%0 Equivalence Trial %T Cardiorespiratory monitoring with a wireless and nonadhesive belt measuring diaphragm activity in preterm and term infants: A multicenter non-inferiority study. %A Scholten AWJ %A Zhan Z %A Niemarkt HJ %A Vervoorn M %A van Leuteren RW %A de Jongh FH %A van Kaam AH %A Heuvel ERVD %A Hutten GJ %J Pediatr Pulmonol %V 58 %N 12 %D 2023 Dec 5 %M 37795597 %F 4.09 %R 10.1002/ppul.26695 %X BACKGROUND: We determined if the heart rate (HR) monitoring performance of a wireless and nonadhesive belt is non-inferior compared to standard electrocardiography (ECG). Secondary objective was to explore the belt's respiratory rate (RR) monitoring performance compared to chest impedance (CI).
METHODS: In this multicenter non-inferiority trial, preterm and term infants were simultaneously monitored with the belt and conventional ECG/CI for 24 h. HR monitoring performance was estimated with the HR difference and ability to detect cardiac events compared to the ECG, and the incidence of HR-data loss per second. These estimations were statistically compared to prespecified margins to confirm equivalence/non-inferiority. Exploratory RR analyses estimated the RR trend difference and ability to detect apnea/tachypnea compared to CI, and the incidence of RR-data loss per second.
RESULTS: Thirty-nine infants were included. HR monitoring with the belt was non-inferior to the ECG with a mean HR difference of 0.03 beats per minute (bpm) (standard error [SE] = 0.02) (95% limits of agreement [LoA]: [-5 to 5] bpm) (p < 0.001). Second, sensitivity and positive predictive value (PPV) for cardiac event detection were 94.0% (SE = 0.5%) and 92.6% (SE = 0.6%), respectively (p ≤ 0.001). Third, the incidence of HR-data loss was 2.1% (SE = 0.4%) per second (p < 0.05). The exploratory analyses of RR showed moderate trend agreement with a mean RR-difference of 3.7 breaths/min (SE = 0.8) (LoA: [-12 to 19] breaths/min), but low sensitivities and PPV's for apnea/tachypnea detection. The incidence of RR-data loss was 2.2% (SE = 0.4%) per second.
CONCLUSIONS: The nonadhesive, wireless belt showed non-inferior HR monitoring and a moderate agreement in RR trend compared to ECG/CI. Future research on apnea/tachypnea detection is required.