%T Comparison of two new HTLV-I/II screening methods, Abbott Alinity i rHTLV-I/II and Diasorin LIAISON® XL murex recHTLV-I/II, to Abbott architect rHTLVI/II assay.
%A Guiraud V
%A Crémoux F
%A Leroy I
%A Cohier J
%A Hernandez P
%A Mansaly S
%A Gautheret-Dejean A
%J J Clin Virol
%V 164
%N 0
%D 2023 07 14
%M 37148674
%F 14.481
%R 10.1016/j.jcv.2023.105446
%X Diagnosis of Human T-cell Lymphotropic Virus (HTLV) types I and II infection requires sequencial testing with firstly a screening using an Enzyme immunoassay followed by a confirmatory test.
To compare the performances of the Alinity i rHTLV-I/II (Abbott®) and LIAISON® XL murex recHTLV-I/II serological screening tests to the ARCHITECT rHTLVI/II test followed if positive by HTLV BLOT 2.4, MP Diagnostics as the reference.
119 serum samples from 92 known HTLV-I infected patients and 184 from uninfected patients with HTLV were analyzed in parallel with, Alinity i rHTLV-I/II, LIAISON® XL murex recHTLV-I/II and ARCHITECT rHTLVI/II.
Alinity i rHTLV-I/II and LIAISON® XL murex recHTLV-I/II exhibited a total agreement with ARCHITECT rHTLVI/II for both positive and negative samples. Both tests are suitable alternatives for HTLV screening.