%0 Journal Article %T Gafchromic EBT3 film provides equivalent dosimetric performance to EBT-XD film for stereotactic radiosurgery dosimetry. %A Smyth L %A Alves A %A Collins K %A Beveridge S %J Phys Eng Sci Med %V 0 %N 0 %D 2024 May 13 %M 38739345 %F 7.099 %R 10.1007/s13246-024-01430-z %X The accurate assessment of film results is highly dependent on the methodology and techniques used to process film. This study aims to compare the performance of EBT3 and EBT-XD film for SRS dosimetry using two different film processing methods. Experiments were performed in a solid water slab and an anthropomorphic head phantom. For each experiment, the net optical density of the film was calculated using two different methods; taking the background (initial) optical density from 1) an unirradiated film from the same film lot as the irradiated film (stock to stock (S-S) method), and 2) a scan of the same piece of film taken prior to irradiation (film to film (F-F) method). EBT3 and EBT-XD performed similarly across the suite of experiments when using the green channel only or with triple channel RGB dosimetry. The dosimetric performance of EBT-XD was improved across all colour channels by using an F-F method, particularly for the blue channel. In contrast, EBT3 performed similarly well regardless of the net optical density method used. Across 21 SRS treatment plans, the average per-pixel agreement between EBT3 and EBT-XD films, normalised to the 20 Gy prescription dose, was within 2% and 4% for the non-target (2-10 Gy) and target (> 10 Gy) regions, respectively, when using the F-F method. At doses relevant to SRS, EBT3 provides comparable dosimetric performance to EBT-XD. In addition, an S-S dosimetry method is suitable for EBT3 while an F-F method should be adopted if using EBT-XD.