%0 Journal Article %T Magnetic compression anastomosis gastrojejunostomy: feasibility and efficacy of a novel device in a swine model. %A Gagner M %A Krinke T %A Lapointe-Gagner M %A Buchwald JN %J Surg Obes Relat Dis %V 0 %N 0 %D 2024 May 20 %M 38879418 %F 3.709 %R 10.1016/j.soard.2024.04.451 %X BACKGROUND: Means of addressing technical challenges in forming gastrojejunostomy (GJ) anastomoses and maintaining their patency are sought.
OBJECTIVE: Evaluation of preclinical feasibility and healing efficacy of a novel linear magnetic compression anastomosis (MCA) device to form a patent GJ versus sutured jejunal enterotomy (JE) sites in swine.
METHODS: Single-center veterinary testing facility.
METHODS: Feasibility of 3 prototype sizes (4, 6, and 8 cm) of a metal MCA device (MCAD) to form a patent GJ was evaluated over 6 weeks. A distal magnet was laparoscopically inserted in the jejunum, a proximal magnet was placed gastroscopically in the stomach; magnets were aligned to gradually form an anastomosis, self-detached, and be expelled. At necropsy, MCAs were assessed for patency and compared with JE tissues to evaluate wound healing.
RESULTS: MCADs aligned at the GJ location without complications. In 5/6 MCAD pairs, dislodgement occurred between 7 and 26 days; expulsion 13-31 days; 1 MCAD pair was retained in the stomach. At necropsy, all pigs were healthy, gaining a mean 15.0 kg. Anastomoses were not adequately patent in 2/4 pigs receiving the 4-cm or 6-cm MCADs because their linear length was too small. But, anastomoses of both pigs receiving the 8-cm MCADs maintained full patency. Minimal inflammation and fibrosis were seen in MCA specimens versus sutured enterotomies.
CONCLUSIONS: A novel linear MCA device was feasible and effectively created a patent GJ anastomosis in swine with minimal inflammation and fibrosis. The MCAD may be appropriate for clinical evaluation.