%0 Journal Article %T Exploring the anatomical factors influencing testes elasticity via ultrasound shear wave elastography: Preliminary results. %A Lin YY %A Mao L %A Li J %A Zhu ZM %A Luo YH %A Zhou XH %A Qiu SD %A Chen F %J Rev Int Androl %V 21 %N 4 %D 2023 Oct-Dec 7 %M 37422973 %F 0.66 %R 10.1016/j.androl.2023.100367 %X The aim of this study is to evaluate the anatomical factors influencing elasticity values of normal testicular parenchyma using shear wave elastography (SWE).
This study examined 68 healthy male volunteers (117 testes in which standard transverse axis ultrasonography views could be obtained) via conventional scrotal ultrasonography and SWE. Both the mean (EMean) and standard deviation (ESD) elasticity values were acquired.
In the standard transverse axis view of the rete testis at the mid-lateral edge of the testes, the EMean values in 2mm the testicular parenchyma from the rete testis and the testicular capsule at the same level as the rete testis were all significantly larger than in the central zone (P<0.001, P<0.001, respectively). The EMean value in the testicular parenchyma 2mm from the testicular capsule on the line formed approximately 45° below the horizontal line of the rete testis was significantly larger than in the rete testis approximately 45° above the horizontal line (P<0.001). In two standard transverse axis views, the ESD values in other regions were significantly larger than those in the central zones (all P<0.001). Also, the EMean values in the transmediastinal arteries were larger than those of the surrounding normal testicular parenchyma (P<0.001).
Based on SWE, factors including the testicular capsule, the density of testicular fibrous septa, the depth of the Q-Box™, and the transmediastinal artery may influence the testes elasticity measurement.