.1 s, 1112 ± 179 photograms) obtained lower scanning times (P < 0.01) and number of photograms (P < 0.01) than OBL (130.3 ± 19.4 s, 1293 ± 161 photograms) and Zig-zag (125.7 ± 22.1 s, 1316 ± 160 photograms) groups.
CONCLUSIONS: The scanning patterns tested influenced scanning accuracy, time, and number of photograms of the complete-arch scans obtained by using the IOS tested. The zig-zag and O-Lock scanning patterns are recommended to obtain complete-arch implant scans when using the selected IOS.