%0 Observational Study %T Feasibility and measurement stability of smartwatch-based cuffless blood pressure monitoring: A real-world prospective observational study. %A Han M %A Lee YR %A Park T %A Ihm SH %A Pyun WB %A Burkard T %A Cho MC %A Camafort M %A Yang E %A Stergiou GS %A Lee HY %A Seo JM %A %J Hypertens Res %V 46 %N 4 %D 04 2023 %M 36781979 %F 5.528 %R 10.1038/s41440-023-01215-z %X Cuffless wearable devices are currently being developed for long-term monitoring of blood pressure (BP) in patients with hypertension and in apparently healthy people. This study evaluated the feasibility and measurement stability of smartwatch-based cuffless BP monitoring in real-world conditions. Users of the first smartwatch-based cuffless BP monitor approved in Korea (Samsung Galaxy Watch) were invited to upload their data from using the device for 4 weeks post calibration. A total of 760 participants (mean age 43.7 ± 11.9, 80.3% men) provided 35,797 BP readings (average monitoring 22 ± 4 days [SD]; average readings 47 ± 42 per participant [median 36]). Each participant obtained 1.5 ± 1.3 readings/day and 19.7% of the participants obtained measurements every day. BP showed considerable variability, mainly depending on the day and time of the measurement. There was a trend towards higher BP levels on Mondays than on other days of the week and on workdays than in weekends. BP readings taken between 00:00 and 04:00 tended to be the lowest, whereas those between 12:00 and 16:00 the highest. The average pre-post calibration error for systolic BP (difference in 7-day BP before and after calibration), was 6.8 ± 5.6 mmHg, and was increased with higher systolic BP levels before calibration. Smartwatch-based cuffless BP monitoring is feasible for out-of-office monitoring in the real-world setting. The stability of BP measurement post calibration and the standardization and optimal time interval for recalibration need further investigation.