%0 Journal Article %T Robot-assisted total knee arthroplasty system provides more precise control of the femoral rotation angle: A retrospective study. %A Yan P %A Duan X %A Lei Y %A Xing F %A Cao R %A Luo S %A Chen Y %A Liu Z %A Wang K %A Yang P %A Tian R %J Int J Med Robot %V 20 %N 3 %D 2024 Jun %M 38733580 %F 2.483 %R 10.1002/rcs.2635 %X BACKGROUND: Rotational alignment in total knee arthroplasty (TKA) is a crucial technical point that needs attention. We conducted a retrospective study to investigate whether a new robot-assisted TKA (RA-TKA) could improve the accuracy of rotational alignment and whether rotational alignment affects postoperative pain and functional evaluation of the knee.
METHODS: A total of 136 consecutive patients who underwent TKA were included in this study. Half of the patients underwent RA-TKA and the other half underwent conventional TKA (CON-TKA) by the same group of surgeons. Collect the relevant parameters.
RESULTS: The postoperative femoral rotation angle (FRA) was -0.72 ± 2.59° in the robot-assisted group and 1.13 ± 2.73° in the conventional group, and were statistically significantly different (p < 0.001).
CONCLUSIONS: This study provides preliminary evidence that the RA-TKA provides more precise control of FRA than CON-TKA, and verifies that tibial rotation angle and combined rotation angle affect postoperative knee pain and functional evaluation.