关键词: Inter-observer variability Proton therapy Range uncertainty Sensitivity analysis Setup uncertainty Inter-observer variability Proton therapy Range uncertainty Sensitivity analysis Setup uncertainty

来  源:   DOI:10.1016/j.phro.2021.11.005   PDF(Pubmed)

Abstract:
Margin concepts in proton therapy aim to ensure full dose coverage of the clinical target volume (CTV) in presence of setup and range uncertainty. Due to inter-observer variability (IOV), the CTV itself is uncertain. We present a framework to evaluate the combined impact of IOV, setup and range uncertainty in a variance-based sensitivity analysis (SA). For ten patients with skull base meningioma, the mean calculation time to perform the SA including 1.6 × 104 dose recalculations was 59 min. For two patients in this dataset, IOV had a relevant impact on the estimated CTV D95% uncertainty.
摘要:
暂无翻译
公众号