%0 Journal Article %T Evaluation of the QIAstat-Dx Respiratory SARS-CoV-2 Panel for detection of pathogens in nasopharyngeal and lower respiratory tract specimens. %A Caza M %A Hayman J %A Jassem A %A Wilmer A %J Diagn Microbiol Infect Dis %V 110 %N 1 %D 2024 May 24 %M 38906032 %F 2.983 %R 10.1016/j.diagmicrobio.2024.116368 %X This study evaluates the performance of the QIAstat-Dx Respiratory SARS-CoV-2 Panel (RS2P) for the detection of respiratory pathogens. RS2P testing was performed on 440 specimens, including 82 negatives and 358 specimens positive for 1 or more targets (520 targets initially detected). Initial testing was performed on multiple platforms during routine laboratory workflow. Specimens with discordant results on RS2P were re-tested on a different platform to obtain a consensus result based on agreement of 2/3 assays. Percent positive, negative and overall agreement (PPA, PNA, POA), as well as concordance by number of targets and CT value range were calculated. RS2P produced valid results in 439 specimens, with a POA of 91.5 % based on consensus results, with 16/31 (51.6 %) discordant specimens with >1 positive target. When individual targets were examined, PPA, PNA and POA were 93.7 %, 99.9 % and 99.6 % compared to consensus results. Overall, RS2P performed well in detection of respiratory pathogens.