%0 Journal Article %T Clinical Implementation and Initial Validation of Respiratory-Gated Stereotactic Body Radiotherapy for Thoracoabdominal Tumors Under Abdominal Compression Using an Anzai Laser-Based Gating Device With Visual Guidance on an Elekta Linear Accelerator. %A Hoshina M %A Noguchi M %A Takanashi M %A Masuda K %A Sugahara S %J Cureus %V 16 %N 5 %D 2024 May %M 38832192 暂无%R 10.7759/cureus.59638 %X We have clinically implemented gated stereotactic body radiotherapy under abdominal compression using an Anzai laser-based gating device with visual guidance in combination with an Elekta linear accelerator. To ensure accuracy, we configured the gating window for each patient by correlating the respiratory curve from the laser sensor and the tumor positions from the 4D computed tomography (CT) images reconstructed with the aid of the respiratory curve. This allowed us to define a patient-specific gating window to keep the tumor displacement below 5 mm from the end-expiration, assuming the reproducibility of the tumor trajectories and the laser-based body surface measurements. Results are summarized as follows: 1) A patient-specific gating window internal target volume (ITV) with a prespecified maximum tumor displacement relative to the end-expiration was obtained by acquiring a 4D CT consisting of 20 phase CT sets and a respiratory curve from the Anzai system. 2) Respiratory hysteresis was managed by setting two different thresholds on the respiratory curve based on the predetermined maximum tumor displacement relative to end-expiration. 3) Abdominal compression increased gating window width, thereby presumably leading to faster gated-beam delivery. 4) Gamma index pass rates in sliding-window gated intensity-modulated radiotherapy (IMRT) were superior to those in gated volumetric modulated arc therapy (VMAT). 5) Intrafraction gated cone-beam computed tomography (CBCT) demonstrated that the tumor appeared to remain within the gating window ITV during the stereotactic gated sliding-window IMRT. In conclusion, we have successfully implemented gated stereotactic body radiotherapy at our clinic and achieved a favorable clinical validation result. More cases need to be evaluated to increase the validity.