%0 Journal Article %T Signal to noise ratio of masticatory muscle activity of functional and non-functional oral tasks. %A Kothari SF %A Devendran A %A Kumar A %A de Caxias FP %A Svensson P %J J Oral Rehabil %V 51 %N 8 %D 2024 Aug 4 %M 38837445 %F 3.558 %R 10.1111/joor.13714 %X BACKGROUND: Electromyographic activity (EMG) of masticatory muscles during wakefulness is understudied. It is unclear if single channel ambulatory EMG devices are sensitive enough to detect masticatory muscle activity (MMA) during wakefulness.
OBJECTIVE: To compare the MMA of various oral tasks recorded with a single channel EMG device ((Grindcare4-datalogger Prototype device) (GC4-β)) and a conventional EMG (cEMG) device.
METHODS: EMG activity of 30 standardised oral tasks was recorded unilaterally from the masseter and anterior temporalis muscle in 24 healthy volunteers using GC4-β and a cEMG device. To compare the EMG data, signal-to-noise ratios (SNR) were calculated as a way to normalise EMG activity across tasks. Analysis of variance was used to compare the SNR between the devices, muscles and oral tasks.
RESULTS: SNR measured from GC4-β was overall significantly higher than the cEMG device (p =.001). The SNR for maximum voluntary contraction (MVC) was significantly higher than all other tasks (p <.001). SNR for temporalis with GC4-β was significantly higher for MVC, hard food, soft food, gum chewing (dominant side), rhythmic clenching and upper lip biting compared to the cEMG device (p <.021). The SNR for masseter with GC4-β was significantly higher for hard food and gum chewing (dominant side), rhythmic clenching, rhythmic biting of an object and yawning compared to the cEMG device (p <.022).
CONCLUSIONS: This study provides novel insight into the EMG patterns of numerous oral tasks enhancing knowledge of physiological differences between the masticatory muscles. Further, single channel EMG devices can effectively measure the EMG activity of various oral tasks during wakefulness.