%0 Journal Article %T Limitations of glycated albumin standardization when applied to the assessment of diabetes patients. %A Lenters-Westra E %A Atkin SL %A Kilpatrick ES %A Slingerland RJ %A Sato A %A English E %J Clin Chem Lab Med %V 0 %N 0 %D 2024 Jun 17 %M 38874995 %F 8.49 %R 10.1515/cclm-2024-0591 %X OBJECTIVE: Glycated albumin (GA) has potential value in the management of people with diabetes; however, to draw meaningful conclusions between clinical studies it is important that the GA values are comparable. This study investigates the standardization of the Norudia Glycated Albumin and Lucica Glycated Albumin-L methods.
METHODS: The manufacturer reported imprecision was verified by performing CLSI-EP15-A3 protocol using manufacturer produced controls. The Japanese Clinical Chemistry Reference Material (JCCRM)611-1 was measured 20 times to evaluate the accuracy of both methods. GA was also measured in 1,167 patient samples and results were compared between the methods in mmol/mol and %.
RESULTS: Maximum CV for Lucica was ≤0.6 % and for Norudia ≤1.8 % for control material. Results in mmol/mol and % of the JCCRM611-1 were within the uncertainty of the assigned values for both methods. In patient samples the relative difference in mmol/mol between the two methods ranged from -10.4 % at a GA value of 183 mmol/mol to +8.7 % at a GA value of 538 mmol/mol. However, the relative difference expressed in percentage units ranged from of 0 % at a GA value of 9.9 % to +1.7 % at a GA value of 30 %.
CONCLUSIONS: The results in mmol/mol between the two methods for the patient samples were significantly different compared to the results in %. It is not clear why patient samples behave differently compared to JCCRM611-1 material. Valuable lessons can be learnt from comparing the standardization process of GA with that of HbA1c.