%0 Journal Article %T The influence of contrast media on calcium-based imaging of the spine in dual-layer CT. %A Rahn S %A Skornitzke S %A Melzig C %A Reiner T %A Stiller W %A Heussel CP %A Kauczor HU %A Weber TF %A Do TD %J Sci Rep %V 14 %N 1 %D 2024 08 14 %M 39143146 %F 4.996 %R 10.1038/s41598-024-69743-3 %X This study aimed to evaluate the impact of contrast media application on CT attenuation of the bone using a novel calcium-only imaging technique (VCa) from dual-layer spectral detector CT (DLCT), which enables CT-based bone mineral density measurement unimpeded by soft tissue components. For this, true non-contrast (TNC) and venous phase images (VP) of n = 97 patients were acquired. CT attenuation of the first lumbar vertebra (L1) was measured in TNC-VCa, VP-VCa, and in virtual non-contrast images (VNC). CT attenuation was significantly higher in VP-VCa than in TNC-VCa (p < 0.001), although regression analyses revealed a strong linear association between these measures (R2 = 0.84). A statistical model for the prediction of TNC-VCa CT attenuation was established (TNC-VCa[HU] = - 6.81 + 0.87 × VP-VCa[HU]-0.55 × body weight[kg]) and yielded good agreement between observed and predicted values. Furthermore, a L1 CT attenuation threshold of 293 HU in VP-VCa showed a sensitivity of 90% and a specificity of 96% for detecting osteoporosis. The application of contrast media leads to an overestimation of L1 CT attenuation in VCa. However, CT attenuation values from VP-VCa can be used within CT-based opportunistic osteoporosis screening eighter by applying a separate threshold of 293 HU or by converting measured data to TNC-VCa CT attenuation with the given regression equation.