%0 Journal Article %T Repeatability of Automatic Measurements by an Anterior Segment Swept-Source OCT Biometer in Patients With Keratoconus. %A Abicca I %A Schiano-Lomoriello D %A Gilardi M %A Giannini D %A Dinu V %A Coutinho C %A Savini G %J J Refract Surg %V 40 %N 7 %D 2024 Jul %M 39007818 %F 3.255 %R 10.3928/1081597X-20240514-03 %X UNASSIGNED: To evaluate the repeatability of automatic measurements of a new anterior segment optical coherence tomographer (ANTERION; Heidelberg Engineering) and their agreement with an anterior segment optical coherence tomography device combined with Placido disc corneal topography (MS-39; CSO) in patients affected by keratoconus.
UNASSIGNED: Fifty-four consecutive patients were included. Three measurements were performed with the ANTERION and one with the MS-39. Repeatability was assessed by means of within-subject standard deviation, coefficient of variation (CoV), and intraclass correlation coefficient (ICC). Agreement was investigated with the 95% limits of agreement. The paired t-test and Wilcoxon matched-pairs test were performed to compare the measurements of the different devices.
UNASSIGNED: Repeatability of ANTERION measurements was high, with an ICC greater than 0.98 for all parameters. Many parameters revealed a CoV of less than 1% and a CoV within 5% was obtained for astigmatism measurements. The ANTERION measured a significantly higher corneal power and the MS-39 more negative posterior keratometric values. These differences were mirrored by a moderate agreement for mean simulated keratometry and poor agreement for total corneal power and posterior keratometry.
UNASSIGNED: The ANTERION revealed high repeatability of automatic measurements and good agreement with the MS-39 for many parameters in patients affected by keratoconus, but for most parameters the two instruments cannot be considered interchangeable. [J Refract Surg. 2024;40(7):e445-e452.].