%0 Journal Article %T Prediction of high-intensity focused ultrasound (HIFU)-induced lesion size using the echo amplitude from the focus in tissue. %A Zhou Y %A Gong X %A You Y %J Phys Eng Sci Med %V 0 %N 0 %D 2024 Jun 1 %M 38822970 %F 7.099 %R 10.1007/s13246-024-01449-2 %X In the realm of high-intensity focused ultrasound (HIFU) therapy, the precise prediction of lesion size during treatment planning remains a challenge, primarily due to the difficulty in quantitatively assessing energy deposition at the target site and the acoustic properties of the tissue through which the ultrasound wave propagates. This study investigates the hypothesis that the echo amplitude originating from the focus is indicative of acoustic attenuation and is directly related to the resultant lesion size. Echoes from multi-layered tissues, specifically porcine tenderloin and bovine livers, with varying fat thickness from 0 mm to 35 mm were collected using a focused ultrasound (FUS) transducer operated at a low power output and short duration. Subsequent to HIFU treatment under clinical conditions, the resulting lesion areas in the ex vivo tissues were meticulously quantified. A novel treatment strategy that prioritizes treatment spots based on descending echo amplitudes was proposed and compared with the conventional raster scan approach. Our findings reveal a consistent trend of decreasing echo amplitudes and HIFU-induced lesion areas with the increasing fat thickness. For porcine tenderloin, the values decreased from 2541.7 ± 641.9 mV and 94.4 ± 17.9 mm2 to 385(342.5) mV and 24.9 ± 18.7 mm2, and for bovine liver, from 1406(1202.5) mV and 94.4 ± 17.9 mm2 to 502.1 ± 225.7 mV and 9.4 ± 6.3 mm2, respectively, as the fat thickness increases from 0 mm to 35 mm. Significant correlations were identified between preoperative echo amplitudes and the HIFU-induced lesion areas (R = 0.833 and 0.784 for the porcine tenderloin and bovine liver, respectively). These correlations underscore the potential for an accurate and dependable prediction of treatment outcomes. Employing the proposed treatment strategy, the ex vivo experiment yielded larger lesion areas in bovine liver at a penetration depth of 8 cm compared to the conventional approach (58.84 ± 17.16 mm2 vs. 44.28 ± 15.37 mm2, p < 0.05). The preoperative echo amplitude from the FUS transducer is shown to be a reflective measure of acoustic attenuation within the wave propagation window and is closely correlated with the induced lesion areas. The proposed treatment strategy demonstrated enhanced efficiency in ex vivo settings, affirming the feasibility and accuracy of predicting HIFU-induced lesion size based on echo amplitude.