%0 Journal Article %T Validation of Artificial Intelligence (AI)-Assisted Flow Cytometry Analysis for Immunological Disorders. %A Lu Z %A Morita M %A Yeager TS %A Lyu Y %A Wang SY %A Wang Z %A Fan G %J Diagnostics (Basel) %V 14 %N 4 %D 2024 Feb 14 %M 38396459 %F 3.992 %R 10.3390/diagnostics14040420 %X Flow cytometry is a vital diagnostic tool for hematologic and immunologic disorders, but manual analysis is prone to variation and time-consuming. Over the last decade, artificial intelligence (AI) has advanced significantly. In this study, we developed and validated an AI-assisted flow cytometry workflow using 379 clinical cases from 2021, employing a 3-tube, 10-color flow panel with 21 antibodies for primary immunodeficiency diseases and related immunological disorders. The AI software (DeepFlowâ„¢, version 2.1.1) is fully automated, reducing analysis time to under 5 min per case. It interacts with hematopatholoists for manual gating adjustments when necessary. Using proprietary multidimensional density-phenotype coupling algorithm, the AI model accurately classifies and enumerates T, B, and NK cells, along with important immune cell subsets, including CD4+ helper T cells, CD8+ cytotoxic T cells, CD3+/CD4-/CD8- double-negative T cells, and class-switched or non-switched B cells. Compared to manual analysis with hematopathologist-determined lymphocyte subset percentages as the gold standard, the AI model exhibited a strong correlation (r > 0.9) across lymphocyte subsets. This study highlights the accuracy and efficiency of AI-assisted flow cytometry in diagnosing immunological disorders in a clinical setting, providing a transformative approach within a concise timeframe.