%0 Journal Article %T A multianalyte LC-MS/MS method for accurate quantification of Nitrosamines in Olmesartan tablets. %A Pawar N %A Bhardwaj A %A Vora A %A Sharma S %J J Chromatogr A %V 1732 %N 0 %D 2024 Sep 13 %M 39088899 %F 4.601 %R 10.1016/j.chroma.2024.465176 %X This research summaries the development, optimization and validation of liquid chromatography tandem mass spectrometric (LC-MS/MS) method for concurrent measurement of seven nitrosamines viz; NDMA, NDEA, NDIPA, NDPA, NEIPA, NMPA & NMBA in Olmesartan tablet. Controlling these nitrosamines at trace levels is imperative for ensuring the safety of drug substances and products for consumption. Various regulatory authorities stress the significance of utilizing highly sensitive analytical methods to precisely measure nitrosamines at trace levels. The method applied effective chromatographic separation and optimized parameters for mass spectrometric detection. Detection was carried out using APCI positive ion mode. Chromatographic separation was achieved using a Thermo Accucore PFP column (150 mm x 4.6 mm, 2.6 ยต), with a simple gradient elution of mobile phase consisting of 0.1 % formic acid in water (mobile phase A) and methanol (mobile phase B). The total run time was 20 min, with a flow rate of 0.800 mL/min. The method was validated according to the International Council on Harmonisation (ICH Q2 (R2)) guidelines. The established method demonstrated excellent linearity (R2> 0.99) and sensitivity for all the nitrosamines. Detection and quantification limits were sufficiently low for trace nitrosamine levels having good S/N ratio. The method showed good accuracy in Olmesartan tablet samples, with recoveries ranges between 80 % to 120 %. The new analytical approach has exceptional repeatability and reliability, making it possible to precisely quantify the levels of seven nitrosamines in Olmesartan medoxomil tablets in a single analytical run.