%0 Comparative Study %T Antibiotic cement-coated rigid locked nails in infected femoral and tibial nonunion. Reoperation rates of commercial versus custom-made nails. %A Garabano G %A Pereira S %A Alamino LP %A Munera MA %A Ernst G %A Bidolegui F %A Pesciallo CA %J Injury %V 54 %N 0 %D Nov 2023 18 %M 36858895 %F 2.687 %R 10.1016/j.injury.2023.02.033 %X BACKGROUND: The objective of this study is to assess bone union, infection control, and reoperation rates in a series of patients with infected femoral or tibial nonunion treated with antibiotic-cement-coated rigid nails and to compare the results obtained with custom-made nails versus commercial nails.
METHODS: We retrospectively analyzed a series of consecutive patients with infected nonunion of the femur or the tibia treated with antibiotic-cement-coated rigid nails between January 2010 and 2020. We assessed patients' distinctive characteristics, initial injury, type of nail used (custom-made nail with vancomycin or commercial nail with gentamicin), success rate (bone union + infection control), reoperation rate, and failure rate. Comparative analyses were conducted between reoperated and non-reoperated patients regarding the type of nail used. A multivariate regression analysis was performed to assess the risk variables that impacted reoperation rates.
RESULTS: We included 54 patients with 22 (40.74%) infected femoral nonunions and 32 (59.25%) tibial nonunions, who were treated with 38 (70.37%) custom-made antibiotic-cement coated nails and 16 (29.62%) commercial nails. Bone union and infection control were achieved in 51 (94.44%) cases. The reoperation rate was 40.74% (n = 22), and the failure rate was 5.55% (n = 3). The use of custom-made nails was associated with a higher risk of reoperation (Odds Ratio 4.71; 95% Confidence Interval 1.10 - 20.17; p = 0.036).
CONCLUSIONS: Antibiotic-cement-coated nails reached a 94.44% success rate. Nails manufactured in the OR coated with vancomycin cement were associated with a higher risk of reoperation than commercial nails loaded with gentamicin cement.
METHODS: III comparative, observational, non-randomized.