%0 Comparative Study %T Combined Dual-Energy and Single-Energy Metal Artifact Reduction Techniques Versus Single-Energy Techniques Alone for Lesion Detection Near an Arthroplasty. %A Chandrasekar S %A Colak C %A Obuchowski NA %A Primak AN %A Karim W %A Subhas N %J AJR Am J Roentgenol %V 215 %N 2 %D 08 2020 %M 32374668 %F 6.582 %R 10.2214/AJR.19.22084 %X OBJECTIVE. The purpose of this study was to compare a combined dual-energy CT (DECT) and single-energy CT (SECT) metal artifact reduction technique with a SECT metal artifact reduction technique for detecting lesions near an arthroplasty in a phantom model. MATERIALS AND METHODS. Two CT phantoms with a cobalt chromium sphere attached to a titanium rod, simulating an arthroplasty, within a background of soft-tissue attenuation containing spherical lesions (range, 10-20 mm) around the head and stem of different attenuations from the background (range of attenuation, 10-70 HU) were scanned with a single CT scanner individually (unilateral) and together (bilateral) with the following three dose-equivalent techniques: the currently used clinical protocol (140 kVp, 300 Reference mAs); 100 kVp; and DECT (100 kVp and 150 kVp with a tin filter). Three radiologists reviewed the datasets to identify lesions. Nonparametric AUC was estimated for each reader with each technique. Multireader ANOVA was performed to compare AUCs. Multiple-variable logistic regression analysis was used to identify factors affecting sensitivity and specificity. RESULTS. Accuracy was lower (p < 0.001) for the DECT 130-keV technique than for the 100-, 70-, and 140-kVp techniques. Sensitivity was higher with unilateral arthroplasties (p = 0.037), with greater contrast differences from background (p < 0.001), and with the SECT 100-kVp technique versus other techniques (p < 0.001). The difference in specificities of modalities was not statistically significant (p = 0.148). CONCLUSION. Combining DECT and SECT techniques does not provide additional benefits for lesion detection as opposed to using SECT alone.