%0 Journal Article %T SOFIA Aspiration System as first-line Technique (SOFAST): a prospective, multicenter study to assess the efficacy and safety of the 6 French SOFIA Flow Plus aspiration catheter for endovascular stroke thrombectomy. %A Gandhi D %A Chen H %A Zaidi S %A Sahlein DH %A Maidan L %A Kreitel K %A Miller TR %A Rahimi S %A Al Shekhlee A %A Woo HH %A Toth G %A Schirmer C %A Loh Y %A Fiorella D %J J Neurointerv Surg %V 0 %N 0 %D 2024 Jun 27 %M 38937082 %F 8.572 %R 10.1136/jnis-2024-021811 %X BACKGROUND: Mechanical thrombectomy (MT) is the standard of care for acute ischemic stroke (AIS) patients with large vessel occlusion (LVO). The SOFAST study collected clinical evidence on the safety and efficacy of the 6 French SOFIA Flow Plus aspiration catheter (SOFIA 6F) when used as first-line treatment.
METHODS: This was a prospective, multicenter investigation to assess the safety and efficacy of SOFIA 6F used for first-line aspiration. Anterior circulation LVO stroke patients were enrolled. The primary endpoint was the final modified Thrombolysis in Cerebral Infarction (mTICI)≥2b rate. Secondary endpoints included first-pass and first-line mTICI≥2b rates, times from arteriotomy to clot contact and mTICI≥2b, and 90-day modified Rankin Scale (mRS)≤2. First-line and final mTICI scores were adjudicated by an independent imaging core lab. Safety events were assessed by an independent clinical events adjudicator.
RESULTS: A total of 108 patients were enrolled across 12 centers from July 2020 to June 2022. Median age was 67 years, median National Institutes of Health Stroke Scale (NIHSS) was 15.5, and 56.5% of patients received intravenous thrombolytics. At the end of the procedure, 97.2%, 85.2%, and 55.6% of patients achieved mTICI≥2b, ≥2c, and 3, respectively. With SOFIA 6F first-line aspiration, 87.0%, 79.6%, and 52.8% achieved mTICI≥2b, ≥2c, and 3, respectively. After the first pass, 75.0%, 70.4%, and 50.9% achieved mTICI≥2b, ≥2c, and 3, respectively. Median times from arteriotomy to clot contact and successful revascularization were 12 and 17 min, respectively. At 90 days, 66.7% of patients achieved mRS≤2.
CONCLUSIONS: First-line aspiration with SOFIA 6F is safe and effective with high revascularization rates and short procedure times.