%0 Journal Article %T Accurate In Vivo Bowman's Thickness Measurement Using Mirau Ultrahigh Axial Resolution Line Field Optical Coherence Tomography. %A Lawman S %A Mason S %A Kaye SB %A Shen YC %A Zheng Y %J Transl Vis Sci Technol %V 11 %N 8 %D 08 2022 1 %M 35930278 %F 3.048 %R 10.1167/tvst.11.8.6 %X The purpose of this study was to assess the accuracy, repeatability, and performance limits of in vivo Mirau ultrahigh axial resolution (UHR) line field spectral domain (LF-SD) optical coherence tomography (OCT) for the measurement of Bowman's and epithelial thickness, and to provide a reference range of these values for healthy corneas.
Volunteers with no history and evidence of corneal disease were included in this study. An in vivo graph search image segmentation of the central cornea was obtained at the normal interface vector orientation. The Mirau-UHR-LF-SD-OCT system used has an axial resolution down to 2.4 µm in air (1.7 µm in tissue), with an A-scan speed of 204.8 kHz and a signal to noise ratio (sensitivity) of 69 (83) dB.
Nine volunteers were included, one of whom wore contact lenses. The repeatability of mean Bowman's and epithelial thicknesses were 0.3 and 1.0 µm, respectively. The measured 95% population range for healthy in vivo thickness was 13.7 to 19.6 µm for the Bowman's layer, and 41.9 to 61.8 µm for the epithelial layer.
The measured thicknesses of Bowman's layer and the corneal epithelium using the Mirau-UHR-LF-SD-OCT were both accurate, with the range for healthy in vivo thicknesses matching prior confocal and OCT systems of varying axial resolutions, and repeatable, equaling the best value prior reported.
T1. Development of a commercially viable clinical UHR OCT technology, enabling accurate measurement and interpretation of Bowman's and epithelial layer thickness in clinical practice.